Centuris™ Pro Bed

Instructions for Use

LI900B1

194568

Rev. 

6

Содержание Centuris Pro

Страница 1: ... 194568 EN Centuris Pro Bed Instructions for Use LI900B1 194568 Rev 6 ...

Страница 2: ...rks of Hill Rom Services Inc Hillrom Centuris Pro and AutoContour are trademarks of Hill Rom Services Inc MCM is a trademark of Hill Rom SARL Sabina Viking Golvo et LowBase are trademarks of Liko R D AB Hill Rom reserves the right to make changes to the design characteristics and models without prior notice The only warranty Hill Rom makes is the express written warranty extended on the sale or re...

Страница 3: ... 10 Precautions for transport and storage 11 Technical specifications 12 Overview 14 General Symbols 16 Function Symbols 17 Electrical controls 19 Installing the patient Before placing the patient on the bed 21 Accessories and peripheral equipments 21 Mattress 22 Recommended accessories 25 Recommended additional parts 25 Recommended patient lifts 26 Recommended bed dining tables 26 Endboards 26 In...

Страница 4: ...e AD294A 47 Telescopic IV pole AD298A AD299A 47 Linen holder 48 Drainage bag holder pins 49 Oxygen Cylinder Holder AC959A AD101A AD102A 49 Pivoting 3L Bottle Holder AC962A 50 Monitor stand AD244B 51 Syringe driver holder AC963A 51 IV line manager support AD286A 52 X ray transparent adjustable head section AD242A 53 Chrome plated IV hook AC953A 55 Label holder AC325A 55 Movement Transfer Braking st...

Страница 5: ...68 6 Instructions for Use Electric Bed Centuris Pro Page iii PROPRIETARY AND CONFIDENTIAL DRAFT 02 novembre 2020 Electromagnetic conformance 68 Complies with electromagnetic emission standards 68 Electromagnetic conformance 69 Complies with electromagnetic emission standards 69 Compliance with electromagnetic immunity 70 Recommended separation distances 72 ...

Страница 6: ...PROPRIETARY AND CONFIDENTIAL DRAFT 02 novembre 2020 Page iv 194568 6 Instructions for Use Electric Bed Centuris Pro ...

Страница 7: ...ons The structure of the Instructions for Use For every type of use Hillrom beds provide patients with optimal comfort and greater independence for a feeling of well being that is conducive to a swift recovery They are also easy to use for caregivers PATIENT CAREGIVER EASE OF USE INDEPENDENCE WELL BEING COMFORT SAFETY INSTALLATION MOBILIZATION HELP WITH CARE MOVEMENT ...

Страница 8: ... text emphasizes a word or phrase highlights special information or explains very important instructions The symbols below represent different risks or hazards Symbol Description Warning This symbol indicates that the failure to follow the associated recommendation can put the patient or the user in danger or damage the equipment Caution This symbol indicates that the failure to follow the associa...

Страница 9: ...o identify your bed model its serial number SN HRPXXXXXXXXX its UDI and its date of manufacture refer to the identification label see Electrical characteristics page 13 Your bed model LI900B1 is composed of a chassis sleep surface whose reference REF starts with CS900B1 and two endboards a headboard and a footboard REF CS900B1XXXXXX CS900 Centuris Pro B Version 1XXXXXX unique 7 figure numerical co...

Страница 10: ...ection against power outages The electric Trendelenburg is not an emergency function fitted with Ø 150mm casters can be used to transfer patients Intended Users The Centuris Pro beds are designed to be used by Qualified Staff Patients and Visitors can also use the Centuris Pro medical beds depending on authorization given by Qualified Staff First use Before using the bed it is essential to have a ...

Страница 11: ...arged before using the bed without the mains power supply make sure that all the moving parts are in good working order make sure that the bed has been cleaned and disinfected see Decontamination page 61 Risk prevention General recommendations In general check that nothing e g objects accessories or power cable or any persons e g children limbs will interfere with the movement of the mobile parts ...

Страница 12: ... from particular behavioral difficulties e g agitation mental confusion loss of sense of direction obsessive behavior old patients weakness etc properly trained medical staff should ascertain how the siderails should be used irrespective of the model or type whether the patient should be monitored closely or immobilized and whether the patient helpers should be left in position in order to ensure ...

