Standard Features

Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)

 27 

Lockout

The Lockout control (padlock icon) disables the bed articulating function (for both 
patient and caregiver). The lockout of any bed articulating function will also lock out the 
Chair control and the Bed Flat control. The Tilt and Reverse Tilt controls and the 
transport pod Hi/Lo controls are locked out if the bed up/down control is locked out.

To activate

—at the same time, press the 

Lockout

 control and the 

Up

 or 

Down

 control 

of the applicable function. Both patient and caregiver controls are locked out. A single beep sounds and 
the applicable indicator light comes on to let you know that the function is locked out. 

NOTE:

If you press the Lockout control and do not press an Up or Down control within a few seconds, or if you 
do not complete the lockout procedure correctly, the bed will sound a triple beep to let you know that 
the function is not locked out.

To deactivate

—at the same time, press the 

Lockout

 control and the 

Up

 or 

Down

 control of the 

applicable function. Both patient and caregiver controls are unlocked. A single beep sounds and the 
applicable indicator light turns off to let you know that the function is no longer locked out.

NOTE:

If you attempt to use a locked-out control, a triple beep will sound to notify you to check the lockouts.

When the bed CPR is activated, any controls that are locked out will become unlocked.

Follow your facility’s protocols for lockouts to reduce the likelihood of unauthorized use of the bed 
controls.

Caregiver Controls on the Intermediate Siderail

Controls
A

Chair

B

Bed flat

C

Tilt

D

Reverse tilt

E

Width retract with indicator

F

Width extend with indicator

G

Foot longer

H

Foot shorter

I

Knee up/down with lockout indicator

J

Head up/down with lockout indicator

K

Bed up/down with lockout indicator

L

Lockout 

Indicators
M

Nurse  call 

N

Service required

O

Battery  charge  level

P

Bed  not  down

Q

Lockout indicators

Содержание Compella

Страница 1: ...Compella Bariatric Bed System Instructions for Use Product No P7800 178951 REV 6 ...

Страница 2: ......

Страница 3: ...the bed to the transport position Press and hold this control until the control s indicator is green and you hear a single beep C Steer Pedal indicator when this indicator is green the bed is in steer mode and can be moved D Battery Charge indicator Item Feature Page A Retract with the siderails raised press and hold this control to retract the width of the bed Press and hold the control until the...

Страница 4: ...o set the Bed Exit alert to notify you when the patient attempts to get out of the bed 51 G Bed Exit Out of Bed mode with the Enable key active press this control to set the Bed Exit alert to notify you when the patient has left the bed 48 or 51 H Alert Silence when a Bed Exit alert is armed press this control to stop monitoring the patient movement temporarily 30 seconds 49 or 51 I Alert Volume w...

Страница 5: ...ight side of the display to select Height or Weight 3 Press Increase or Decrease as applicable to enter the correct height or weight 4 The unit automatically adjusts the pressures for the set height and weight If necessary use the arrow keys to move the cursor to the applicable pressure setting head seat or foot and use the Increase and Decrease controls to adjust the settings To return all pressu...

Страница 6: ...iv Compella Bariatric Bed System Instructions for Use 178951 REV 6 NOTES ...

Страница 7: ... rental of its products Notice to Users and or Patients in EU Any serious incident that has occurred in relation to this device should be reported to the manufacturer and the competent authority of the Member State in which the user and or patient is established This manual 178951 was originally released and supplied in English For a list of available translations contact Hill Rom Technical Suppor...

Страница 8: ...ntelliDrive Line of Site Point of Care SideCom and SlideGuard are registered trademarks of Hill Rom Services Inc Replace this manual 178951 if it is damaged and or can not be read For product support or to order additional copies of this manual 178951 contact your distributor local Hill Rom representative or go to www hillrom com Reference Documents Compella Bariatric Bed System Service Manual 178...

Страница 9: ...Therapy Support Surface 1 Introduction 1 Symbols 2 Product Symbols 2 Acronyms 15 Safety Information 15 Quick View List of Bed Features 18 Information Indicators 19 Audible Indicators 19 Brake Not Set 19 Visual Indicators 19 Battery Charge Indicator for Bed Frame Articulations 19 Service Required 20 Bed Not Down 20 Hip Position Locator earlier bed versions 20 Head Pivot Locator 20 Line of Site Head...

Страница 10: ...wn 26 Standard Caregiver Controls 26 Caregiver Controls on the Intermediate Siderail 27 Lockout 27 Bed Up Down 28 Head Up Down 28 Knee Up Down 28 Tilt and Reverse Tilt 29 Bed Flat 29 Cardiac Chair 29 Battery Back Up 29 Brake and Steer Controls 30 To Activate 31 Patient Restraint and Drainage Bag Holders 31 Patient Restraints 31 Drainage Bag Holders 32 Equipment Sockets 32 Night Light 32 Head Angle...

Страница 11: ...the Scale 44 Weigh the Patient 44 Changing the Scale Units 45 Scale B Display 45 Bed Setup 45 Scale Reference Position 46 To Zero the Scale 47 Weigh the Patient 47 Magnification Mag Mode 48 Unstable Weight 48 Single Mode Bed Exit Alert System 48 To Activate the Bed Exit Alert System 49 To Silence the Bed Exit Alert System without Deactivating the System 49 To Deactivate the Bed Exit Alert System 5...

