The Vest® Airway Clearance

System, Model 205

Instructions for Use

Содержание Vest 205

Страница 1: ...The Vest Airway Clearance System Model 205 Instructions for Use ...

Страница 2: ......

Страница 3: ...information in this manual is confidential and may not be disclosed to third parties without the prior written consent of Hill Rom The information in this manual is subject to change without notice Hill Rom makes no commitment to update or keep current the information contained in this manual Units with a 51 XXXXX serial number are manufactured by HILL ROM INC 4349 CORPORATE ROAD CHARLESTON SC 294...

Страница 4: ...on 2020 04 First Printing 2006 3M is a trademark of 3M Company Cough Pause and Hill Rom are registered trademarks of Hill Rom Services Inc Hillrom is a trademark of Hill Rom Services Inc Underwriters Laboratories Inc is a registered trademark of Underwriters Laboratories Inc The UL logo is a registered trademark of Underwriters Laboratories Inc The Vest is a registered trademark of Hill Rom Servic...

Страница 5: ...atures 13 Air Pulse Generator 13 Push Bar 14 Remote Control 14 Height Adjustment Release Lever 14 Locking Casters 14 Stand 14 Storage Basket 14 Air Hose Connector Ports 14 Air Hoses 14 Control Panel 14 Contour Foam Optional 14 Assembly 15 Instructions for Use 16 Put on the Disposable Single Patient Use SPU Vests 16 Wrap SPU Vest 16 Full SPU Vest 17 Connect the Air Hoses 17 Set Up the Air Pulse Gen...

Страница 6: ...t and Use the New Ramp Mode 46 Use the Ramp Mode 49 Check the Hour Meter Reading 51 Change the Language 52 Disable the Program Modes Program and Ramp Modes 53 Set Custom Default Settings 54 Move the Air Pulse Generator 57 Raise or Lower the Air Pulse Generator 59 Cleaning 60 General Cleaning 61 Steam Cleaning 61 Cleaning Hard to Clean Spots 61 Disinfecting 62 Maintenance 62 Expected Life 62 Dispos...

Страница 7: ...structions for Use 140643 REV 16 Table of Contents An Air Hose Comes out of the Air Pulse Generator or Disposable Garments during Operation 67 Screen Shows RESTARTING 67 Specifications 68 Classification and Standards 69 Frequently Asked Questions 75 ...

Страница 8: ...vi The Vest Airway Clearance System Model 205 Instructions for Use 140643 REV 16 Table of Contents NOTES ...

Страница 9: ...ainage Therapy According to AARC guidelines specific indications for external manipulation of the thorax include evidence or a suggestion of retained secretions evidence that the patient is having difficulty with the secretion clearance or presence of atelectasis caused by mucus plugging In addition The Vest Airway Clearance System is also indicated for external manipulation of the thorax to promo...

Страница 10: ...identifies special procedures or precautions that persons must obey to help prevent equipment damage PRODUCT SYMBOLS These symbols may or may not be used on your model of The Vest Airway Clearance System Symbol Definition Type B equipment with an F type applied part according to EN 60601 1 The Vest garment is the applied part and the maximum temperature on it will not exceed 107 1 F 41 7 C under o...

Страница 11: ...setting Pressure setting Time setting Up arrow button increases the Frequency Pressure or Time setting Down arrow button decreases the Frequency Pressure or Time setting Remote Control Port Attention Consult accompanying documents Applicable to UL60601 1 compliant units Class II equipment double insulated according to EN 60601 1 Protected against ingress of an object of diameter 12 0 mm and vertic...

Страница 12: ...o 601 1 The Vest Airway Clearance System Model 205 conforms to the European Medical Device Directive 93 42 EEC Units made before May 3 2008 The Vest Airway Clearance System Model 205 conforms to the European Medical Device Directive 93 42 EEC Units made on and after May 3 2008 and before September 14 2018 The Vest Airway Clearance System Model 205 conforms to the European Medical Device Directive ...

Страница 13: ...er Batch number Serial number Medical Device Authorized Representative in the European Community Unique Device Identification Temperature for transport and storage Relative humidity for transport and storage Atmospheric pressure for transport and storage Fragile This way up Keep Dry Symbol Definition ...