Страница 13: ... by the Hospital Bed Safety Workgroup guide and the standard EN 60601 2 52 the mattress label on page 22 lists the mattresses recommended for use on the Centuris Pro to offer the safest conditions The therapeutic benefits of the other therapeutic mattresses listed in page 22 outweigh the residual risk of entrapment or fall incurred by their use Other mattresses may be used but the manufacturer mus...

Страница 14: ...the electrostatic discharges are prevalent we recommend using an antistatic caster The mains power supply for the bed must comply with relevant standards NF C 15 100 and NF C 15 211 France International Electrotechnical Commission IEC 364 for other locations Check that the bed s power requirements on the identification label see Electrical characteristics page 13 correspond to the power supply vol...

Страница 15: ...er clean or service the bed without unplugging it from the mains power supply and disconnecting the battery The battery backup must never be left in direct contact with fire placed in liquid or discarded in a refuse bin In the event of the battery being damaged see De commissioning page 66 This label indicates that the bed must never be used with an oxygen tent or in explosive atmospheres presence...

Страница 16: ...utside the corresponding ranges recommended by Hill Rom in hyperbaric chambers in explosive atmospheres in the presence of flammable gases or vapors with oxygen tent type respiration devices or devices that extend below the sleep surface outdoors or to transport a patient in a vehicle moving the bed over soft ground or inappropriate surfaces moving the bed along slopes of over 10 with or without a...

Страница 17: ... other During shipment a the bed must be a Transport does not include the transfer of the bed between wards with or without patients Climatic restrictions on transport and storage When stored the bed must be in the lowered position all functions locked out covered brakes applied and all moving parts secured protected from fluid ingress in the lowered position all functions locked out covered brake...

Страница 18: ... open L 2362 mm a Length of long siderail protection B 1421 mm b Height of long siderail protection without mattress S 385 mm ab Length of the head half siderail protection B1 499 mm a Length of the foot half siderail protection B2 631 mm a Height of long siderail protection without mattress S1 393 mm a Low position 150 d diameter casters cd h 397 mm a High position 150d diameter casters cd H 768 ...

Страница 19: ...clination in relation to sleep surface f SWL 220 kg the maximum patient weight varies according to the mattress and accessories 155 kg as per EN60601 2 52 acute care 185 kg as per EN60601 2 52 other environments g Without mattress or accessories Characteristic 120V 230V Voltage 120V AC 230V AC Frequency 50 60 Hz 50 60 Hz Power supply unit maximum power load 300 VA 300 VA Power supply unit fuse rat...

Страница 20: ...d model H Extension linen holder a B Headboard I Bumper 4 C 2 sockets for I V pole and patient helper J Central brake and steer bar control D Control pendant a K 150 mm diameter single band casters E Head section CPR control a L HRP and identification labels F General lock out unit of the electrical functions M Caregiver half siderails controls a G Footboard N Patient half siderails controls a D A...

Страница 21: ... Removable metal siderails a a Equipment varies depending on bed model G Head section angle indicator B Headboard H Electric function lock out unit C Control pendant I 150 mm diameter single band casters D 2 sockets for I V pole and patient helper J Footboard E HRP and identification labels K Bumper 4 F Head section CPR control L Central brake and steer bar control A B D I L K J C E F H G ...

Страница 22: ... manual Type B Equipment DO NOT BIN follow the local recycling regulations Direct Current Danger do not use Alternating Current Recyclable Material Earth leakage circuit breaker rating Total weight authorized during moving Body Mass Index 17 Patient weight 40 kg Patient height 146 cm Atmospheric pressure limits Hygrometry limits Temperature limits Maximum patient weight Protective earth Safe worki...

Страница 23: ... LITS compliant NF ENVIRONNEMENT compliant bed Compliant with the ROHS Europe regulations Compliant with the ROHS China regulations Unique Device Identifiant Medical Device Duty cycle Head section CPR Information page 45 Headboard position Information page 26 Siderail lock Information page 38 Siderail assembly lock Information page 38 Siderail release Information page 38 Electric functions lockout...

Страница 24: ... Pro Do not sit or climb on the linen holder Information page 47 Do not sit or climb on the extension Information page 27 References of recommended mattresses Information page 22 and page 22 Patient helper position Information page 33 Caster control Information page 57 Earth continuity and earth leakage current ...

Страница 25: ...vembre 02 2020 Introduction Specifications Electrical controls 1 1 1 Functions available only to the caregiver General lock out unit Control pendant Raise AutoContour Lower AutoContour Reverse Trendelenburg 1 Lock control 1 Trendelenburg 1 Raise thigh section Raise HiLow Lower HiLow Lower thigh section ...