Страница 12: ...rol Panel 61 Turn Assist Mode 62 Deflate and Inflate the Seat Section for Patient Exit or Entry 62 Continuous Lateral Rotation Therapy CLRT Mode 63 Deflate and Inflate the Side Bolsters and Foot Section 65 Automatically using the Caregiver Width Adjust controls 65 Manually using the Air Supply Unit 65 Deflate the Side Bolsters and Foot Section for Patient Transport 66 Compella Foam Surface with In...

Страница 13: ...P7801 75 To Remove 75 To install 75 Line Managers P7512 76 Accessories for International 76 Oxygen Tank Holders AC959 AD101 and AD102 76 Pivoting 3 Liter Cylinder holder AC962 76 Telescopic IV Poles AD165 and AD148 77 To Use the IV Pole with Four Hooks AD165 77 To Use the IV Pole with Two Hooks AD148 77 Syringe Driver Holder AC963 78 To Adjust the Holder Position 78 Cleaning and Disinfecting 78 Re...

Страница 14: ...85 Troubleshooting 86 Solve a System Alarm Condition on a Compella Therapy Surface 86 Solve a System Alarm Condition on a Compella Foam Surface 87 Surface and Air Supply Unit Power Failure and Alarm Conditions 87 Power Failure 87 Alarm Alert System 87 Notification Priority 88 Powered Width Extensions will Not Extend or Retract 89 The Head section Will not Rise or lower 89 The Bed CPR Handle Does N...

Страница 15: ...be used to assist in treatment or prevention of pulmonary or other complications associated with immobility to assist in treatment or prevention of pressure injuries or for any other use where medical benefits may be derived from Continuous Lateral Rotation Therapy CONTRAINDICATION Contraindication Use of powered air surfaces for patients with unstabilized spinal cord injury could cause serious in...

Страница 16: ...E sets apart special data or important instruction clarification CONTRAINDICATION WARNING or CAUTION A CONTRAINDICATION identifies situations or actions that may have an effect on the patient s safety A WARNING identifies situations or actions that may have an effect on patient or user safety To ignore a warning could cause patient or user injury A CAUTION identifies special procedures or precauti...

Страница 17: ...ows the direction to turn the CPR mechanism to deflate the surface page 21 CPR control label for the bed frame shows how to operate the CPR control to lower the head section page 21 Caregiver Control Panel Nurse Call control page 53 Service Required indicator page 20 Bed battery charge status page 19 Chair position control page 29 Bed flat control page 29 Tilt control page 29 Reverse tilt control ...

Страница 18: ...h extend control and indicator page 34 Err 6 Bed width alert Fully expand or retract the bed width page 34 FlexAfoot bed length adjustment page 35 Bed Up and Down controls page 28 Bed Not Down indicator comes on when the upper frame is not in the lowest position page 28 Head Up and Down controls page 28 Symbol Description ...

Страница 19: ...ulation controls page 27 Caregiver Control Pod Scale Weigh control page 44 and page 47 Zero control page 44 Zero control Scale B page 47 Magnification Mode control Scale B page 48 Frame Setup control and indicator Scale B page 46 Raise Bed control and indicator Scale B page 46 Scale Reference indicator Scale B page 46 Symbol Description ...

Страница 20: ...ntrol and indicator page 43 Alert Silence control and indicator page 49 and page 51 Enable key page 43 page 48 and page 51 Bed Exit Alert System Patient Position mode page 51 Bed Exit Alert System Bed Exiting mode page 51 Bed Exit Alert System Out of Bed mode page 48 and page 51 Bed Exit Alert System Volume and Tone Control page 50 and page 52 Symbol Description ...

Страница 21: ...Hip Position Locator on earlier versions of the bed page 20 Head Pivot Locator page 20 Tilt Reverse Tilt angle page 21 Head angle page 21 Patient Controls Nurse Call control page 53 Head Up and Down controls page 28 Knee Up and Down controls page 28 Nurse Call Voice indicator page 53 Symbol Description ...

Страница 22: ...ed so the drive wheels raise off the ground page 36 Transport Position control and indicator lowers the bed to the transport position the drive wheels are on the ground page 36 Steer pedal indicator page 36 Bed Frame Shows how to raise the push handles page 36 or 41 Shows how to stow the push handles page 36 or 41 Before Transport sequence page 36 After Transport sequence page 36 Identifies the ma...

Страница 23: ...entifies patient restraint location ankle page 31 Head end brake pedal page 30 Head end steer pedal page 30 Do not use the IV pole in this location Do not store cords here Crush warning must consult accompanying documents Identifies battery installation location Identifies mains fuse Refer to the bed manufacturer s user manual for compatible support surfaces page 54 Symbol Description ...

Страница 24: ... cord International No Equipment Zone Safe Working Load for the bed this includes the weight of the patient support surface and accessories that are on the bed Patient minimum and maximum weight range Patient minimum and maximum weight range Total bed weight including the safe working load is 950 kg 2094 lb the bed weight excluding the safe working load is 450 kg 992 lb minimum Support Surface Sho...

Страница 25: ...tric Bed System Instructions for Use 178951 REV 6 11 Hand wash only Bleach with specification Bleach as needed Do not dry clean Do not tumble dry Tumble dry no heat Drip dry Machine wash Safe working load Symbol Description ...

