WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
8
Quick View™ List of Features
Indicator Light—Visual Identification
E
Centrella®
Bed with a different
nurse call system
—
a.
Connect the two-cable
end of the WatchCare™
adapter to the bed.
b.
Connect the 37-pin end of the adapter cable to the
facility’s communication cable.
c.
Connect the remaining end (the 1/4" cable
connector) of the adapter cable to the facility’s
equipment call jack designated for the WatchCare™
System.
Status
Indicator
Light
Solid green light—identifies that a smart pad is present and
is being monitored. No moisture is detected at this time.
Flashing amber light—identifies that the smart pad is wet.
This visual alert will project on the floor.
Solid white light—identifies that the monitor system is on,
but the reader does not detect any smart pads on the bed.
Alternating white and green light—identifies that the
monitor system can not operate effectively because more
than four smart pads are detected.
No light—the monitor system is not active. Make sure the
bed is plugged into a power outlet.
Item
Description