Страница 26: ...TIAL DRAFT novembre 02 2020 Page 20 194568 6 Instructions for Use Electric Bed Centuris Pro 1 1 Functions available only to the caregiver Commande de verrouillage 1 Caregiver half siderail control Patient half siderail control ...

Страница 27: ...falls of the patient patient in state of confusion patient s learning ability persons lacking the mental capacity to recognize unsafe actions unauthorized persons check the list of recommended mattresses on the label on the adjustable head section All persons authorized to use the bed s functions must be capable of doing so in a safe and controlled manner In case of doubt the bed s functions must ...

Страница 28: ...Rom recommends the mattresses listed below which are compatible with the safety recommendations see Risk prevention page 5 Mattress label for bed with half siderails Mattress label for bed with long siderails AD271B Folding mattress clamp When installing a mattress extension cushion the clamp must be folded to avoid any contact with the lower limbs 1 2 3 ...

Страница 29: ...mps on the foot section for beds with half siderails or the fixed clamps on the head and foot sections for beds with long siderails Other mattresses may be used but the manufacturer must always be consulted to make sure that the bed mattress siderail combination does not affect the bed s performance its suitability for use or its safety characteristics Users must check the compatibility of the pat...

Страница 30: ... density foam mattress 198 x 90 x 14 cm excluding UK and Italy without handles ASS030 NP100 WD dual density foam mattress 198 x 85 x 14 cm excluding UK and Italy with handles ASS032 NP100 WD dual density foam mattress 198 x 90 x 14 cm excluding UK and Italy with handles ASS022XT NP100 WD dual density foam mattress 198 x 85 x 14 cm UK and Italy only without handles ASS033 NP150 WD viscoelastic foam...

Страница 31: ...e head section AD244B Monitor stand AD271B Pair of metal siderails without attachments AD286A IV line manager support AD290B a Foot egress handle AD296B a Head egress handle AD294A FIXED IV pole AD298A Telescopic IV pole with four hooks AD299A Telescopic IV pole with four hooks AD312A Net for siderail AD271 AD288A Foot gap panels AC968A Equipotential connecting cable AD270B Removable frame AD276A ...

Страница 32: ...must point towards the sleep surface If the headboard is installed in the bed frame the wrong way round the risk of entrapment increases 2020003 Sabina II EE sit to stand lift 2020004 Sabina II EM mobile lift 2040015 Viking M mobile lift 2040013 Viking XL mobile lift 2000014 Golvo 8000 mobile lift 2000015 Golvo 8008 mobile lift 2000019 Golvo 8008 LowBase mobile lift TA270 Bed dining table TA519 Be...

Страница 33: ...rly the use of the accessories installed at the head of the bed e g IV poles helpers etc can incur risks for the patient The headboard can be removed for easier access to the patient s head Bed frame extension Do not sit or climb on the extension The extension can be pulled out by 20 cm in intermediate steps of 4 cm Cushion for extensions is available as an additional mattress Part number Name ASS...

Страница 34: ...TARY AND CONFIDENTIAL DRAFT novembre 02 2020 Wall stop AD277A Located at the head of the bed the extractible wall bumper protects the bed against the risk of impacts with walls or technical ducts Store the bumper during transfers Take out the wall stop Store the wall stop ...

Страница 35: ...ment is reached Caregivers need to assess whether patients can be left unattended with access to the functions on the control pendant Control pendant The control pendant can be stowed under the siderail If the bed was not originally fitted with a control pendant they can be ordered as an additional part with the P N AD282A It can be placed on the right hand side of the bed If the patient pendant i...

Страница 36: ...cts accessories power cables or persons especially children are under the sleep surface and that none of the patient s limbs protrude beyond the edges of the sleep surface An intermittent beep sounds when one of the bed s movements is hindered When descending to the low position make sure that the drainage devices do not come into contact with the floor Use the HiLow feature of the sleep surface t...

Страница 37: ...les preventing the section from being lowered or moving e g limbs electric cables foreign bodies or accessories An intermittent beep sounds when one of the bed s movements is hindered When the thigh section is fully raised the foot section is inclined at an angle of approximately 3 from the sleep surface Electric AutoContour The AutoContour simultaneously raises the head section and the thigh sect...