Страница 26: ...it only Conforms to European Medical Device Directive The CE mark was first applied in 2018 Compella Bariatric Bed frame Therapy Surface and Air supply unit Conforms to European Medical Device Directive 93 42 EEC The CE mark was first applied in 2015 Conforms to European Medical Device Directive 93 42 EEC NAWI EN45501scale XX identifies the date of manufacture The CE mark was first applied in 2015...

Страница 27: ...Underwriters Laboratories Inc with respect to Electric Shock Fire Mechanical and other specified Hazards only in accordance with ES60601 1 IEC EN60601 1 and CAN CSA C22 2 No 60601 1 Type B applied part according to IEC 60601 1 Support surface only Type BF applied part Bed and patient pendant only According to IEC 60529 Rating for protection against fluid ingress and identified as equipment that is...

Страница 28: ...1 REV 6 Do not dispose as Unsorted Municipal Waste WARNING yellow and black CAUTION white and black Must consult the accompanying documents Alternating current Dangerous voltage Equipment emits electromagnetic energy Protective Earth Equipotentiality Symbol Description ...

Страница 29: ...g to established clinical assessment protocols Warning A thorough risk assessment and protocol is necessary to determine the appropriate surface for the patient s condition Warning Do not use the bed with patients who weigh more than 454 kg 1000 lb or are wider than the fully extended surface width Injury or damage could occur Warning Air pressure in the support surface is controlled automatically...

Страница 30: ...s dropped or receives a sudden impact equipment damage could occur Warning After exposure to extreme high or low temperatures let the air supply unit equilibrate for at least one hour before use Warning The air supply unit circulates room air when it operates Exposure to smoke may cause the unit to fail Do not permit patients or visitors to smoke cigarettes or other substances while using this dev...

Страница 31: ...d and found to be safe to operate Warning If the battery backup does not operate correctly the bed does not articulate when you press an articulation control plug the bed into AC power so that you can use the bed controls if necessary Warning Do not connect the power cord to an extension cord or multiple outlet strip There is a risk of overheating and fire may cause injury or damage Warning Whenev...

Страница 32: ...to cause birth defects or other reproductive harm For more information go to www P65Warnings ca gov QUICK VIEW LIST OF BED FEATURES The Compella Bariatric Bed System has these features CPR controls at patient right foot Powered width adjustment with manual override Integrated scale Item Feature Page Item Feature Page A CPR frame and surface controls 21 E IntelliDrive XL Transport System transport ...

Страница 33: ...ke Not Set The Brake Not Set is an audible alert only When the bed is plugged into AC power and you release the brakes the alert will sound To silence the alert either unplug the bed for transport or set the brakes WARNING Warning Unless you are to transport a patient always set the brakes when the bed is occupied Make sure the brakes are set before any patient transfer on to or off the bed Failur...

Страница 34: ... air surface will stay inflated but it will not adjust pressures The Bed Exit and Scale functions will not operate Service Required Hip Position Locator earlier bed versions The hip position label on the intermediate siderails identifies the correct position of the patient s hips while on the bed Correct placement of the patient increases the effectiveness of the SlideGuard Technology This is desi...

Страница 35: ...der the bed Refer to this label to make it easier to put the legs of the equipment under the bed STANDARD FEATURES CPR CONTROLS NOTE CPR is a two step process if the bed is equipped with an air surface There is a CPR label on the patient right hand side of the footboard that shows the locations of the CPR controls for the air surface and the bed frame NOTE Earlier versions of the bed may have a di...

Страница 36: ...ng released The surface will start to deflate NOTES When the air surface is completely deflated the bed s sleep deck can be used as a backboard The headboard and or patient helper can be removed for patient access for patient intubation or to insert a central line To Inflate the Air Surface after CPR 1 Turn the air surface CPR mechanism counter clockwise until it locks into position 2 On the air s...

Страница 37: ...g Evaluate patients for entrapment and fall risk according to facility protocol and monitor patients appropriately Make sure that all siderails are fully latched when in the raised position Failure to do either of these could result in serious injury or death NOTE Siderails are intended to be a reminder to the patient of the bed s edges not a patient restraining device When appropriate Hill Rom re...

Страница 38: ...ails give an estimated degree of bed tilt To Lower a Siderail Lift up on the recessed blue release handle that is on the lower part of the main siderail support The siderail has a dampening mechanism that slowly lowers the siderail NOTE Gently leaning into the siderail may make it easier to latch and unlatch in some situations For example this may be helpful immediately after you have fully retrac...

Страница 39: ...ng bed articulation a static buildup may occur Standard Patient Controls The patient controls are on the patient side of the intermediate siderails The standard patient controls include Head Up Down and Knee Up Down NOTE The patient head up down control includes the Auto Contour feature When the patient raises or lowers the head section the head and knee sections raise or lower at the same time to...

Страница 40: ...outside of the head and intermediate siderails There are three sets of controls Instruct visitors not to operate the caregiver controls They may assist the patient with the patient controls 1 The bed position control panel on the head siderail includes bed up down head up down and knee up down 2 The bed position control panel on the intermediate siderail includes bed up down head up down and knee ...

