Hillrom Helion Rack Скачать руководство пользователя страница 1

Instructions for use

Helion

Video Management System

ENGLISH

en-US

Read the instructions for use carefully before using the 
product and keep them safe for future reference.

Содержание Helion Rack

Страница 1: ...Instructions for use Helion Video Management System ENGLISH en US Read the instructions for use carefully before using the product and keep them safe for future reference ...

Страница 2: ...This page is intentionally blank ...

Страница 3: ...he following sales units The manual is provided by VIDEOMED S r l in electronic PDF format on digital media A paper copy of the manual is available upon request for qualified technical and medical personnel Telephone 39 049 9819113 Fax 39 0434 030689 Document number 80028000 Language ID 030 Version C Material number 773607 Date of publication 2021 09 01 Product designation REF Helion Main Unit R V...

Страница 4: ...elion 4 80028000_030_C 773607 2021 09 01 VIDEOMED S r l declines any liability for improper use of the system and or for damage caused as a result of operations not covered by the technical documentation ...

Страница 5: ... by the user GENERAL CONSIDERATIONS All operating instructions and recommendations described in this manual must be complied with Clinical personnel must be trained on all operating procedures and safety standards prior to using the system SIGNAL WORDS Residual dangers that may occur while using the product are identified in the document with a signal word The safety measures required and potentia...

Страница 6: ...Helion 6 80028000_030_C 773607 2021 09 01 This page is intentionally blank ...

Страница 7: ...l devices 23 4 4 Obligations and prohibitions 23 4 4 1 Personnel prohibitions 24 4 5 Technical data 25 4 6 Measurement and weight layout 30 4 7 System components 35 4 7 1 Main unit 35 4 7 2 Conference unit 36 4 7 3 4K unit 36 4 7 4 4K Plus unit 36 4 7 5 Control software 37 5 Operation 38 5 1 First system start 38 5 2 Preliminary checks 38 5 3 System start up 38 5 4 Connection to the sources 39 5 5...

Страница 8: ...electing a patient present in the list 66 6 6 1 2 Entering a new patient 67 6 6 1 3 Entering an emergency patient 68 6 6 1 4 Searching for a patient present in a list 69 6 6 1 5 Patient master data modification 69 6 6 1 6 Accessing the worklist 70 6 6 2 Check List 71 6 6 3 Preset 72 6 6 3 1 Preset setting 73 6 6 3 2 Enabling Preset 74 6 6 4 Multiview 75 6 6 4 1 Multiview setting 76 6 6 5 Audio con...

Страница 9: ...eomed srl Via Cesare Battisti 31 C 35010 Limena Italy ALGOAOAIBKHIALFNCJDK AHFINKEOJEDJDJIEBEFK AJDKNDFDKNLJIJJGLDMK AMFBIMFBAFFHHMEDHDJK ACOGAOKEKCMKIOMKICEK 01 00615521031695 21 1234567890 11 210811 7d hillrom co uk 1234567890 2021 08 11 Made In Italy FUSES RATING 2X 250VAC T3 15A E464244 INPUT 100 240V 1 1 2 0A 50 60 Hz VR401111 1DT Helion Main Unit RSD 116001 VIDEOMED MANAGEMENT SYSTEM Videome...

Страница 10: ...1 2 0A 50 60 Hz VR401111 2T Helion Main Unit ARS 116001 VIDEOMED MANAGEMENT SYSTEM Videomed srl Via Cesare Battisti 31 C 35010 Limena Italy ALGOAOAIBKHIALFNCJDK AHFINKEOJEDJCKODNNFK AKFKNDFCKPKFEMPFGLMK AMFAJJEBFMDJKEFKPFJK ACOGEGIMKKMOEOKGEKEK 01 00615521031619 21 1234567890 11 210811 7d hillrom co uk 1234567890 2021 08 11 Made In Italy FUSES RATING 2X 250VAC T3 15A E464244 INPUT 100 240V 1 1 2 0...