Страница 38: ...under the sleep surface the patient s limbs are within the sleep surface there is enough space between the head of the bed and the partition especially for Trendelenburg no accessories IV pole in particular may come into contact with the fittings check that the drainage devices do not come into contact with the floor Trendelenburg Reverse Trendelenburg The electrical Trendelenburg Reverse Trendele...

Страница 39: ...s accessory must only be fitted at the head of the bed Fixed patient helper AD810A Safe working load 75 kg 1 Do not position the patient helper at the outside of the bed See incorrect position shown below The patient helper can be fitted into either of the two square sockets at the head of the bed Adjustable patient helper AD811A Safe working load 75 kg 1 The adjustable patient helper can be place...

Страница 40: ... the patient lift some of his her weight so as to assist the nursing staff with their work This position is not designed to allow patients to transfer themselves alone Failure to do so could result in material damage or injury position 1 blue tuck away position position 2 blue normal egress position position 3 yellow patient transfer aid position position 4 red incorrect risk of bed tipping 2 3 4 ...

Страница 41: ...he patient Adjust the height of the handle until there is a right angle at the elbow It is easier for the patient to change position in the bed making for greater comfort and independence Place the patient handle on the patient helper arm when not in use in order eliminate any obstruction see photo below If the bed is equipped with both an adjustable patient helper AD081D AD811A and an IV Pole AD1...

Страница 42: ...t in and out of the bed with greater ease and in safety Incompatible with half siderails Extracting the egress handle Lowering the egress handle If the bed was not originally fitted with egress handles they can be ordered as an accessory with the P N AD290B foot section and AD296B head section Incompatible with half siderails Assistance when moving to a chair CLICK 1 2 2 1 CLICK ...

Страница 43: ...opriately Ensure that all siderails are fully latched when in the raised position Siderails are intended to show patients where the edges of the bed are They are not patient restraining devices When appropriate Hill Rom recommends that medical personnel determine the proper methods necessary to ensure a patient remains safely in bed without being constantly observed Do not place accessories respir...

Страница 44: ...s for Use Electric Bed Centuris Pro PROPRIETARY AND CONFIDENTIAL DRAFT novembre 02 2020 Raising a siderail Lowering the siderail Removing the long siderails Installing the the long siderails CLIC 1 2 3 1 3 2 1 2 3 1 2 1 CLIC CLIC ...

Страница 45: ...gned to cover Centuris Pro medical beds for adults fitted with AD271B metal siderails is secured by press studs and zip fasteners The tightening effect reduces the risk of the patient s head or limbs passing between the bars of the siderails while providing a bright environment for the patient Half siderails Raising a half siderail Siderail in low position Siderail in high position CLIC 1 1 ...

Страница 46: ...bre 02 2020 Lowering a half siderail Foot gap panels AD288A In order to mitigate the risks incurred by patient egress through the gaps at the foot of the bed between the half siderails and the foot panel Hill Rom has developed a kit of two detachable panels one for each side designed to block this gap 2 1 ...

Страница 47: ...els The panels are not designed to restrain or immobilize the patient in the bed Check that the panels are correctly installed The authorized medical personnel must consider the use of siderails depending on the state of health and behavior of the patient according to a protocol that indicates in which situations and when the panels can be used 1 1 ...

Страница 48: ... at the head of the bed and remove from the foot of the bed when not in use Fittings for the restraining strap handles 1 Do not attach the restraining straps to any part of the bed particularly the siderails other than those provided for this purpose When the patient is restrained by the straps the electric functions must be locked out When the patient is restrained with an abdominal strap a syste...

Страница 49: ...ining devices may become entangled and injure the patient or even cause death especially if the patient is agitated and confused Whenever containment devices are used the patient must be observed in accordance with legal requirements and protocol Restraining devices must be secured to the articulated sections of the bed using appropriate attachment points in order to avoid injury to the patient Ne...

Страница 50: ...l keypad press and hold the lock symbol then press the function to be inhibited The indicator light of the corresponding function comes on to indicate that the function is locked out 1 Locking out the thigh section adjustment control will also lock out the AutoContour when the adjustable head section function is activated To enable an electrical function from a half siderail keypad press and hold ...

Страница 51: ...CPR Never allow a non qualified person to operate this function and check that no obstacles e g limbs accessories objects power cables or persons are under the head section This function is used in emergencies e g reanimation cardiac massage or in the event of a power cut It is operated by a handle located centrally and bilaterally under the sleep surface or under the head section if the bed is fi...