Страница 41: ... sound a triple beep to let you know that the function is not locked out To deactivate at the same time press the Lockout control and the Up or Down control of the applicable function Both patient and caregiver controls are unlocked A single beep sounds and the applicable indicator light turns off to let you know that the function is no longer locked out NOTE If you attempt to use a locked out con...

Страница 42: ...west position NOTE The Bed Not Down indicator comes on when the bed is not at the lowest bed height Head Up Down The caregiver can adjust the head section to specific angles The maximum travel for the head section is 50 To raise press and hold the Head Up control until the bed is at the desired height To lower press and hold the Head Down control until the bed is at the desired height NOTE The Aut...

Страница 43: ...the system stops its articulations When all sections are flat and the bed is level the system stops and sounds one beep NOTE If any frame function is locked the Bed Flat control will not operate Cardiac Chair CAUTION Caution Do not transport a patient with the bed in the Chair position Equipment damage could occur The caregiver can adjust the bed to a cardiac chair position When you press the chai...

Страница 44: ... bed into an AC power outlet whenever possible Brake and Steer Controls WARNING To help prevent injury and or equipment damage obey these warnings Warning Unless you are to transport a patient always set the brakes when the bed is occupied Make sure the brakes are set before any patient transfer on to or off the bed Warning Make sure your foot is firmly on the pedal during brake and steer operatio...

Страница 45: ...ient Restraint and Drainage Bag Holders Patient Restraints WARNING To help prevent personal injury and or equipment damage obey these warnings Warning Make sure restraints are attached to the articulating sections of the bed at the correct attachment points Warning Do not adjust the width of the bed while restraints are in use The bed facilitates the use of ankle waist wrist and chest restraints H...

Страница 46: ... Sockets There are four equipment sockets for the attachment of accessories They are at each corner of the bed The equipment sockets can be used to mount IV poles and oxygen tank holders NOTE Everything on or attached to the bed will be included in the scale reading This includes IV poles and items attached to the poles pumps and drainage bags the patient helper trapeze and items attached to the h...

Страница 47: ... width is extended that may interfere with or contact the bed when lowering Warning Keep the patient within the surface perimeter during width adjustments The Head Angle Alert control is on the caregiver control pod next to the display When set if the head section goes below 30 these will occur The display will flash five times An audible alert will come on The alert indicator will flash Set the A...

Страница 48: ...und NOTE If the bed is not fully retracted or extended you will not be able to raise the head of the bed you will only be able to lower it Manual Width Control The bed width can be adjusted manually when the bed is not connected to AC power such as during a transport Each siderail has its own blue release lever to use to manually adjust the bed width To manually retract the bed do these steps in t...

Страница 49: ...Both indicators next to the Retract and Extend Width controls will flash if the bed is not in the correct position and will continue to flash until the bed is fully extended or fully retracted A triple beep will sound every 10 seconds until the bed is adjusted to the fully extended or fully retracted position Err 6 will flash on the caregiver control pod until the bed is fully extended or fully re...

Страница 50: ...hanism deploys or stows based on the bed height The system is activated when the bed brake is set to steer the bed is disconnected from AC power the bed is in Transport mode and the caregiver presses an enable switch and applies pressure to the push handles that are at the head end of the bed This allows the caregiver to propel the bed during patient transport with minimally applied force To put t...

Страница 51: ...occur The powered transport system is intended for indoor use only Outdoor use may cause temporary or permanent damage to the powered drive mechanism Before you transport the bed make sure that the power cord hoses and other equipment are correctly stowed For transport a fully charged bed battery is preferred however if the bed battery charge is low put the bed in the correct position before any t...

Страница 52: ...trol on page 34 to adjust the width of the bed if AC power is not available It takes approximately 2 minutes for the side bolsters to deflate There must be power to the bed and surface during this time 4 When the side bolsters have deflated turn the air supply unit off NOTE The air surface will stay inflated during transport To help with lateral transfers Max Inflate the surface before you disconn...

Страница 53: ...1 Access the transport pod at the patient head right corner of the bed 2 Adjust the transport pod into a position for easy access 3 Press the green Transport control 1 to lower the bed to the transport position the power drive wheels on the ground When the bed reaches the Transport position you will hear a single beep and the amber Transport indicator on the transport pod will turn green NOTES If ...

Страница 54: ...sition the brake pedal is in steer and the battery indicator has a minimum of one 1 bar If you need to align the bed into position with fine small side to side movements raise the bed so that the drive wheels are no longer on the ground Once the bed is in the correct position set the brakes Disengage Transport Mode After the transport do as follows 1 Press the Disengage control at the top of the t...

Страница 55: ...lengthen or shorten the foot end of the bed make sure that the magnets on the surface power cord do not interfere with the movement Warning Do not push or pull the bed by IV poles siderails or other equipment Use the push handles headboard or footboard There are push handles at the head end of the bed that can be used during transport These handles can be stowed when they are not in use Transport ...

Страница 56: ... unit and its hoses over the footboard and set the unit and hoses on the foot or side section of the bed 11 At the head end of the bed unplug the power cords for both the bed and the air supply unit Use the blue cord hooks at the head section to store cords during transport 12 Put the brake steer pedals in the Steer or Neutral position as necessary 13 Transport the patient After the transport do a...

