Maintenance
WARNING:
Only facility-authorized personnel should service the Excel Care™ Bariatric Bed.
Servicing by unauthorized personnel could result in personal injury or equipment
damage.
WARNING:
Before bed use, correct any differences noted during the examination. Failure to
do so could result in patient injury or equipment damage.
The amount of time the bed is used dictates the amount of maintenance necessary. To
make sure the bed operates correctly, do the periodic bed examinations. Before bed use,
correct differences noted during the examination.
Do the following examination after the first two weeks of use and then, at a minimum,
every six months:
1. Make sure all bolts are fully tightened.
2. Make sure all pins are in the locked position and are fully connected.
3. Examine all welds.
4. Make sure all electrical connections are connected.
Do this examination every six months:
1. Use light grease to lubricate all slides and pivot points.
2. Examine all electrical wiring for any frays, kinks, and/or deterioration.
3. Examine the covering of the bed’s control panel, including the patient lockout
features, to make sure the covering is not cracked or damaged.
37