Страница 11: ... 11 Made In Italy FUSES RATING 2X 250VAC T1 6A E464244 INPUT 100 240V 0 2 0 48A 50 60 Hz AC500920K 2 Helion Rack 230V 800116 COMPONENT ACCESSORY Videomed srl Via Cesare Battisti 31 C 35010 Limena Italy ALGOAOAIBKHKALFNCJDK AHFINKEOJADJDJLDOOFK AKAINDFCLNMIELOJICFK AMFBJMBJFOHEHNJNGAJK ACOKAKAOACMCAOEMEKEK 01 00615521031893 21 1234567890 11 210903 7d hillrom co uk 1234567890 2021 09 03 Made In Ital...

Страница 12: ...Processes ANSI AAMI IEC 62366 1 2015 Medical devices Part 1 Application of usability engineering to medical devices AAMI ANSI HE75 2009 Human Factors Engineering Design Of Medical Devices AAMI ANSI ES60601 1 2005 R 2012 and A1 2012 C1 2009 R 2012 and A2 2010 R 2012 Medical electrical equipment part 1 general requirements for basic safety and essential performance IEC 60601 1 2005 Mod General II ES...

Страница 13: ...ription Regulation EU 2017 745 Regulation on medical devices amending Directive 93 42 EEC will enter into force on 26 May 2021 93 42 EEC Medical Devices Directive MDD and f m 2007 47 EC EN 1041 2008 Information provided by the manufacturer of medical devices EN ISO 13485 2016 Management systems for the quality of medical devices EN ISO 14971 2012 Applying risk management to medical devices EN ISO ...

Страница 14: ... or resale It is the user s responsibility to keep this documentation intact so that it may be consulted throughout the entire lifespan of the system 2 2 Updates VIDEOMED S r l reserves the right to update the instructions for use at any time with modifications and or translations without prior notice Contact the Customer Service office of VIDEOMED S r l to be updated on the latest version of the ...

Страница 15: ...peration of the system and ensures that it is used solely for the intended purpose Qualified Personnel Authorized persons who are generally employees of the manager or who have acquired their skills through professional training in the medical sector who are able to evaluate their work and recognize potential risks based on their professional experience and knowledge of safety regulations Where re...

Страница 16: ...e a medical device Indicates the Unique Device Identification UDI code is composed of UDI DI 01 and UDI PI 11 date of production 21 serial number MEDICAL GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH ANSI AAMI ES60601 1 2005 AMD 1 2012 CAN CSA C22 2 No 60601 1 2014 Consult instructions for use IFU A copy of the IFU is available on this website...

Страница 17: ... Lead and Di 2 ethylhexyl phthalate DEHP which are known to the State of California to cause cancer and Lead and Di 2 ethylhexyl phthalate DEHP which are known to the State of California to cause birth defects or other reproductive harm For more information go to www P65Warnings ca gov CAUTION All safety information must be complied with for a safe use of the Helion Video Management System CAUTION...

Страница 18: ...ivities on the electrical equipment carried out in a non compliant manner or incorrect replacement of components may compromise the efficiency of the solutions adopted The Helion product is a Class A electromedical device according to CEI 60601 1 2 CISPR 11 it is suitable for use in a specific electromagnetic environment The customer and or the user of the product must ensure that it is used in an...

Страница 19: ... level The quality of the mains supply must be that typical of a commercial and or hospital environment Pulses IEC 61000 4 5 0 5 1 kV differential mode 0 5 1 2 kV in common mode IEC 60601 1 2 Test level Mains power quality should be typical of a commercial or hospital environment Conducted disorders induced by RF fields IEC 61000 4 6 3 V 150 kHz to 80 MHz 6V ISM frequencies IEC 60601 1 2 Test leve...

Страница 20: ...the assistance service and the availability of spare parts VIDEOMED applies a certified quality management system according to EN ISO 13485 to all its business processes which is a guarantee of highest quality product and accessory reliability ease of use functional design optimization for the intended purpose Test frequency MHz Modulation Minimum IMMUNITY level V m Applied IMMUNITY level V m 385 ...

