Electrically Operated General Ward Bed
Page 2 - 1
Operation and Maintenance Manual
(154588 REV 1)
2
Chapter 2
Troubleshooting Procedures
Getting Started
WARNING:
Only facility-authorized persons should service the bed. Service by
unauthorized persons could cause injury or equipment damage.
Begin each procedure in this chapter with step 1. Follow the sequence outlined
(each step assumes the previous step has been completed). In each step, the
normal operation of the product can be confirmed by a
Yes
or
No
answer to the
statement. Your response will lead to another step in the procedure, a repair
analysis procedure (RAP), or a component replacement. If more than one
component is listed, replace them in the given order.
To collect data about the problem, start with
Initial Actions
.
To identify a problem and to make sure of the repair after you complete each
corrective action (part replacement or adjustment, connector installation, etc.),
do the
Function Checks
.
To make sure the repair corrected the problem, do the
Final Actions
after the
Function Checks.
If troubleshooting procedures do not identify the problem, contact
Medicraft Hill-Rom Technical Support.