Страница 14: ... be stacked on the bottom package Date of manufacture Manufacturer Manufacturer and date of manufacture Non sterile For multiple use on a single patient Physician prescription required for US only Only for products to the Federative Republic of Brazil InMetro Net Connection Corporation NCC logo Tear open here Maximum safe working load Symbol Definition ...

Страница 15: ...The Vest Airway Clearance System Model 205 7 Instructions for Use 140643 REV 16 Symbols Mass of equipment weight Symbol Definition ...

Страница 16: ... of the conditions below carefully consider and assess the patient s case before you decide to use The Vest Airway Clearance System Intracranial pressure ICP 20 mm Hg or patients in whom increased intracranial pressure is to be avoided Uncontrolled hypertension Hemodynamic instability Pulmonary edema associated with congestive heart failure Bronchopleural fistula Subcutaneous emphysema Large pleur...

Страница 17: ...on below to help prevent injury and or equipment damage Warning Unplug this system immediately after you use it Warning Do not use the system near flammable chemicals or products including flammable anaesthetics To do so could cause personal injury or equipment damage Warning Adult supervision is required to use the therapy on children Warning Federal USA law restricts this device to sale by or on...

Страница 18: ...ized therapy regiments involving manually or mechanically assisted coughing or other techniques in conjunction with The Vest Airway Clearance System Model 205 therapy Please consult your physician to determine if additional therapy is appropriate Warning To help prevent cross contamination replace the single patient use disposable garment between patients Warning Only authorized persons should ser...

Страница 19: ...card The Vest Airway Clearance System Model 205 product in accordance with your local government regulations Warning To protect the patient and caregiver from the total acoustical energy from the use of this device make sure of these The device is not used by a patient and or a caregiver more than a total of 12 hours each day When the device is in use keep a distance of at least 1 64 0 5 m between...

Страница 20: ...ommended that patients refrain from doing any therapy for a period of time after eating If patients experience these symptoms it is recommended that they take breaks during therapy Warning Patients that have external ports or equipment for example catheters feeding tubes tracheostomy tubes should take care to cushion the site and or support the equipment before initiating therapy Warning For patie...

Страница 21: ...n correctly connected the Air Pulse Generator controls the disposable garment and supplies pulsations to the chest wall A Air Pulse Generator F Stand B Push Bar G Storage Basket C Remote Control H Air Hose Connector Ports D Height Adjustment Release Lever I Air Hoses E Four Locking Casters J Control Panel Alternate Stand D ...

Страница 22: ...he Locking Casters keep the stand and Air Pulse Generator from unintentionally moving STAND The Stand holds the Air Pulse Generator STORAGE BASKET The Storage Basket can be used to store disposable garments AIR HOSE CONNECTOR PORTS The Air Hoses connect to the Air Hose Connector Ports on the front of the Air Pulse Generator AIR HOSES The Air Hoses connect the disposable garment to the Air Pulse Ge...

Страница 23: ...in the top of the stand 2 Align the holes on the bottom of the Air Pulse Generator with the posts on the rubber mat NOTE The screen on the Air Pulse Generator faces the side of the stand that the basket is on 3 Install the four screws to attach the Air Pulse Generator to the stand 4 Tighten but do not over tighten the four screws with the supplied wrench ...

Страница 24: ...p SPU Vest deflated begin to lay it flat on the bed so it is perpendicular with the patient s torso 3 Roll the patient away from you and on the open portion of the Wrap SPU Vest 4 Put the Wrap SPU Vest in position just beneath the patient s underarms and continue to pull the Wrap SPU Vest around the patient s torso 5 Roll the patient to a supine position 6 Pull both ends of the Wrap SPU Vest aroun...

Страница 25: ...to each other You should be able to fit a fist between the Full SPU Vest and the body before you inflate the garment 4 Make sure there is approximately 3 to 4 8 cm to 10 cm between the bottom edge of the front of the Full SPU Vest and the patient 5 Adjust the hook and loop fasteners at the shoulder straps so the bottom edge of the Full SPU Vest is level with the top of the patient s hip bone CONNE...

Страница 26: ...e Install the foam as follows NOTE The contour foam is not sterile 2 Cut out a center hole to form a donut shape in the foam 3 Put the open hole of the donut directly over the port site 4 Connect the Air Hoses to the disposable garment Wrap SPU Vest a Remove the tear label from the disposable garment ...

