Service Part Replacement Procedures
1.1
Reader Replacement
CAUTION:
Caution
—Failure to wear an antistatic strap could cause equipment
damage.
7.
Put on the antistatic strap.
8.
Remove the P.C. board from the reader.
1.
Make sure the bed is unplugged and you have done the steps for
“Setup” on page 25.
2.
Near the reader, remove the cable tie(s) that holds the power cable
with the other cables on the bed.
3.
Remove the heat shrinks from
the antenna cable connections.
4.
Disconnect the antenna-reader cables from the antenna cables.
5.
Remove the reader from the bed.
VersaCare® Bed
Progressa® Bed
6.
Remove the six screws that
attach the back cover to the
reader.