CAUTION
Always protect the LCD. When the main board is outside of the device
case, never rest the board on the LCD.
CAUTION
Observe screw torque specifications, especially with screws that secure
directly into plastic standoffs.
CAUTION
To avoid mismatching screws and holes, keep the screws for each
piece with that piece as you remove modules and circuit assemblies.
Required tools and equipment
•
#1 Phillips bit
•
Torx T10 bit
•
Torque driver calibrated for 3.0 in-lb ±0.5 in-lb
•
Torque driver calibrated for 4.0 in-lb ±0.5 in-lb
•
Torque driver calibrated for 6.0 in-lb ±0.5 in-lb
•
Needlenose pliers
•
Tweezers
•
Spudger/nylon probe
Disassemble the external housing
No.
Item
No.
Item
1
Screw, M3 x 0.5 shoulder
6
Side panel
28 Disassembly and repair
Welch Allyn® Connex® ProBP™ 3400 digital blood pressure device
Содержание Welch Allyn Connex ProBP 3400
Страница 1: ...Welch Allyn Connex ProBP 3400 digital blood pressure device Service Manual Software version 1 04 XX ...
Страница 8: ...4 Symbols Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 12: ...8 Safety Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 20: ...16 Overview Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 22: ...18 Controls and indicators Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 24: ...20 Connections Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 26: ...22 Power up sequence Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 42: ...38 Disassembly and repair Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 46: ...42 Reassembly Welch Allyn Connex ProBP 3400 digital blood pressure device ...
Страница 64: ......