Страница 27: ...t to hold the Air Hoses in position NOTE The garment can be worn with the Air Hose Connector Ports anywhere around the chest that is comfortable Full SPU Vest Slide the other end of each Air Hose over the disposable garment Air Hose Ports To help keep the Air Hoses in position slightly twist the Air Hoses as you push them over the Full SPU Vest Air Hose Ports NOTE The Vest Airway Clearance System ...

Страница 28: ...tructions for Use SET UP THE AIR PULSE GENERATOR 1 Install the Remote Control into the Air Pulse Generator NOTE Use of the Remote Control is optional 2 Plug the Power Cord into the power inlet on the back of the Air Pulse Generator 3 Plug the Power Cord into an applicable power outlet ...

Страница 29: ...it This screen shows for 10 seconds or until you press the ON button NOTE To show the total therapy hours press the OFF button when the unit is on the main menu 1 Press the Up arrow above NORMALfor the standard mode The screen will show the previously programmed Frequency Pressure and treatment Time settings or the default settings depending on how it was programmed 2 Confirm the settings on the s...

Страница 30: ...rrow buttons until the prescribed treatmentTime shows on the screen The treatment Time may be set between 1 and 60 minutes NOTE It is not necessary to stop the treatment to adjust the settings If the backlight turns itself off press any button to turn it back ON 3 Press the ON button The disposable garment will inflate NOTE If the unit is left in this mode for 10 minutes it will turn off 4 Press t...

Страница 31: ...n You must press ON to continue to the main screen 8 When the treatment session is complete a The Session Complete message shows b The pulsations stop c The disposable garment deflates d Immediately unplug the system from its power source e Remove the Air Hoses from the Air Hose Ports on the disposable garment f Remove the disposable garment from the patient SETTING AND USING A NEW PROGRAM MODE Ne...

Страница 32: ...g Point 1 proceed to Programming Point 2 by pressing the ON button to move to a previous Programming Point press the OFF button Adjust the settings to match the prescription and repeat for all 8 Programming Points 4 If the prescription does not call for all 8 Programming Points after Programming Point 2 set the Time to 0 MIN for the following segments press the ON button to advance through the rem...

Страница 33: ... garment will deflate b To resume treatment press the ON button or the Remote Control again 9 If it is necessary to end the treatment session before it is complete press the OFF button NOTE Pressing the OFF button does not turn off the screen 10 When the OFF button is pressed the disposable garment deflates and the Incomplete X Min Remain message shows on the screen You must press ON to continue t...

Страница 34: ... will inflate 3 Press the ON button again to begin the treatment program The set treatment Time shows as it counts down to zero for each Programming Point 4 The therapy will automatically end after it has gone through each of the Programming Points 5 If it is necessary to pause the treatment during a session do as follows a Press the OFF button or the Remote Control once The unit will stop the pul...

Страница 35: ...Immediately unplug the system from its power source e Remove the Air Hoses from the Air Hose Ports on the disposable garment f Remove the disposable garment from the patient SETTING AND USING A NEW RAMP MODE Allows the caregiver capability to create a new Ramp Program to ease the patient from a lower setting to a higher setting within the percent of the total therapy session 1 Press the Down arrow...

Страница 36: ...ON button to advance to the next screen 6 Review the settings on the screen Frequency and Pressure for the final setting Ramp End are shown as treatment Time a To adjust the Frequency setting press the left Up or Down arrow buttons until the prescribed setting shows on the screen Frequency or Hertz Hz may be set between the Ramp Start setting and 20 cycles per second b To adjust the Pressure setti...

Страница 37: ...wn to zero NOTE Therapy settings cannot be changed during Ramping stage 10 If it is necessary to pause the treatment during a session do as follows a Press the OFF button or the Remote Control once The unit will stop the pulsations but the settings will continue to show on the screen NOTE When the unit is paused the disposable garment will deflate b To resume treatment press the ON button or the R...

Страница 38: ...e shows b The pulsations stop c The disposable garment deflates d Immediately unplug the system from its power source e Remove the Air Hoses from the Air Hose Ports on the disposable garment f Remove the disposable garment from the patient USING THE RAMP MODE Allows the caregiver to run a previously created Ramp Program See Set and Use the New Ramp Mode on page 46 1 Press the Up arrow above RAMP t...

