187692(10) - EN - P280 Mattress Instructions for Use
Electromagnetic Emissions Compliance
Electromagnetic Immunity Compliance
Manufacturer's recommendations and declaration – electromagnetic emissions
The P280 Mattress is intended for use in the electromagnetic environment specified
below. The customer or the user of the P280 should make sure it is used in such an
environment.
Emission test
Compliance Electromagnetic environment - Recommendations
RF emissions
CISPR 11
Group 1
P280 Mattress uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The P280 Mattress is suitable for use in all
establishments.
Harmonic emissions
IEC 61000-3-2
Class A
Emissions of voltage
fluctuations/flicker
IEC 61000-3-3
Compliant
Manufacturer's recommendations and declaration – electromagnetic immunity
The P280 Mattress is intended for use in the electromagnetic environment specified
below. The customer or the user of the P280 Mattress should make sure it is used in such
an environment.
Immunity Test
Test Level
IEC 60601
Compliance
Electromagnetic Environment -
Recommendations
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV on contact
± 2 kV, ± 4 kV, ± 8 kV
and ± 15 kV in the air
± 8 kV on contact
± 2 kV, ± 4 kV, ± 8 kV
and ± 15 kV in the air
The relative humidity must be at
least 5%.
Electric fast
transients / Burst
IEC 61000-4-4
± 2 kV for the power
supply lines
±1kV for the
input/output lines
(100 kHz Repetition
Frequency)
± 2 kV for the power
supply lines
± 1 kV for the
input/output lines
(100 kHz Repetition
Frequency)
Mains power quality should be
that of a typical commercial or
hospital environments.
Surge
IEC 61000-4-5
±1 kV between
line(s)
±2 kV between
line(s) and earth
±1 kV between
line(s)
±2 kV between
line(s) and earth
The quality of the electric power
supply should be the equivalent
of that found in commercial or
hospital environments.
Magnetic field at
the frequency of
the mains power
supply
(50/60 Hz)
IEC 61000-4-8
30 A/m
60 Hz
50 Hz
30 A/m
60 Hz
50 Hz
Power frequency magnetic
fields should be at levels
characteristics of a typical
location in a typical commercial
or hospital environment.