4.1 Counterweight
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 6
TotalCare® Bariatric Bed Service Manual (man299ra)
4.1
Counterweight
Tools required:
1/2" socket
Ratchet
Removal
1. Set the brakes.
2. Raise the bed to the highest position.
SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so can result in
personal injury or equipment damage.
3. Unplug the bed from its power source.
4. Remove the head end covers. Refer to “Cover—Head End” in the
TotalCare® Bed System Service Manual (man112)
5. Install the Hi-Lo cylinder braces. Refer to “Hi-Lo Cylinder Brace—
TotalCare® Bed System Service Manual (man112)
6. Remove the three nuts (A) securing the counterweight (C) to the bed (D)
WARNING:
The counterweight weighs in excess of 75 lb (34 kg). Use two people
when removing or installing the counterweight. Failure to do so can
result in personal injury.
7. Remove the counterweight (C).
Содержание TotalCare P1830A
Страница 2: ......
Страница 6: ...Revisions Page iv TotalCare Bariatric Bed Service Manual man299ra NOTES...
Страница 26: ...Chapter 2 Troubleshooting Procedures Page 2 2 TotalCare Bariatric Bed Service Manual man299ra NOTES...
Страница 30: ...Chapter 3 Theory of Operation Page 3 2 TotalCare Bariatric Bed Service Manual man299ra NOTES...
Страница 48: ...Chapter 5 Parts List Page 5 2 TotalCare Bariatric Bed Service Manual man299ra NOTES...
Страница 50: ...Warranty Chapter 5 Parts List Page 5 4 TotalCare Bariatric Bed Service Manual man299ra NOTES...
Страница 61: ...Head Siderail Module Chapter 5 Parts List TotalCare Bariatric Bed Service Manual man299ra Page 5 15 5 NOTES...
Страница 74: ...Chapter 6 General Procedures Page 6 2 TotalCare Bariatric Bed Service Manual man299ra NOTES...
Страница 92: ...Chapter 7 Accessories Page 7 2 TotalCare Bariatric Bed Service Manual man299ra NOTES...