Страница 39: ...ng a session do as follows a Press the OFF button or the Remote Control once The unit will stop the pulsations but the settings will continue to show on the screen NOTE When the unit is paused the disposable garment will deflate b To resume treatment press the ON button or the Remote Control again 6 If it is necessary to end the treatment session before it is complete press the OFF button NOTE Pre...

Страница 40: ... the disposable garment from the patient CHECK THE HOUR METER READING The Hour Meter Reading will show on the screen when you first plug your device in You can also access the Hour Meter Reading as follows With the device plugged in press the ON button and then press the OFF button The Hour Meter Reading will show on the screen CHANGE THE LANGUAGE 1 Press the Up arrow button above NORMAL The Norma...

Страница 41: ...e the main screen is visible unplug the unit and plug it back in The unit will automatically reconfigure in the applicable language DISABLE THE PROGRAM MODES PROGRAM AND RAMP 1 Press the left Up arrow button above NORMAL The Normal mode screen shows 2 Press the two outside Down arrow buttons for a minimum of 3 seconds or until the screen changes 3 Press the right Up arrow button above Programs 4 P...

Страница 42: ...er be able to see the Program or Ramp mode programs 8 To switch back to Program Mode repeat the steps starting at step 2 NOTE Press the Up arrow button above pgm enable at step 4 to enable the Program Mode it will turn pgm enable to PGM ENABLE Set Facility Custom Default Settings 1 Press the left Up arrow button above NORMAL The Normal mode screen shows 2 Press the two outside Down arrows for a mi...

Страница 43: ...adjust each by using the respective the Up and Down arrow buttons and press the ON button 8 To adjust the default settings for Program Mode set each Programming Point sequentially Press the ON button to advance to the next Programming Point 9 Adjust Frequency Pressure and Time for each Program Point 1 through 8 or change the time to zero after Programming Point 3 to make the previous point the las...

Страница 44: ... the right Down arrow below ACCEPT 15 Once you have accepted the change it will automatically re boot the system and go back to the main screen 16 To switch back to previous patient default mode repeat the steps and select prev pt instead of default USE THE VEST AIRWAY CLEARANCE SYSTEM SOFTWARE VERSION 1 22 Make sure the unit is set up correctly Once power is connected the unit will show a blank s...

Страница 45: ...creen If the settings match the prescribed treatment continue to step 3 Otherwise adjust the settings to match the prescription a To adjust the Frequency setting press the left Up or Down arrow buttons until the prescribed treatment shows on the screen The Frequency or Hertz Hz may be set between 5 and 20 cycles per second b To adjust the Pressure setting press the middle Up or Down arrow buttons ...

Страница 46: ...ion do as follows a Press the OFF button or Remote Control once The unit will stop the pulsations but the settings will continue to show on the screen NOTE When the unit is paused the disposable garment will deflate b To resume treatment press the ON button or the Remote Control again 6 If it is necessary to end the treatment session before it is complete press the OFF button twice NOTE Pressing t...

Страница 47: ...ment SET AND USE THE NEW PROGRAM MODE New Program Mode allows the programming of Program A and Program B modes each with up to eight 8 points with various settings within a single therapy session It also allows the programming of the Cough Pause feature reminder to cough NOTE All the settings can be adjusted You can program up to 8 different Programming Points each with its own combination of Freq...

Страница 48: ... duration time between cough pauses A zero 0 time will turn off the Cough Pause feature 5 Press ON to go to the next screen 6 Press the Up or Down arrow to select the pause duration This will set the duration the unit pauses for coughs 7 Press ON to go to the next screen 8 For Programming Point 1 if the settings match the prescribed treatment continue to step 9 Otherwise adjust the settings to mat...

Страница 49: ...garment will inflate 11 Press the ON button again to begin the treatment program The set treatment Time shows as it counts down to zero for each Programming Point 12 When the time for a cough pause is reached these will happen a The unit will pause b The display will change to show the pause duration timer c The garment will deflate 13 While the pause happens the patient should cough as prescribed...

Страница 50: ...r a cough pause is reached these will happen a The unit will pause b The display will change to show the pause duration timer c The garment will deflate 18 While the pause happens the patient should cough as prescribed 19 When the timer reaches zero the garment will inflate and therapy will resume NOTE If additional time is needed press the OFF control to pause the unit To resume therapy before th...