Страница 21: ...ion Inform the operator of any defects 3 4 1 Preparing the system Pull the plug of the power cord out of the socket For permanent installations turn off the main switch place on the front panel of the rack 3 4 2 Cleaning the system 1 Wear the required personal safety equipment 2 Using suitable means remove coarse dirt from the rack or the units for the installation without rack 3 With a clean dry ...

Страница 22: ...by the operator Repair and disposal of the MD must be performed by qualified service technicians as required To control the on off and intensity of compatible surgical lights Contraindications The system shall not be used for the purpose of findings and diagnostics The system shall not be used for checking vital body functions The system shall not be used to make reports The product has not been d...

Страница 23: ...be used with medical devices from other manufacturers The operation of these devices is described in the relevant operating instructions Only medical devices e g Room Camera compliant with IEC standard 60601 1 can be connected to the system If a medical device is installed at a later date installation must be carried out as specified in IEC standard 60601 1 Environmental conditions for the operati...

Страница 24: ...fy system components without the permission of the Manufacturer use the system as a support point even if not operational resulting in the risk of falling and or damaging the system itself use the system outside the room conditions permitted see paragraph Technical data CAUTION VIDEOMED S r l is not liable for damage caused to things or persons if it is ascertained that the system has been used in...

Страница 25: ...tage Input Output 4000 V AC Input FG 1500 V AC Enclosure IP20 Environmental conditions Operating temperature 10 40 C Operating relative humidity range 30 to 75 Operating atmospheric pressure range 54 0 kPa to 106 0 kPa Storage temperature 40 70 C Storage relative humidity range 10 to 100 including condensation Storage atmospheric pressure range 50 0 kPa to 106 0 kPa Max Operation Altitude 5000 mt ...

Страница 26: ... W Protection Short circuit protection Overload protection Overcurrent protection Overvoltage protection Insulation Voltage Input Output 4000 V AC Input FG 1500 V AC Environmental conditions Operating temperature 10 40 C Operating relative humidity range 30 to 75 Operating atmospheric pressure range 54 0 kPa to 106 0 kPa Storage temperature 40 70 C Storage relative humidity range 10 to 100 includi...

Страница 27: ...ply 100 240 V 50 60Hz AC Power Consumption 30 W Protection Short circuit protection Overload protection Overcurrent protection Overvoltage protection Insulation Voltage Input Output 4000 V AC Input FG 1500 V AC Environmental conditions Operating temperature 10 40 C Operating relative humidity range 30 to 75 Operating atmospheric pressure range 54 0 kPa to 106 0 kPa Storage temperature 40 70 C Stor...

Страница 28: ... 450 mm Power Supply 100 240 V 50 60Hz AC Power Consumption 30 W Protection Short circuit protection Overload protection Overcurrent protection Overvoltage protection Insulation Voltage Input Output 4000 V AC Input FG 1500 V AC Environmental conditions Operating temperature 10 40 C Operating relative humidity range 30 to 75 Operating atmospheric pressure range 54 0 kPa to 106 0 kPa Storage tempera...

Страница 29: ...Operating atmospheric pressure range 70 0 kPa to 106 0 kPa Storage temperature 40 70 C Storage relative humidity range 10 to 100 including condensation Storage atmospheric pressure range 50 0 kPa to 106 0 kPa Rack internal component Two fans forced ventilation produced 2410 min Cubic Meters per Minute CMM each Rack internal component Isolation transformer power 1000 VA Max Operation Altitude 3000 ...

Страница 30: ...System description 30 80028000_030_C 773607 2021 09 01 4 6 Measurement and weight layout MAIN UNIT Dimensions 133 x 430 x 450 mm Unit Weight 13 5 kg Rack Dimensions optional Rack Brackets optional ...

Страница 31: ...System description 80028000_030_C 773607 2021 09 01 31 CONFERENCE UNIT Dimensions 44 x 430 x 450 mm Unit Weight 8 kg Rack Brackets optional Rack Dimensions optional ...