Страница 57: ...apply to your scale system NOTE Everything on or attached to the bed will be included in the scale reading This includes items on the headboard footboard siderails IV poles and drainage bag restraint holders SCALE A DISPLAY The scale system continuously weighs the patient however the weight does not continuously show on the display You must press the Weigh control to view the patient s weight NOTE...

Страница 58: ... not touching anything that could affect the patient weight other beds or carts headwall miscellaneous drainage lines etc The scale system is now ready for you to zero the scale or weigh the patient To Zero the Scale 1 Press the Enable key 2 Press and hold the Zero control until 00 0 shows on the display HOLD will show until 00 0 shows 3 Release the control when 00 0 shows 4 After you release the ...

Страница 59: ...he bed is plugged into AC power 2 Put all standard linens blankets and pillows on the bed A list of these items posted near the bed may be helpful for future reference 3 Make sure the bed is not touching anything that could affect the patient weight other beds or carts headwall miscellaneous drainage lines etc 1 Make sure the Enable key indicator is off 2 Press and hold the Zero control After appr...

Страница 60: ...etup or Release Brake indicator Repeat step 1 or 2 as identified by the flashing indicator NOTE If the Brake Not Set alert is sounding you will not hear the triple beep so we recommend that you release the brake last The scale system is now ready for you to zero the scale or weigh the patient 1 Press and hold the Frame Setup control until you hear a single beep These will occur a The head section ...

Страница 61: ...Not Set alert will sound and the Scale Reference indicator will come on solid 3 Press and hold the 0 T control Hold will show on the display 4 Release the control when 00 0 shows 5 After you release the control the display will flash CALC Do not touch the bed until the display stops flashing CALC When CALC stops flashing you will see 0 0 on the display the 0 T control indicator will be green and y...

Страница 62: ...giver control pod that is on each intermediate siderail If your bed is equipped with the Compella Low Air Loss Surface with Turn Assist and CLRT Continuous Lateral Rotation Therapy Bed Exit should be used in the Normal mode only It should not be used with the CLRT mode active as it may cause false alerts NOTE If the bed is in Transport Mode the power drive wheels are on the ground and you attempt ...

Страница 63: ...ystem is in the Silence Mode you can change the position of the patient or assist the patient out of the bed Alert Silence can be used during Turn Assist To silence the alert system before it sounds press the Enable key until its indicator is on solid and then press the Alert Silence control until its indicator is on solid To silence the alert system after it sounds press the Alert Silence control...

Страница 64: ...ctivate a Bed Exit mode a minimum of 113 kg 250 lb must be on the bed 2 Activate the alert by having the caregiver exit the bed 3 Press and hold the Volume control 4 While you press the Volume control press the Out of Bed control 5 Press and release the Out of Bed control until the desired tone is reached 6 Clear the alert condition Press the Enable key until its indicator is on solid and then pre...

Страница 65: ...tion Bed Exiting and Out of Bed Patient Position Mode this mode sounds an alert when the patient moves towards either siderail moves away from the head section or sits up in bed This mode should be used when a caregiver wants to be notified when the patient begins to move Bed Exiting Mode this mode sounds an alert when a patient moves away from the center of the bed towards an egress point This mo...

Страница 66: ...ible alert or send a nurse call alert While the system is in the Silence Mode you can change the position of the patient or assist the patient out of the bed Alert Silence can be used during Turn Assist To silence the alert system before it sounds press the Enable key until its indicator is on solid and then press the Alert Silence control until its indicator is on solid To silence the alert syste...

Страница 67: ...e on the bed 2 Activate the alert by having the caregiver exit the bed 3 Press and hold the Volume control 4 While you press the Volume control press the Out of Bed control 5 Press and release the Out of Bed control until the desired tone is reached 6 Clear the alert condition SIDECOM COMMUNICATION SYSTEM NURSE CALL CONTROL A Nurse Call control is on the caregiver and patient control panels and th...

Страница 68: ...is speaking NOTE The Nurse Call controls are always active and their indicators are green when a Nurse Call has not been initiated The Nurse Call controls can not be locked out SURFACE FEATURES AND CONTROLS SAFETY INFORMATION WARNING Obey all warnings and cautions throughout the manual and also the safety information below to help prevent injury and or equipment damage General Warning Use of the t...

Страница 69: ...urface products that have not been designed by Hill Rom for the Compella Bariatric Bed System Use of surface products other than those designed for the Compella Bariatric Bed System could substantially reduce the effectiveness of the safety features incorporated into the system Air Supply Unit Warning There is a risk of asphyxiation due to entanglement with cords Route the power cord under the bed...

Страница 70: ...ied as part of the device or specified as being compatible with the device Warning Check that the power ratings of the supply mains are sufficient to power the air supply unit Warning Use of accessories transducers and cables other than those specified with the exception of transducers and cables sold by the manufacturer of the device as replacement parts for internal components may result in incr...

Страница 71: ...the air supply unit on the footboard make sure that the mounting bracket is installed on the footboard correctly Otherwise equipment damage could occur if the air supply unit becomes dislodged 5 Hang the air supply unit on the mounting bracket on the footboard CAUTION Caution Make sure that the surface being connected is correctly matched to the air supply unit to make sure that all functions oper...

