100
Synclara™ Cough System, Model PSC1 User Manual (196286 REV E)
Specifications
E
LECTROMAGNETIC
I
MMUNITY TO
W
IRELESS
C
OMMUNICATIONS
The Synclara™ Cough System, Model PSC1 is intended for use in the electromagnetic
environment specified below. The customer or user of the Synclara™ Cough System,
Model PSC1 should make sure it is used in such an environment.
Sides Tested
Frequency
(MHz)
Test Severity Level
Test Distance
(m)
Front, back, left, right
385
27 V/m, 50% PM 18 Hz
0.3
Front, back, left, right
450
28 V/m, FM ± 5 kHz,1 kHz
0.3
Front, back, left, right
710
9 V/m, 50% PM 217 Hz
0.3
Front, back, left, right
745
9 V/m, 50% PM 217 Hz
0.3
Front, back, left, right
780
9 V/m, 50% PM 217 Hz
0.3
Front, back, left, right
810
28 V/m, 50% PM 18 Hz
0.3
Front, back, left, right
870
28 V/m, 50% PM 18 Hz
0.3
Front, back, left, right
930
28 V/m, 50% PM 18 Hz
0.3
Front, back, left, right
1720
28 V/m, 50% PM 217 Hz
0.3
Front, back, left, right
1845
28 V/m, 50% PM 217 Hz
0.3
Front, back, left, right
1970
28 V/m, 50% PM 217 Hz
0.3
Front, back, left, right
2450
28 V/m, 50% PM 217 Hz
0.3
Front, back, left, right
5240
9 V/m, 50% PM 217 Hz
0.3
Front, back, left, right
5500
9 V/m, 50% PM 217 Hz
0.3
Front, back, left, right
5785
9 V/m, 50% PM 217 Hz
0.3