Страница 32: ...System description 32 80028000_030_C 773607 2021 09 01 4K UNIT Dimensions 44 x 430 x 450 mm Unit Weight 5 5 kg Rack Brackets optional Rack Dimensions optional ...

Страница 33: ...System description 80028000_030_C 773607 2021 09 01 33 4K PLUS UNIT Dimensions 44 x 430 x 450 mm Unit Weight 5 5 kg Rack Brackets optional Rack Dimensions optional ...

Страница 34: ...System description 34 80028000_030_C 773607 2021 09 01 RACK UNIT Dimensions 800 x 600 x 757 mm Unit Weight 64 kg ADJUSTABLE FOOT HEIGHT ...

Страница 35: ...tion ports are located on the back of the unit They are divided into the following sections The connection cables are supplied VIDEOMED S r l 1 MAIN UNIT 2 CONFERENCE UNIT 3 4K UNIT or 4K PLUS UNIT 1 2 3 Function Description ROUTING Enables to distribute the different sources present in the room to the recipient monitors PROCEDURE DOCUMENTATION It enables to document the operation by temporarily s...

Страница 36: ...EOMED S r l 4 7 3 4K unit The new identification plates are still missing The 4K unit allows full management of signals with 4K Ultra HD resolution The following connection ports are located on the back of the unit They are divided into the following sections The connection cables are supplied VIDEOMED S r l 4 7 4 4K Plus unit The 4K Plus unit allows full management of signals with 4K Ultra HD res...

Страница 37: ...PTZ Roomcam surgical lights with surgical video camera The complete functions present in the Control Software are described in detail in the User Interface chapter of these instructions for use 1 2 3 No Function Description Image 1 VIDEO ROUTING The BLUE button identifies the section of the VIDEO ROUTING function This function allows the distribution of video signals connected to all monitors inst...

Страница 38: ...s The delivery of the Helion Video Management System is validated by a document signed by the operator Once the system has been commissioned the instructions contained in this manual are binding for the user 5 2 Preliminary checks Before each use check the components of the control screen in regards to monitor assembly stability loose parts on the monitor body visible damage in particular abrasion...

Страница 39: ...sent DVI 3G HD SD SDI CVBS Composite VGA The technical drawing shows an example of the connection plates installed on the pendant panels 5 5 System shutdown To shut down the system proceed as follows In the event of system shutdown it is possible to force system deactivation by keeping the finger on the button until it is completely switched off and then restart the device by following the start u...

Страница 40: ... off button installed inside the operating room typically on a pendant panel or wall unit Thanks to this solution the operator can manage the entire video management system without having to access the Technical rack The switch off of the units inside the Rack will therefore be reserved for technical and authorized trained VIDEOMED personnel for service or maintenance sessions ...

Страница 41: ...e video conference the streaming and the advanced modes Privacy Mode Do Not Disturb and Lecture Mode are also present 2 MAIN SECTION Area for the functions to select the source and identify the monitors The structure of the area varies according to the control functions activated 3 SELECTION BAR Bar located at the bottom of the screen with which the following functions can be selected Video Routin...

Страница 42: ...e monitor settings can be accessed brightness intensity of the entire display screen contrast the difference in brightness between the different light and dark areas of the screen The menu control settings are located to the side or at the bottom of the monitor depending on the model purchased Refer to the user manual of the touchscreen monitor for further information Information regarding the ser...

Страница 43: ... signals can be routed to any monitor in the operating room The main Video Routing screen is divided as follows To send a video signal to a monitor drag the relevant image from the Source List A available and drop it into one of the enabled monitors B by using the Drag Drop system The Preview of the video signal sent will be displayed in the relative Monitor icon and updated periodically A B A lis...

Страница 44: ...rge or reduce the preview of the video signal of each connected source If the preview is enlarged it will only appear momentarily on the screen To display the Live Preview of one of the signals available in the Source List proceed as follows Step Image 1 Press the icon in the corresponding Preview The enlarged image selected is displayed 2 Press to return to standard view of the Video Routing sect...