Страница 52: ...of all the unprotected metal parts need to be equalized The bed must be connected to the electrical installation To equalize potentials if a grounded power connection is unavailable connect the equipotential cable AC968A to the connection terminal on the bed and the device Equipotential cable AC968A It is fitted with two POAG WB 6 DIN type connectors and a 2 m long yellow and green cable This cabl...

Страница 53: ...e supports and is used to hold IV bags Safe working load Refer to the value indicated on the IV pole Telescopic IV pole AD298A AD299A The IV pole is mounted in the angle supports and is used to hold IV bags Ensure that the IV pole is positioned facing towards the bed and not outwards as shown in the following illustrations Using the IV pole AD298A To adjust the height or angle of the IV pole 1 2 3...

Страница 54: ...g the IV pole AD299A To adjust the height or angle of the IV pole Linen holder The linen holder must not be used to support luggage or as a seat even for young children Do not sit or climb on the linen holder Safe working load 15 kg 1 1 The safe working load specifications allow for a substantial safety margin 1 2 3 4 5 ...

Страница 55: ...gned to accept an oxygen cylinder and must only be fitted on the patient helper supports at the head end of the bed outside the sleep surface It can be rotated through 80 Each type of holder corresponds to a cylinder model and must never be used with a different cylinder See below 1 The safe working load specifications allow for a substantial safety margin OR AC959A for cylinder model B5 Ø140 AD10...

Страница 56: ...the bed to go through a doorway position the holder in front of the bed otherwise place it and the cylinder on the mattress remember to put the holder in its normal position after moving the bed Pivoting 3L Bottle Holder AC962A The bottle holder is designed to accept a 3 litter bottle and can be fitted on the supports at the foot end of the bed outside the sleep surface It can be rotated through 8...

Страница 57: ...onitor ensure that the folded table is located on the outer edge of the bed The table must be folded away when moving the bed If the bed is in Trendelenburg or Reverse Trendelenburg any devices must be placed on the monitor stand To fit a monitor stand Syringe driver holder AC963A Safe working load 15 kg 2 1 The safe working load specifications allow for a substantial safety margin 2 The safe work...

Страница 58: ...n the knob position the tablet as required and then tighten the knob IV line manager support AD286A This accessory must be fitted by an authorized technician Please refer to the fitting instructions provided with the accessory when fitting the accessory A Line Manager is on each side of the head end of the bed The Line Manager helps to keep lines such as IV lines suction lines etc together and awa...

Страница 59: ... The radiologist is responsible for deciding on the best solution to take the X ray according to the medical target and the hospital s protocol adapted to the patient s illness NOTE For patients weighing more than 100kg the user must adjust the angle of the head section and the position of the patient to produce quality images NOTE Incompatible with side rails and AD271B Installing the accessory 1...

Страница 60: ...tte in the sideways direction 9 Adjust the position of the cassette using the right and left straps so that the retaining bar is positioned on the edge of the mattress 10 Adjust the cassette positioning buckle Wind the right strap around the mattress and put the buckle on the upper edge of the mattress Once it has been adjusted using the right and left straps this buckle is used to position the to...

Страница 61: ...the left strap to extract the cassette support 2 Raise the retaining bar and take out the cassette 3 Pull on the right strap to insert the cassette support 4 Hook the buckle of the right strap on its storage hook Chrome plated IV hook AC953A This accessory is used to hold the IV bag to the patient helper AD810A or AD811A Label holder AC325A This additional part is used as a place to holder to inse...

Страница 62: ...Page 56 194568 6 Instructions for Use Electric Bed Centuris Pro PROPRIETARY AND CONFIDENTIAL DRAFT novembre 02 2020 ...

Страница 63: ...e that it does not move The brake bar located at the foot of the bed or the bilateral pedals at the head end simultaneously control all four casters including one steering caster It has three positions STOP to prevent the bed from moving NEUTRAL to move the bed in all directions STEERING for easier movement in a straight line Using the bar in the steering position Three wheels turn freely NEUTRAL ...

Страница 64: ...ace the patient in a stable and comfortable position do not fully raise the head section Never try to move the bed by pulling on the power cable or you may damage it A damaged power cable is an electric shock hazard Never use the patient helper or the IV stand to move the bed The bed should only be moved while in the transport position by two people one at each end so as to ensure that there is al...

Страница 65: ...er cable Always correctly store the power cable Failure to follow this recommendation may result in damage to the cable by crushing and create the risk of electric shock The power cable must be hooked in place before moving the bed Attachment with cable tie AD292A Removable frame AD270B The detachable tube helps to guide the bed when transferring ...

