Clean the Mattress Assembly
Coverlet (Excel Air™ Low Airloss Therapy Surface Only)
Between patients, and for heavy soilage such as visible signs of body fluids and/or
substances, do the following to completely clean and disinfect the coverlet:
1. Machine-wash the coverlet with chlorine bleach (50 ppm to 150 ppm) or detergent and
an effective intermediate level disinfectant, such as PRO-TECH™
1
disinfectant
cleaner. (For customers in the US, the disinfectant should be registered with the
Environmental Protection Agency.)
NOTE:
2.5 oz of bleach per 10 g of water is approximately 100 ppm of available chlorine.
a. Use the bleach or disinfectant as instructed in the manufacturer’s instructions.
b. To determine the amount of bleach or disinfectant to use, determine the amount of
water in the washer, and follow the manufacturer’s dilution instructions.
c. During the wash cycle, soak the coverlet in the disinfectant or bleach.
d. Let the coverlet rinse thoroughly in clean water.
CAUTION:
Do not dry the coverlet using high temperatures. Air dry, or select a low or non-
heat dry cycle such as air fluff. High temperatures could destroy the coating that
makes the coverlet waterproof yet breathable.
2. Use the lowest temperature setting of the dryer to dry the coverlet. Do not use high
temperatures.
For light soilage when there are no visible signs of body fluids or substances, do the
following to sanitize the coverlet:
1. Wipe down the coverlet with chlorine bleach (50 ppm to 150 ppm) or mild detergent
and warm water followed by an approved intermediate level disinfectant, such as
PRO-TECH™ disinfectant cleaner:
a. Let the bleach or disinfectant remain in contact with the surface as instructed in the
manufacturer’s instructions.
b. Remove the bleach or disinfectant, and rinse with warm water.
2. Let the coverlet completely air dry.
Bottom Shell and Vapor Barrier
Wipe down the bottom shell and vapor barrier with an approved disinfectant, and let them
air dry.
1. PRO-TECH™ is a trademark of Central Solutions, Inc.