8.7.1 Basic Specifications
Instrument specifications
Analog input specifications
Logic input specifications
According to the specifications for Logic probe 9320-01, MR9321-01
Measured specifications
8.7 Specifications
Number of inputs
Analog 1 input × 2CH
Logic 4 inputs × 2CH
Maximum sampling rate
400 kS/s
Input terminal
voltage input (CH1, CH2)
Logic input (CH A-1 to CH A-4, CH B-1 to CH B-4)
External trigger input (CH1, CH2)
Output terminal
External trigger output (CH1, CH2)
Power outage protection
After recovering from a power outage, the instrument automatically returns to
the state held before the outage.
Input terminal
Terminal block
Input resistance
1M
Ω
±1%
Input capacity
10 pF ±5 pF (at 100 kHz)
Low pass filter
OFF/5 Hz ±50%/50 Hz ±50%/500 Hz ±50%/5 kHz ±50% (-3dB)
Range
50 V/20 V/10 V/5 V/2 V/1 V/500 mV/200 mV/100 mV/50 mV
Indication possible up to 110% of each range
AD optical resolution
12bit
Accuracy
23 °C ± 5°C, set at 30 minutes after power is turned on.
DC accuracy
; ± 0.5%f.s. (filter 5 Hz ON)
Frequency
; DC - 40 kHz, at 40 kHz -3dB
Temperature
; at DC ± 0.03%f.s./ °C
Guaranteed accuracy
period
1 year
Pre-heating time
30 minute
Noise
2 mVp-p (typ), 4 mVp-p (max) (at 50 mV range, filter OFF, input short circuit)
Common mode removal
comparison
Above 80 dB (50/60 Hz signal source resistance below 100
Ω
)
Input method
Uneven input (floating)
Maximum input voltage
AC33 Vrms DC70 V
Maximum rated voltage to
earth
AC33 Vrms DC70 V
Measured value
The maximum value of an analog waveform in 1 second (Peak), the minimum
value (Bottom), Peak-peak value (pp), maximum electric potential (Ap), crest
Factor (CF), Mean (MEAN), root mean square value (RMS), analog waveform
per second, and instantaneous value of logic waveform (INST)
RMS measurement
Calculation of 1 data per second at a fixed 10 kS (accuracy not specified)
Содержание 2300 Smart Site
Страница 1: ...SMART SITE Instruction Manual 2300 February 2011 Revised edition 5 2300A981 05 11 02H...
Страница 13: ...Contents XII...
Страница 21: ...8 Service...
Страница 87: ...74 4 2304 20 PULSE MODULE...
Страница 105: ...92 5 2304 21 PULSE MODULE...
Страница 145: ...132 8 2321 20 WAVEFORM MODULE 8 1 4 Dimension Diagrams 103 9 1 Unit mm 96 1 110 1 103 3 1 94 6 1 68 5 1 4 1...
Страница 207: ...194 10 2332 20 POWER METER MODULE 10 1 4 Dimension Diagrams 94 5 1 103 3 1 103 9 1 110 1 96 1 45 5 1 4 0 5 Unit mm...
Страница 253: ...240 12 2342 20 OUTPUT MODULE...
Страница 275: ...262 14 2351 AIR MODULE 2352 20 WIRE MODULE...
Страница 285: ...272 15 2353 20 LAN MODULE...
Страница 317: ...304 18 2362 20 DC POWER MODULE...
Страница 345: ...332 20 2392 01 02 MODULE BASE...
Страница 347: ...334 21 Maintenance and Service...
Страница 348: ......
Страница 349: ......
Страница 350: ......
Страница 351: ......
Страница 352: ......
Страница 353: ......
Страница 354: ......
Страница 355: ......
Страница 356: ......
Страница 357: ......
Страница 358: ......
Страница 359: ......
Страница 360: ......
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......
Страница 365: ......
Страница 366: ......