Страница 72: ...t corner of the bed Along the side of the intermediate frame on the patient left hand side of the bed At the patient head left corner of the bed 11 Plug the air supply unit power cord into the AC power outlet Do not put the air supply unit in a position that makes it difficult to unplug the power cord NOTE The Power switch on the caregiver s right side of the air supply unit turns the unit On and ...

Страница 73: ...to and from the Patient Setup screen Help shows the Help screen Arrows selects settings on the Patient Setup screen COMPELLA THERAPY SURFACE The Compella Therapy Surface is a therapeutic system that supplies three zone constant pressure redistribution microclimate management Turn Assist and Continuous Lateral Rotation Therapy CLRT The Compella Bariatric Therapy Surface consists of either Therapy S...

Страница 74: ...ssure Redistribution screen NOTE The settings will be saved when the system returns to the Pressure Redistribution screen Surface Options Max Inflate Mode Max Inflate mode inflates the surface assembly to its maximum pressures NOTE The surface goes into Hyper Inflate mode to inflate the surface after Seat Deflate and Bed Deflate modes it does not go into Max Inflate mode To turn Max Inflate mode o...

Страница 75: ...e Pressure Redistribution screen press Mattress Adjust The Mattress Adjustments screen shows 2 Press Fowler Adjust The Fowler Boost screen shows 3 Press On Off to turn Fowler Boost mode On or Off 4 If necessary adjust the setting as follows To increase the pressure press Increase To decrease the pressure press Decrease Press Default to return to the standard Fowler Boost setting of 30 more than th...

Страница 76: ...elected turn starts and a 30 minute countdown screen shows with two menu options Pause Turn this option stops the turn at the current angle Pause will then change to Resume which permits the turn to continue until completion as indicated by one long beep Exit Turn this option brings the patient back to a flat position NOTE If you retract or extend the bed width during the Turn Assist mode the Paus...

Страница 77: ...screen flashes for 5 seconds 2 The unit goes into Hyper Inflate mode to inflate the surface to 55 mmHg for 10 minutes If the unit is in Pressure Redistribution mode the Hyper Inflate screen shows and highlights Hyper Inflate Continuous Lateral Rotation Therapy CLRT Mode The rotation mode provides gentle side to side continuous lateral rotation therapy CLRT to help prevent and treat pulmonary compl...

Страница 78: ...To do so could cause patient injury 5 At the Siderail warning screen press Yes or No 6 The CLRT Mode screen shows 7 CLRT begins During therapy you have the option to Exit CLRT Pause or Max Inflate NOTE If you retract or extend the bed width during CLRT the Pause function will activate to permit the bolsters to deflate or inflate 8 If Pause is selected the patient will stop rotating and stay at the...

Страница 79: ...e bolsters will deflate for 2 minutes anytime some or all of the frame width extensions are partially or fully retracted Foot Section The foot section will automatically inflate or deflate sufficiently to accommodate any adjustment to the foot extension of the frame Manually using the Air Supply Unit The side bolsters and foot section can be operated manually only when the frame is disconnected fr...

Страница 80: ...t The foot section deflates the Foot Deflating screen shows for 2 minutes and then the Pressure Redistribution screen shows To cancel Foot Deflate prior to completion press Inflate and Exit The foot section inflates the Foot Inflating screen flashes for 5 seconds and then the Pressure Redistribution screen shows After the foot section has been deflated the Deflated Foot indicator shows on the Pres...

Страница 81: ...s Audio Pause To inflate the side bolsters and foot section press Inflate and Exit The Surface Inflating screen flashes the side bolsters and foot section inflate and the unit goes into Pressure Redistribution mode in 5 seconds COMPELLA FOAM SURFACE WITH INFLATABLE SIDE BOLSTERS The foam surface with inflatable side bolsters and foot section is for patient weights up to 454 kg 1000 lb The width of...

Страница 82: ...down starts and the side bolsters deflate 2 If the side bolsters remain deflated for 30 minutes an alarm sounds To mute the alarm for 10 minutes press Audio Pause To inflate the side bolsters on the foam surface do as follows At the Deflate Sides screen press Inflate and Exit The Sides Inflating screen shows the side bolsters inflate and then the Foam Mattress screen shows Foot Section To deflate ...

Страница 83: ...reen shows NOTE If you press No the display shows the Foam Mattress screen 3 When the side bolsters and foot section are deflated the Ready for Transport screen shows 4 If the surface is deflated for more than 30 minutes the alarm sounds To mute the alarm for 10 minutes press Audio Pause To inflate the side bolsters and foot section do as follows Press Inflate and Exit The Surface Inflating screen...

Страница 84: ...America P2217 IV Pole P27601 Oxygen tank holder vertical P7802 Patient Helper P7803A01 Pendant P7803A02 Pendant with Nurse Call P3670A01 Respiratory circuit holder P3670A05 Pressure transducer holder P158 a a Adapter bracket P163 must be installed for use of P158 Infusion Support System ISS P163 ISS socket adapter P7801 Headboard P7512 Line Managers Accessories for International AC959 Oxygen tank ...

Страница 85: ...y into any of the equipment sockets at either the head end or foot end of the bed To Remove Pull the IV pole out from the equipment socket Oxygen Tank Holder Vertical P27601 WARNING Warning Failure to correctly attach the oxygen tank holder could cause it to drop Injury and or equipment damage could occur The oxygen tank holder holds one E size oxygen tank with a regulator The mount location lets ...