Страница 51: ...een 22 When the treatment session is complete a The Session Complete message shows b The pulsations stop c The disposable garment deflates 23 Unplug the system from its power source 24 Remove the Air Hoses from the Air Hose Ports on the disposable garment 25 Remove the disposable garment USE THE PROGRAM MODE Runs a previously created New Program Mode See Set and Use the New Program Mode on page 39...

Страница 52: ... garment will inflate 5 Press the ON button again to begin the treatment program The set treatment Time shows as it counts down to zero for each Programming Point 6 The therapy will automatically end after it has gone through each of the Programming Points 7 If it is necessary to pause the treatment during a session do as follows a Press the OFF button or the Remote Control once The unit will stop...

Страница 53: ...prescribed 10 When the timer reaches zero the garment will inflate and therapy will resume NOTE If additional time is needed press the OFF control to pause the unit 11 To resume therapy before the cough pause timer expires you can press ON or squeeze the Remote Control 12 If it is necessary to end the treatment session before it is complete press the OFF button twice NOTE Pressing the OFF button d...

Страница 54: ...he Air Hoses from the Air Hose Ports on the disposable garment 17 Remove the disposable garment SET AND USE THE NEW RAMP MODE Creates a new Ramp Program to ease the patient from a lower setting to a higher setting during a limited period of the therapy session WARNING Warning Consult your physician if not using the prescription as provided Failure to do so can result in personal injury 1 Press the...

Страница 55: ...ed treatment shows on the screen The Frequency or Hertz Hz may be set between 5 and 20 cycles per second b To adjust the Pressure setting press the middle Up or Down arrow buttons until the prescribed treatment Pressure shows on the screen Pressure may be set between 1 and 10 7 Press the ON button to advance to the next screen 8 Review the settings on the screen Frequency and Pressure for the fina...

Страница 56: ...nflate 11 Press the ON button again to begin the therapy The screen will show the word Ramping during the ramp portion of the program The total treatment Time shows as it counts down to zero NOTE Therapy settings can not be changed during Ramping stage 12 If it is necessary to pause the treatment during a session do as follows a Press the OFF button or the Remote Control once The unit will stop th...

Страница 57: ...ue to the main screen 15 When the treatment session is complete a The Session Complete message shows b The pulsations stop c The disposable garment deflates 16 Unplug the system from its power source 17 Remove the Air Hoses from the Air Hose Ports on the disposable garment 18 Remove the disposable garment USE THE RAMP MODE Runs a previously created Ramp Program See Set and Use the New Ramp Mode on...

Страница 58: ...to the next screen 5 Press the ON button The disposable garment will inflate 6 Press the ON button again to begin the program The screen will show the word Ramping during the ramp portion of the program The total treatment Time shows as it counts down to zero NOTE Therapy settings can not be changed during Ramping stage 7 If it is necessary to pause the treatment during a session do as follows a P...

Страница 59: ...essed twice the disposable garment deflates and the Incomplete X Min Remain message shows on the screen You must press the ON button to continue to the main screen 10 When the treatment session is complete a The Session Complete message shows b The pulsations stop c The disposable garment deflates 11 Unplug the system from its power source 12 Remove the Air Hoses from the Air Hose Ports on the dis...

Страница 60: ...ton 2 Press the Up arrow button above NORMAL The Normal mode screen shows 3 Press the two outside Down arrow buttons for a minimum of 3 seconds or until the screen changes 4 Press the middle Down arrow button until the desired language shows in all capital letters 5 To accept the change press the far right Down arrow button 6 Once you have accepted the change the system will automatically re boot ...

Страница 61: ...s by the user 1 Press the ON button 2 Press the left Up arrow button above NORMAL The Normal mode screen shows 3 Press the two outside Down arrow buttons for a minimum of 3 seconds or until the screen changes 4 Press the right Up arrow button above Programs 5 Press the right Up arrow button above pgm disable it will turn to PGM DISABLE 6 To accept the change press the right Down arrow below ACCEPT...

