MAINTENANCE
8
MAINTENANCE
8
CHECKING AND REPLACING THE AIR INLET FILTER
The air inlet filter is used to filter ambient air entering through the back of the compressor.
The compressor air inlet filter needs to be changed every three to six months, or as needed.
NOTE:
The compressor is shipped with an air inlet filter already installed.
Place the patient on an alternate means of ventilation.
If operating in Stationary or Extended Range Configuration, power off the ventilator.
Power off the compressor.
To inspect the filter, locate the filter on the back of the unit.
Remove the used filter and inspect the filter.
Replace the filter if it is blocked by dirt or dust or is otherwise
contaminated. If replacing the filter, obtain an approved replacement
filter. For more information about purchasing replacement filters, refer to
"Accessories and Replacement Parts" on page 147.
If the filter is clean, return it to the back of the compressor.
NOTES:
• Reusing contaminated filters is not recommended.
• Dispose of used filters in accordance with your facility's guidelines or local regulations.
• The filter might need to be changed more frequently in particulate-filled environments.
WARNING:
• Do not power on or use the compressor without the filters and condensation tray properly installed.
• Do not insert foreign objects into any part of the ventilation system.
CAUTION:
Use only an approved filter with the compressor.
4
5
CHECKING AND REPLACING THE COOLING FILTER ASSEMBLY
The cooling filter is used to filter air that enters the compressor to cool it during operation.
The cooling filter assembly (including the cooling filter and cooling filter cover) needs to be changed every three
to six months, or as needed.
NOTE:
The compressor is shipped with a cooling filter assembly already installed.
Place the patient on an alternate means of ventilation.
If operating in Stationary or Extended Range Configuration, power off the ventilator.
Power off the compressor.
Locate the cooling filter cover on the back of the compressor.
Remove the cooling filter assembly by pulling the cooling
filter cover directly out from the back of the compressor.
Inspect the used cooling filter.
Replace the cooling filter assembly if the filter is blocked
by dirt or dust or is otherwise contaminated. If replacing
the cooling filter assembly, obtain a replacement. For more
information about purchasing replacement parts, refer to
"Accessories and Replacement Parts" on page 147.
Replace the cooling filter assembly on the back of the compressor by pushing it back into place until it is
secure.
NOTES:
• Reusing contaminated filters is not recommended.
• Dispose of used filters in accordance with your facility's guidelines or local regulations.
• The filter might need to be changed more frequently in particulate-filled environments.
WARNING:
• Do not power on or use the compressor without the filters and condensation tray properly installed.
• Do not insert foreign objects into any part of the ventilation system.
CAUTION:
Use only an approved filter with the compressor.