Страница 66: ...Page 60 194568 6 Instructions for Use Electric Bed Centuris Pro PROPRIETARY AND CONFIDENTIAL DRAFT novembre 02 2020 ...

Страница 67: ... excess water on the connectors Refer to the recommendations of the cleaning product manufacturer Thoroughly dry before reusing Failure to implement one or more of these recommendations may lead to damage or deterioration preventing use of the bed and rendering the warranty void Recommendations for cleaning and disinfection The following recommendations are not designed to replace existing cleanin...

Страница 68: ...econtamination record should be kept for each bed mentioning month ward and room number bed reference number cleaning frequency materials and products used Sleep surface Clean and disinfect every day Clean and disinfect after patient departure or before patient transfer Clean and disinfect thoroughly after the departure of an infected patient or recommended every two months CLIC ...

Страница 69: ...uld not be used Formaldehyde or phenol based products and solvents of any kind toluene xylene or acetone Never use abrasives cleaning powder or cleaning pads that may damage components Recommended Cleaning and Disinfection Method Always wipe downward working from the cleanest to the dirtiest areas Do not scrape surfaces Keep wipes damp wet as many times as needed and do not wring out too much wate...

Страница 70: ...omponents control unit actuators lateral caregiver units half siderails with keypads remote controls and control cluster arms do not use accessories such as high pressure hoses A It is preferable to use soft non metallic brushes B and microfiber support C in such a way as to reduce the pressure to an acceptable level Steam cleaning areas prevent water and steam from getting into connectors that ar...

Страница 71: ...oblems with actuators e g blockage contact our after sales service Preventive maintenance A service manual and a catalog of spare parts are supplied on delivery but can also be obtained on demand from Hill Rom After Sales Hill Rom guarantees that the original functional parts or parts performing equivalent functions will remain available for 7 years after the corresponding range goes out of produc...

Страница 72: ...ions and observations the date of the next inspection must be recommended every time the bed is serviced De commissioning The device and its accessories should be cleaned and disinfected before de commissioning Customers should adhere to all federal state regional and or local laws and regulations as it pertains to the safe disposal of medical devices and accessories If in doubt the user of the de...

Страница 73: ...pplying to class I medical devices in accordance with see Declaration of Conformity the Essential Requirements of the directive 93 42 EEC the General Safety and Performance Requirements of the Regulation EU 2017 745 regulation Complies with standards NF S 90 312 1984 EN 60601 1 2006 A1 2013 IEC 60601 1 2005 A1 2012 EN 60601 1 2 2015 IEC 60601 1 2 2014 EN 60601 2 52 2010 IEC 60601 2 52 2009 applica...

Страница 74: ...romagnetic immunity is always relative and standards are based on anticipated environments of usage If the user notices that the device behaves unusually and especially if this behavior is intermittent and occurs when in the vicinity of radio or TV transmitters cell phones or electrosurgical equipment this may be a sign of electromagnetic interference If such behavior occurs users must try to move...

Страница 75: ...nvironment Guide RF emissions CISPR 11 Group 1 The Centuris Pro only uses radio electric power for its internal functions Consequently it only produces very weak RF emissions that are unlikely to cause interference with nearby electronic equipment CISPR 11 RF emissions Class A The Centuris Pro can be used in all places other than domestic premises and premises that are directly connected to the lo...

Страница 76: ...l environment Voltage surges IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode The quality of the main power supply must be that of a typical commercial or hospital environment Magnetic field at the frequency of the mains power supply 50 60 Hz IEC 61000 4 8 30 A m 60 Hz 30 A m 60 Hz The magnetic field at the frequency of the mains supply must be characte...

Страница 77: ...n due to structures objects and persons a The field levels of fixed transmitters such as radio telephone bases cell wireless and terrestrial mobile radios amateur radios and AM FM and TV communication radios cannot be theoretically evaluated precisely Site measurements are required in order to obtain the electromagnetic environment due to fixed RF transmitters If the field level measured in the wo...

Страница 78: ...Maximum assigned power output of the transmitter W Separating distance versus the frequency of the transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 24 0 1 0 37 0 37 0 74 1 1 12 1 12 2 33 10 3 67 3 67 7 37 100 11 6 11 6 23 3 For transmitters with a maximum power output that is not in the list above the recommended separation distance in meters m can be calculate...