Страница 62: ...NOTE Press the Up arrow button above pgm enable at step 4 to enable the Program Mode it will change pgm enable to PGM ENABLE SET CUSTOM DEFAULT SETTINGS Allows the user to set the default settings for the modes of operation These settings will be the initial settings for all users 1 Press the ON button 2 Press the Up arrow button above NORMAL The Normal mode screen shows 3 Press the two outside Do...

Страница 63: ...nge the default settings of Program A press the Up arrow above PGM A or press the Down arrow below PGM B and go to step 10 c To change the default Ramp settings press the Down arrow below RAMP and go to step 18 d Press ON to accept changes or to advance to the next screen and go to step 21 9 To set default Frequency Pressure and Time settings for Normal Mode adjust each by using the respective Up ...

Страница 64: ...umes therapy after cough pause duration expires 15 Press ON to go to the next screen 16 Adjust Frequency Pressure and Time for each Program Point 1 through 8 or change the time to zero after the last desired Program Point to make the previous point the last point 17 Press the ON button to advance to the next default setting and return to step 8 18 Adjust the Ramp Time Percent for the Default Ramp ...

Страница 65: ...the change it will automatically re boot the system and go back to the main screen 23 To switch back to previous patient default mode repeat steps 3 through 5 and select prev pt instead of default MOVE THE AIR PULSE GENERATOR 1 Unplug the Air Pulse Generator from its power source 2 Stow the power cord on the holder on the back of the stand CAUTION Caution Failure to lower the Air Pulse Generator t...

Страница 66: ... 16 Use The Vest Airway Clearance System Software Version 1 22 6 Press down on the brake release to set the brakes on all four casters 7 Adjust the height of the Air Pulse Generator to the desired position 8 Plug the Air Pulse Generator into an applicable power outlet Brake set ...

Страница 67: ...E AIR PULSE GENERATOR 1 Hold the Height Adjustment Release Lever on the side of the stand 2 Pull the Height Adjustment Release Lever upwards toward to the shelf 3 Push down or pull up on the shelf to adjust the height 4 When the desired height is reached release the Height Adjustment Release Lever Alternate Base press down on foot lever to raise lower shelf ...

Страница 68: ...nt do not use it and replace that component before the next therapy session Warning Do not use harsh cleansers solvents or detergents Warning Use cleaning and disinfectant products in accordance with the manufacturer s instructions We recommend that you clean the unit with detergent and warm water Do not use excessive liquid or harsh cleansers CAUTION Caution Do not use harsh cleansers solvents or...

Страница 69: ...nt We recommend that you clean the Air Pulse Generator and Remote Control with a soft cotton cleaning pad that is slightly moist with detergent and warm water Do not use excessive liquid or harsh cleansers Do not immerse The Vest Airway Clearance System in water or let liquids enter the Air Pulse Generator Make sure to clean the inside and outside of these Ends of the Air Hoses Connection Ports on...

Страница 70: ...e tests and examinations annually Unplug the Air Pulse Generator from its power source Examine the overall condition of the unit for damage or missing parts Examine the Power Cord and connector for cuts scrapes or other damage Do electrical safety tests in accordance with facility protocols Clean and disinfect the unit see Cleaning on page 60 Connect the Air Pulse Generator to a disposable garment...

Страница 71: ...t warnings cleaning and disinfection instructions in this manual Always ensure that the product is unplugged or powered off before disposal Always check and comply with all local national regulations and facility protocols for disposal information If applicable batteries should be recycled Never dispose of batteries which contain substances that can be dangerous for the environment and health Othe...

Страница 72: ...000 Medium package of 1 33 to 43 84 cm to 109 cm P300632000 Large package of 1 43 to 53 109 cm to 135 cm P300633000 Extra Large package of 1 53 to 67 135 cm to 170 cm P300634000 XX Large package of 1 67 to 75 170 cm to 190 cm P300629005 X Small package of 5 19 to 23 48 cm to 58 cm P300630005 Small package of 5 23 to 33 58 cm to 84 cm P300631005 Medium package of 5 33 to 43 84 cm to 109 cm P3006320...

Страница 73: ...ult Small 31 to 36 79 cm to 91 cm P300215000 Adult Medium 36 to 45 91 cm to 114 cm P300220000 Adult Large 45 to 52 114 cm to 132 cm Part Number Description 300177000 Contour foam kit assembly P300177005 Foam contour box of 5 143512 Caster 142232 Remote Control Assembly P004982 Disposable Air Hose 1 single P0049825 Disposable Air Hose 10 case 142319 Power Cord 142200 Rubber Mount Pad 140660 Screw K...

