Test
Silver
Gold
LED
Voltage
Leak test
Calibration
Accuracy
A/D noise
Pump inflation
Dump
Valve control
Inflation linearity
Overpressure
Valve/pump current
Residual pressure
Bluetooth
Required equipment
Material no.
Description
Silver
Gold
Qty
407691
Service test box
1
407672
BP test volume repair fixture 113670
1
407841
Adapter, Y tube, PTSS
1
4500-30
Blood pressure hose, 5 ft
1
3400-925
USB cable, 2.4 m (8 ft)
1
200-2000IN
Pressure meter, NETECH
1
103521
Welch Allyn Service Tool CD
1
44 Functional verification tests
Welch Allyn® Connex® ProBP™ 3400 digital blood pressure device