Clean the Excel Air™ Low Airloss Therapy Surface Filter
CAUTION:
Keep the filters clean. Failure to do so could result in equipment damage.
To keep the Excel Air™ Low Airloss
Therapy Surface operating correctly, make
sure the filter is clean. After 800 hours of
operation, when the air supply unit is turned
on, the Replace Filter screen shows as a
reminder replace the filters. To replace the
filters later, press the Remind me later control. When the air supply unit is turned on
again, the Replace Filter screen will show.
1. Unplug the air supply unit from its power source.
2. At the bottom of the air supply unit, open the two filter grill covers, and remove the
filters. Do not unscrew the filter assembly from the air supply unit.
NOTE:
The white filter housing is the air intake filter. The black filter housing is the cooling fan
filter.
3. Wash the filters in mild detergent, and let them air dry.
4. If the filters cannot be cleaned, or if the filters are damaged, replace them.
5. Set the clean or new filters in the filter housings, and close the filter grill covers.
6. Install the white filter housing under the side with three connectors.
7. Install the black filter housing under the side with six connectors.