Страница 74: ...e USA call Hill Rom at 800 426 4224 Outside of the USA contact your distributor or local Hill Rom representative or go to www hill rom com When you contact Hill Rom be prepared to give the serial number from the product identification label The product identification label is on the rear panel of the Air Pulse Generator TROUBLESHOOTING WARNING Warning Only authorized persons should service The Ves...

Страница 75: ...e problem is not corrected contact Hill Rom SCREEN SHOWS PLEASE CALL FOR SERVICE MESSAGE The system has experienced an unexpected event Do as follows 1 Unplug the Power Cord from the Air Pulse Generator and then plug it in again 2 If the problem is not corrected contact Hill Rom AN AIR HOSE COMES OUT OF THE AIR PULSE GENERATOR OR DISPOSABLE GARMENTS DURING OPERATION 1 Fully disconnect the Air Hose...

Страница 76: ...t 35 lb 16 kg 22 lb 10 kg Stand height lowest position 29 73 6 cm Stand height highest position 39 99 1 cm Disposable garment material Wrap SPU Vest and Full SPU Vest Polyurethane coated nylon Electrical requirements 100 V AC to 230 V AC 50 to 60 Hz 3 4 Amps 100 V AC 2 0 Amps 230 V AC Fuse requirement 4 A 5 x 20 mm fast blow high breaking fuse Condition Range Temperature 13 F to 158 F 25 C to 70 C...

Страница 77: ...SA C22 2 No 60601 1 CAN CSA C22 2 No 60601 1 11 IEC EN 60601 1 IEC EN 60601 1 6 IEC EN 60601 1 11 IEC EN 62366 1 EN ISO 13485 Technical and Quality Assurance Devices with serial numbers 51 XXXXX and 52 XXXXX UL EN IEC 60601 1 CAN CSA C22 2 No 601 1 EN ISO 13485 Equipment Classification Class II Degree of Protection Against Electric Shock BF with type F applied part Classification According to Dire...

Страница 78: ...uidance RF emissions CISPR 11 Group 1 The Vest Airway Clearance System Model 205 uses RF energy only for its internal func tion Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The Vest Airway Clearance System Model 205 is suitable for use in all establishments including domestic establishments and tho...

Страница 79: ...EC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s Not applica ble Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 0 UT for 0 5 cycle at 0 45 90 135 180 225 270 and 315 0 UT for 1 cycle 70 UT for 25 30 cycles single phase at 0 0 UT for 250...

Страница 80: ...ss than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with this symbol a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast can not be predicted theoretically with accuracy To assess the electromagnetic environ...

Страница 81: ...cations equipment Rated Maximum Output Power of the Transmitter Watts W Separation Distance According to the Frequency of the Transmitter Meters m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 7 GHz 0 01 W 0 12 m 0 04 m 0 07 m 0 1 W 0 37 m 0 11 m 0 22 m 1 W 1 17 m 0 35 m 0 70 m 10 W 3 69 m 1 11 m 2 21 m 100 W 11 67 m 3 50 m 7 00 m For transmitters rated at a maximum output power not listed abov...

Страница 82: ...27V m 50 PM 18Hz 0 3 Front back left right 450 28V m FM 5kHz 1kHz 0 3 Front back left right 710 9V m 50 PM 217Hz 0 3 Front back left right 745 9V m 50 PM 217Hz 0 3 Front back left right 780 9V m 50 PM 217Hz 0 3 Front back left right 810 28V m 50 PM 18Hz 0 3 Front back left right 870 28V m 50 PM 18Hz 0 3 Front back left right 930 28V m 50 PM 18Hz 0 3 Front back left right 1720 28V m 50 PM 217Hz 0 3...

Страница 83: ... s physician will prescribe a schedule for the individual patient Is chest physiotherapy still necessary Refer to the physician s order Is postural drainage still necessary Refer to the physician s order What will happen if a patient s treatment session is skipped One missed treatment session of the system may or may not immediately impact the patient s health However the key to help maintain the ...

Страница 84: ...76 The Vest Airway Clearance System Model 205 Instructions for Use 140643 REV 16 Frequently Asked Questions NOTES ...