Make sure that the patient, equipment, and all lines are securely placed within the perimeter of the bed for
intra-hospital transport. The CareAssist™ Bed is not intended to be used to transport a patient in the
Dining Chair™ Position or FullChair® Patient position.
Fully extended IV poles could impact doorways or ceiling fixtures. Lower poles prior to patient transport.
Make sure the Nurse Call system cables are properly connected after transport.
Sleep Surface/Mattress
WARNING:
Some safety features of the CareAssist™ Bed may not function or may not operate as intended
with surfaces manufactured by other companies. Check with the surface manufacturer to
determine those safety features of the bed that have been tested and verified to work properly
with the replacement surface. Failure to do so could result in serious personal injury or damage
to equipment.
NOTE:
Hill-Rom recommends the use of Hill-Rom surfaces that have been designed and tested specifically for
the CareAssist™ Bed. Customers electing to purchase replacement surfaces from other manufacturers
should confirm that the replacement surface, when used in conjunction with the CareAssist™ Bed, meets
applicable regulations, regulatory guidance and technical standards and does not create unacceptable
risks of injury to patients or caregivers. Specifically, Hill-Rom suggests that surfaces utilize dimensions
and construction to minimize gaps where entrapment could occur, provide for sufficient height between
the surface and the top of the siderail to prevent accidental roll-over events, provide appropriate firmness
at the edges of the surface to facilitate safe transfers into and out of the bed, and do not interfere with the
proper operation of siderails.
WARNING:
Sleep surface impermeability could be affected by needle sticks or punctures caused by
improper use of x-ray cassette holders and/or needle sticks. Personal injury or infection could
result.
The sleep surface should be regularly inspected for such damage.
Flammability
WARNING:
Patients should not be allowed to smoke in bed. Sheets and pillows generally do not have flame
resistant properties. Personal injury or equipment damage could occur.
Reduce the possibility of fires by observing fire prevention rules and regulations.
The sleep surface mattress meets the following specifications:
•
CAL TB-117, Requirements, Test Procedure and Apparatus for Testing the Flame Retardance of
Resilient Filling Materials Used in Upholstered Furniture (foam)
•
16 CFR 1632, Standard for the Flammability of Mattresses and Mattress Pads
•
CAL TB-129, Flammability Test Procedures for Mattresses for Use in Public Buildings (models
with fire barrier only)
•
BFD IX-II, Boston Fire Department Mattress Fire Test (models with fire barrier only)