171051(12) - Hillrom™ 900 Electric Bed Instructions for Use
Page 13
Int
rodu
ct
ion
Spe
c
if
ic
at
ions
Electrical characteristics
Characteristic
100V*
120V*
230V*
Voltage
100V AC
120V AC
230V AC
Frequency
50/60 Hz
50/60 Hz
50/60 Hz
Power supply unit maximum power load
300 VA
300 VA
300 VA
Power supply unit fuse rating
2 x 2.5 A T 2 x 2.0 A T 2 x 1.25 A T
Electric shock protection
Class I
Class according to IEC 60601-1
Type B
Protection against harmful ingress of
water (according to IEC 60529)
IPX4
Duty cycle
10% (2min/18min)
a
a. Do not operate electrical functions continuously for more than 2 minutes in any 18 minute
period when the bed is loaded at the safe working load value as this may damage electrical
components. The power supply of the actuator is temporarily cut off if the load factor is
exceeded when using the HiLow.
Conditions required to connect the nurse call system
For more information about the connections required to use the nurse call
function, please refer to the
SideCom® Communication System Design and
Application Manual
(DS059).