Страница 45: ...ayed signal If the icon is gray it indicates that the function is not active To activate the function it is required to select a patient from the list refer to paragraph Selecting a patient present in the list Enables to create snapshots of the image Enables to start stop video signal streaming Enables to activate the full screen display function without latency of the selected source function ava...

Страница 46: ... possible to use an additional quick activation system from the Video Routing function To perform the recording from the Video Routing function proceed as follows In systems where the dual recording channel is enabled this function can be activated on two sources at the same time Step Image 1 Press the box of a Preview to start recording a signal When recording is in progress the button is red The...

Страница 47: ...m one of the connected sources When streaming is active the button is in white background in the box of the selected source and is disabled in the Previews of the remaining sources Pressing in a Preview box also activates the icon that shows the user the link to connect to the streaming session Therefore by using this link each user is able to connect to the streaming session by using applications...

Страница 48: ...st the zoom of the Roomcam proceed as follows 6 3 4 2 Roomcam movement adjustment To adjust the movement of the Roomcam proceed as follows Step Image 1 Use and to adjust the zoom to obtain the desired image view Step Image 1 Use the arrows in the screen to adjust the movement of the Roomcam ...

Страница 49: ...ng Preset To save a specific video camera setting Preset proceed as follows Step Image 1 After adjusting the video camera to the desired position press 2 Enter the name to be assigned and press to confirm The new Preset with the name assigned will appear on the list on the side ...

Страница 50: ...e 50 80028000_030_C 773607 2021 09 01 6 3 4 4 Delete camera setting Preset To delete a video camera setting from the Preset list proceed as follows Step Image 1 Press 2 Press and hold the button until it is deleted ...

Страница 51: ...therefore possible to save and then modify images and videos stored on a 1TB on board support expandable up to 2TB as an option It is then possible to send the recorded material to a dedicated server connected storage systems such as PACS network or mobile storage media The Recording function includes recording still images video recording audio included image and video post processing Step Image ...

Страница 52: ... surgical activity images and video at any time by pressing the icon This way a list will appear on the screen containing all previews of stored files which can then be reproduced and processed by using the functions described in paragraph Snapshot and video playback C A B A source list B view of the two recording channels C list of snapshots and videos stored ...

Страница 53: ...get a Live Preview of the signal and the basic and advanced recording functions will be enabled The following icons are present in the Recording Channel window Channel selection and recording will in no way affect the signals sent to the monitors via Video Routing Icon Function Enables to start stop recording of the displayed signal If the icon is gray it indicates that the function is not active ...

Страница 54: ...dedicated to them Through the green number in the icon it will be possible to see how many media have been associated with that patient Click on the icon to access the storage folder Step Image 1 Press the box of a Recording Channel to start recording a signal When recording is in progress the button is red in the Recording Channel selected The icon is also present in the upper part of the screen ...

Страница 55: ...videos related to the selected patient 2 Press the icon to enlarge the single media 3 A new window appears which depending on the file selected image or video allows you to Icon Function Delete the file Add a comment Play the video Interrupt video playback Extract a part of the video Return to the full view of available functions Create a snapshot from the video being played ...

Страница 56: ...eos proceed as follows Step Image 1 Press to view all images and videos related to the selected patient 2 Press the icon to enlarge the single element 3 Press the icon A cursor appears on the progress bar Press to select the starting point of the video to be extracted and for the end ...

Страница 57: ...Interface 80028000_030_C 773607 2021 09 01 57 4 Once the cutting points are determined extract the portion of the video by pressing A new video appears in the photo video list of the same patient Step Image ...

Страница 58: ... the captured images and videos is displayed The following icons are present in the Export window The standard system configuration involves exporting to a device connected to the USB port To export proceed as follows Icon Function It enables to export to the PACS system Enables to save on a device connected to the USB port It enables to export via LAN to another device Step Image 1 Select the exp...

