System errors
Symptom
Possible cause
Suggested action
#00000001
A system EEPROM error occurred
Cycle the device power.
Replace the main board.
#00000002
A ModF error occurred
Cycle the device power.
Verify that J22 on the blood
pressure board and J1 on the
main board are properly
connected.
Replace the blood pressure
board.
Replace the main board.