Страница 86: ...t may fall Warning Do not remove or install the Patient Helper arm assembly while it is in a position over a patient For use the Patient Helper trapeze must be locked in the center position The complete Patient Helper weighs 21 kg 46 3 lb Use correct lifting techniques and or ask for assistance when you install or remove the Patient Helper The Patient Helper can be used to help assist patients wit...

Страница 87: ...le locks into position To Remove 1 Pull theT handle that is on the Patient Helper mount and turn the arm assembly so that the trapeze handle is over the side of the bed 2 Loosen the clamp that attaches the trapeze handle assembly to the horizontal arm and remove the trapeze handle assembly from the horizontal arm 3 Remove the pull pin and horizontal arm from the arm assembly 4 Remove the arm assem...

Страница 88: ...tly aligned and that there is sufficient slack in lines for patient movement and surface rotation Warning Do not attach height sensitive devices or transducers with drainage capacity Warning Do not exceed the 2 2 lb 1 kg load capacity of each holder Warning Do not use a holder to move the bed The pressure transducer holder and respiratory circuit holder are used to hold respiratory ventilator circ...

Страница 89: ...ical orientation and raises or lowers the pumps or bags with respect to the system frame The head end of the system has attaching points for two mobile ISS poles Each pole can support one infusion pump plus two liters of intravenous solution Before you install the ISS pole P158A it is necessary to install the P163 ISS socket adapter Headboard P7801 The headboard is available with or without a user...

Страница 90: ...dations Make sure that the oxygen tank holder is correctly inserted into one of the equipment sockets at any of the four corners of the bed Make sure that the oxygen tank is correctly inserted in the holder Do not use a different oxygen tank model from the model specified The tank could fall or interfere with other operations Make sure that the holder is in a safe position before you adjust the ti...

Страница 91: ...th Four Hooks AD165 1 Insert the IV pole vertically into any of the equipment sockets at either the head end or foot end of the bed 2 To adjust the height of the pole do these a Loosen the knob on the pole as you hold the lower section of the pole b Hold the upper section of the pole just below the plastic sleeve c Push the sleeve upwards and adjust the pole to applicable height d Tighten the knob...

Страница 92: ...The potential for electrical shock exists with electrical equipment Failure to follow facility protocol could cause death or serious injury Warning Do not reuse wiping material for multiple steps or on multiple products Warning Harmful cleaning solutions may cause skin rash and or irritation upon contact Follow the manufacturer s instructions found on the product label and Safety Data Sheet SDS Wa...

Страница 93: ...g and disinfection process RECOMMENDATIONS For proper cleaning and disinfection staff members should be trained The trainer should carefully read the instructions and follow them when the trainee is being trained The trainee should Be given time to read the instructions and to ask any questions Clean and disinfect the product while the trainer supervises During and or after this process the traine...

Страница 94: ... Air Supply Unit on page 57 and do the installation steps in reverse order STEP 1 Cleaning a As necessary first remove visible soil from the bed and the support surface using a wiping cloth soaked with an approved cleaner disinfectant see Table 1 Approved Cleaners Disinfectants on page 80 Give special attention to seams and other areas where soil may accumulate A soft bristle brush may be used to ...

Страница 95: ...nector assembly and the support surface sleeve area shown in the picture below Do not disconnect the connector NOTE Clean the folds of the support surface sleeve c Examine the following for damage Top support surface cover Bottom support surface cover and attachment straps Zipper closure d The damaged items should be replaced STEP 2 Disinfection a With a new or clean wiping cloth soaked in an appr...

Страница 96: ...Cleaning and Disinfecting on page 78 For rental beds Hill Rom Service personnel will follow the Laundry Wash and Dry Procedure QS02040 CLEAN THE AIR SUPPLY UNIT FILTERS For correct operation of the Compella Therapy Surface keep the filters clean After 800 hours of operation when the air supply unit is turned on the Replace Filter screen shows as a reminder to replace the filters To replace the fil...

Страница 97: ...ntenance schedule refer to the Compella Bariatric Bed System Service Manual 178952 For service and or technical information other than that specified in this manual including fuse replacement circuit diagrams and isolation of mains power refer to the Compella Bariatric Bed System Service Manual 178952 Do annual preventive maintenance procedures to make sure the Compella Bariatric Bed System operat...

Страница 98: ...own power before the final battery charge indicator flashes The batteries will need to be replaced After the batteries are replaced charge them for a minimum of 20 hours before use AIR SUPPLY UNIT FILTERS REPLACEMENT WARNING Warning Failure to clean the filters may cause damage to the air supply unit which could prevent the unit from supplying the correct support surface pressures Patient injury c...

Страница 99: ...f the decommissioned bed or accessory is not fit for use Hill Rom recommends dismantling the bed in accordance with the instructions provided in the service manual Hill Rom recommends that all oil and hydraulic fluids are removed from the product before recycling or disposal if applicable Always check and comply with all local and national regulations and facility protocols when decommissioning a ...

Страница 100: ... the top row the head seat and foot show in the bottom row If no pressures settings are highlighted on the System Alarm screen examine the side bolsters If a pressure setting is highlighted do as follows Refer to the highlighted pressure on the System Alarm screen to determine which section and air cell row of the surface caused the alarm condition Refer to the table below to find which air cells ...