Страница 59: ...e g USB device not present the system displays an EXPORT FAILED error message and the icon appears on the right side of the screen The files that the system failed to export are indicated with a red selection icon as shown in the figure below 2 Press on the right side of the screen to send the files to the selected destinations By pressing the files are sent to the selected destinations and the pa...

Страница 60: ...deos of the selected patient To delete images and videos proceed as follows Step Image 1 Select the videos or images you wish to delete by clicking directly on their thumbnail The outline of the selected thumbnails will be green 2 Press on the right side of the screen to delete the files selected 3 Press to confirm deletion of the selected files ...

Страница 61: ... transmission channels Images or video sources All connected sources are displayed in the input signal bar Swap Button During a conference it is possible to change the signals displayed in the layout selected Layout Button During a multi channel conference it is possible to have various live Previews of the video signals involved for example PiP and PaP Participant selection Contact list The video...

Страница 62: ... the video conference is active the dashboard receiver icon turns green 6 5 1 Select the signals to be sent by video conference From the Source List drag the source you wish to send in video conference in the Primary Channel or Secondary Channel box A B A source list B view of the two video conference channels ...

Страница 63: ...deo conference Press one of the boxes related to the primary and or secondary video conference channel and then to the icon which will be shown inside it to remove the video signal from the video conference The video signal thus removed will no longer be shared with video conference participants ...

Страница 64: ...lative icon depending on the mode on the right side of the screen The icons are described below Icon Description Image Enables to select a name from the contact list Enables to select a name address from those indicated in the calls sent received log Enables to manually enter the recipient s IP address by using the numeric keypad ...

Страница 65: ...Allows access to the screens related to patient data management Allows access to the Check List screens relating to the surgery Allows access to the Preset Workflows screens of the Room configuration Allows access the Multiview setup screens Allows access the Audio setup screens Allows access to the management screen of the operating lights in the operating room This function can only be used if a...

Страница 66: ...the screen displayed when the icon is pressed the list of patients A already entered and divided between those imported by the Worklist if present and those previously manually selected or inserted is displayed 6 6 1 1 Selecting a patient present in the list To select a patient already in the list proceed as follows A Step Image 1 Select the patient ...

Страница 67: ...w patient To enter a new patient proceed as follows 2 Press to confirm the selection Step Image Step Image 1 Select the icon on the right of the screen The window for filling in patient data is displayed 2 Enter the data for the new patient Fields marked with are mandatory ...

Страница 68: ...tionalities available and management the Emergency Patient is comparable to any patient manually entered or by recalling the Worklist To enter an emergency patient proceed as follows 3 Once the mandatory data have been filled in it is possible to save the new patient by pressing or cancel the insertion with Step Image Step Image 1 Select the icon on the right of the screen A new line is displayed ...

Страница 69: ... follows 6 6 1 5 Patient master data modification To modify a patient s master data proceed as follows Step Image 1 Enter the surname or ID in the specific field in the right section Step Image 1 Search for and select the desired patient 2 Press to change the selected patient s master data This option is not available for patients in the Worklist section ...

Страница 70: ...tem is configured to dialogue with the centralized master data management system the patient list relative to the date room surgeon can be called up by using the button To access the worklist proceed as follows Step Image 1 Press the button 2 Wait for the names relative to the criteria selected to appear in the list ...

Страница 71: ...alth Organization as default concerning each stage of the surgical procedure Press the icon to scroll through the questions The instructions refer to the entire procedure it will therefore be possible to exit this section and return to it to validate the instructions relating to the current state of the operating activity It will always be possible to view the fill in status of the Check List via ...

Страница 72: ...reset On the side menu press the icon to access the Preset screens The Preset enable to save room settings configurations It will then be possible to recall the configurations by pressing the relative icons The main screen is divided as follows A A Preset List ...

Страница 73: ...ng Video Recording Channel Conference Channel Setting Audio Operating Light Setting 2 Press the button The configuration window opens 3 Enter the name to be assigned to the new Preset and add a description in the appropriate field below optional 4 Select deselect the settings to be included in the configuration by pressing the relative icon among those listed ...

Страница 74: ...d as follows Press the icon to automatically activate the Preset each time the system is started 5 Press to confirm Step Image Step Image 1 Select the desired Preset from the Preset List The configuration window opens 2 Apply the Preset by pressing Press to perform changes to the Preset Save the changes by pressing ...

Страница 75: ...6 6 4 Multiview On the side menu press the icon to access the Multiview screen The Multiview function enables to combine multiple Inputs up to a maximum of 4 into a single Output signal The main screen is divided as follows A B A Source List B Multiview ...

Страница 76: ...re and Picture Quadview Picture Over Picture Picture In Picture 2 Drag the images one at a time from the Source List and drop them into the corresponding boxes Press which will appear at the top right of the box once selected to remove the image from the box 3 Once the mosaic has been filled in close the Multiview section by pressing The mosaic thus generated will be available in the Source List ...

Страница 77: ... On the side menu press the icon to access the Audio Control screen In the Audio Control section it is possible to set the volume levels of the inputs microphones and outputs speakers in the room The main screen is divided as follows A B A Microphone Inputs B Line Inputs ...

Страница 78: ...the event that the system receives a connection request in video conference the system will automatically silence lines AUX1 AUX2 and activate the video conference line This option is set by default if you wish to disable it contact the Assistance service VIDEOMED S r l Step Image 1 Operate on the volume bar of the microphones or line inputs according to the volume to be modified 2 Press along the...

Страница 79: ...80028000_030_C 773607 2021 09 01 79 6 6 5 2 Disabling microphones and audio To disable the microphones or line inputs proceed as follows Step Image 1 Press to disable the microphone 2 Press to turn off speaker volume ...

Страница 80: ...respectively to the 2 operating lights installed in the operating room The image shows the case where the lamp control system cannot be reached Helion Video Management System assures the option of controlling the Trumpf Medical surgical operating lights thanks to reproduction of the light control console through the graphic interface The graphic interface has been kept as similar to the original l...

Страница 81: ...tion can also be activated Light beam size Shadow setting the automatic function can also be activated the function may vary based on the light model installed 6 6 7 Environmental control panel management On the side menu press the icon to access the operating room control panel management screen The control panel management screen can be used only if Operamed devices are present in the operating ...

Страница 82: ... function Helion Video Management System includes a Lock function to lock the touchscreen by using a PIN To lock the touchscreen proceed as follows Step Image 1 Press to lock the screen 2 Enter the PIN code to unlock the touch screen using the highlighted numeric keypad ...

Страница 83: ...n be set as active by default at system startup or only after a Logout The Login screen requires 2 mandatory fields to fill in Username and Password in order to access the system Once the correct credentials have been entered the system will show the initial Helion graphical interface Video Routing interface Press the icon to Logout and be redirected to the credentials screen ...

Страница 84: ...f new products VIDEOMED S r l offers a waste collection and environmentally sustainable disposal service for all products VIDEOMED S r l Recycling and disposal is performed by VIDEOMED S r l at no additional cost to the operator To inform of the dispatch of decommissioned devices call 39 049 9819113 Assistance is always available to clear any doubts regarding the recycling and disposal of products...

Страница 85: ...e Source List Select the desired layout from PiP PaP PoP and Quadview Then proceed to populate this layout by dragging one at a time the images of the sources and releasing them in the relative boxes To remove an image from a frame press the which will appear at the top right of the frame once selected Once the composition of the mosaic has been completed return to the Video Routing section by pre...

Страница 86: ...ownload the patient list in automatic mode Urgent insertion With this option it is possible to create a study with a random ID name called Emergency Patient 8 6 Selection of signals to be recorded Drag the desired source into the Recording Channel box to enable basic recording functions Recording start stop Instant capture Press to proceed with closing the patient record and exporting the files Se...

Страница 87: ...This page is intentionally blank ...