Страница 101: ...ugged into AC power to operate Power Failure A power failure can occur under these three conditions The air supply unit power cord was disconnected from the AC power outlet A power outage has occurred A fuse has blown During a power failure condition the display is Off and the air cells in the support surface will not inflate but will hold air When power is restored these occur The unit will resum...

Страница 102: ... Timeout alarm is only operational when the system is operated in manual mode Turn Failed medium priority alarm turns are permitted 5 minutes to complete If a maximum turn has not been achieved after 5 minutes the system will sound a Turn Failed alarm High Pressure Low Pressure medium priority alarm if one or more zones fail to maintain pressure for 15 consecutive minutes the system will sound a L...

Страница 103: ...ommunication board error If so the Communication Error message will show If you have pressed the Enter button and you need to see the error message or confirm that there is still an issue turn the air supply unit off and then turn it on If there is still an issue the Communication Error message will show If there is an internal communication board error the system will continue to operate in the M...

Страница 104: ... prevent the handle from fully returning You may need to contact your facility authorized maintenance person or Hill Rom Technical Support to do this inspection SERVICE CALLS WARNING Warning Only facility authorized persons or Hill Rom service technicians should service the bed Service done by unauthorized persons could cause injury and or equipment damage When you call Hill Rom about your unit be...

Страница 105: ...ilow in the low position 6 75 17 1 cm hilow in the high position Hi Lo b Maximum Minimum Transport b Measured from the top of the seat section outer edge to the floor 26 8 68 cm 18 5 47 cm 17 43 cm Patient weight range 113 kg to 454 kg 250 lb to 1000 lb Patient height range 50 to 78 127 cm to 198 cm Safe Working Load including patient support surface and accessories 500 kg 1100 lb Total bed weight...

Страница 106: ...face Surface width 40 to 50 102 cm to 127 cm Surface length inflated 80 to 88 203 cm to 224 cm Surface weight 17 kg 38 lb foam surface 23 6 kg 52 lb therapy surface Latex content User accessible parts surface pumps and accessories are not made with natural rubber latex Operating weight range 113 kg to 454 kg 250 lb to 1000 lb Operating height range 50 to 78 127 cm to 198 cm Air supply unit range o...

Страница 107: ...20 V AC Maximum Current Bed 12 A Air supply unit 1 2 A Frequency Bed 50 60 Hz Air supply unit 50 60 Hz Condition Range Rated Voltage Bed 220 230 240 V AC Air supply unit 220 230 240 V AC Maximum Current Bed 6 A Air supply unit 0 6 A Frequency Bed 50 60 Hz Air supply unit 50 60 Hz Fuse Type 120 V and 230 V models 3 15 A 250 V fast acting 1 There are no user accessible fuses for the bed Refer to the...

Страница 108: ... Degree of protection against the presence of flammable anesthetic mixtures Not for use with flammable anesthetics Mode of operation Continuous operation with intermittent loading 2 minutes ON 18 minutes OFF Sound level 52 dBA patient resting continuous operation measured from patient s perspective with siderails up 65 dBA patient alert continuous operation measured 1 m 39 4 from the bed centered ...

Страница 109: ...other electrical equipment If adjacent or stacked use is necessary observe the P7800 and the other electrical equipment to make sure they operate as intended Make sure the P7800 operates correctly when it is used near other electronic devices Portable and mobile radio frequency RF communications equipment can affect electrical equipment WARNING Warning Portable RF communications equipment includin...

Страница 110: ... Environment Guidance RF emissions CISPR 11 Group 1 The P7800 uses RF energy only for its internal functions Therefore its RF emissions are low and are not likely to cause any interference in nearby electronic equipment RF Emissions CISPR 11 Class A The P7800 is suitable for use in all establishments other than domestic establishments and those directly connected to the public low voltage power su...

Страница 111: ...s to Line s 2 kV Line s to Earth 1 kV Line s to Line s 2 kV Line s to Earth Mains power quality should be that of a typical commercial or hospital environment Voltage Dips IEC 61000 4 11 See Note 1 0 UT for 0 5 cycles at 0 45 90 135 180 225 270 and 315 for single phase mains equipment 0 UT for one cycle 70 UT for 25 50 Hz and 30 60 Hz cycles 0 UT for 0 5 cycles at 0 45 90 135 180 225 270 and 315 f...

Страница 112: ...d RF Immunity IEC 61000 4 3 3 V m 80 MHz to 2700 MHz 10 V m 80 MHz to 2700 MHz Interference may occur in the vicinity of equipment marked with this symbol Note 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Note 3 Field strengths from fixed transmitters such as base stations for radio cellul...

Страница 113: ...Test Level V m 385 380 390 TETRA 400 Pulse modulation 18 Hz 1 8 0 3 27 450 430 470 GMRS 460 FRS460 FM 5 kHz devi ation 1 kHz sine 2 0 3 28 710 704 787 LTE Band 13 17 Pulse modulation 217 Hz 0 2 0 3 9 745 780 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation 18 Hz 2 0 3 28 870 930 1720 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation 2...

Страница 114: ...tput power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 33 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation...

Страница 115: ......

Страница 116: ...i ...