SERVICE MANUAL

Basic Care™ 
Bed

From Hill-Rom

Product No. P1440/P1441

For Parts or Technical Assistance

 man336

USA 1-812-934-1796 Canada (800) 267-2337
International: Contact your distributor.

Содержание Basic Care P1440

Страница 1: ...SERVICE MANUAL Basic Care Bed From Hill Rom Product No P1440 P1441 For Parts or Technical Assistance man336 USA 1 812 934 1796 Canada 800 267 2337 International Contact your distributor...

Страница 2: ......

Страница 3: ...to third parties without the prior written consent of Hill Rom First Edition First Printing 2004 Printed in the USA Allen is a trademark of Industrial Fasteners Inc Basic Care is a trademark of Hill R...

Страница 4: ...tal of its products To order additional copies of this manual man336 refer to the back cover for contact information For countries not listed on the back cover contact your distributor NOTE The back c...

Страница 5: ...Chapter 2 Troubleshooting Procedures Getting Started 2 1 Initial Actions 2 2 Function Checks 2 2 Final Actions 2 4 None of the Bed Functions Operate 2 5 The Head Up Function Does Not Operate 2 6 The H...

Страница 6: ...Frame 3 1 Articulating Sleep Deck 3 1 Siderails 3 1 Motion Control 3 1 Transformer Configurations 3 2 Motor Assemblies 3 3 Battery Backup 3 3 Controls 3 3 Chapter 4 Removal Replacement and Adjustment...

Страница 7: ...djustment 4 19 CPR Cable 4 20 Removal 4 20 Replacement 4 22 Adjustment 4 22 Knee Motor 4 23 Removal 4 23 Replacement 4 24 Hilow Motor 4 25 Removal 4 25 Replacement 4 26 Manual Crank Handle Assembly 4...

Страница 8: ...lacement 4 41 Siderail Retraction Assembly 4 42 Removal 4 42 Replacement 4 43 Siderail Control Assembly 4 44 Removal 4 44 Replacement 4 45 Siderail Control Cable 4 47 Removal 4 47 Replacement 4 49 Sid...

Страница 9: ...e Parts Ordering 5 1 Exchange Policy 5 3 In Warranty Exchanges 5 3 Out of Warranty Exchanges 5 3 Warranty 5 5 Recommended Spare Parts 5 6 Base Frame 5 8 Upper Frame for the Electric Model Basic Care B...

Страница 10: ...edule 6 6 Preventive Maintenance Checklist 6 7 Basic Care Bed Sleep Mode 6 8 CPR Cable Adjustment 6 10 Chapter 7 Accessories Basic Care Bed Accessories 7 1 IV Pole P1445 7 2 Installation 7 2 Removal 7...

Страница 11: ...parts lists in chapter 5 for ordering replacement components Audience This manual is intended for use by only facility authorized personnel Ignoring this restriction can result in severe injury to pe...

Страница 12: ...ow shows a WARNING or CAUTION Figure 1 1 Warning and Caution A WARNING identifies situations or actions that could affect patient or user safety Disobeying a warning could result in patient or user in...

Страница 13: ...pacity maximum 200 lb 91 kg Maximum height of seat section in Tren delenburg position 23 5 59 7 cm Siderail opening size 3 875 9 843 cm Distance between siderails 11 375 28 893 cm Length with standard...

Страница 14: ...to 206 cm Mattress length short model 76 to 77 193 cm to 196 cm Mattress thickness 6 to 7 15 cm to 18 cm Condition Range Temperature 40 F to 158 F 40 C to 70 C Relative humidity RH 10 to 95 Atmospher...

Страница 15: ...fication Standard Technical and Quality Assur ance Standards UL 60601 1 CSA a C22 2 No 601 1 EN 60601 2 38 including amendment 1 EN 60601 1 IEC 60601 1 2 EN ISO 9002 a CSA is a registered trademark of...

Страница 16: ...lectronic equipment RF Emissions CISPR 11 Class A The Basic Care Bed is suitable for use in all establishments other than domestic and those directly connected to the public low voltage power supply n...

Страница 17: ...1000 4 5 1 kV Differential Mode 2 kV Common Mode 1 kV Differential Mode 2 kV Common Mode Mains power quality should be that of a typi cal commercial or hospi tal environment Conducted RF IEC 61000 4 6...

Страница 18: ...or inter ruption the Basic Care Bed should be switched to operate from the backup battery Note 1 UT is the AC mains voltage prior to application of the test level Note 2 The compliance levels in the...

Страница 19: ...Chapter 1 Introduction Basic Care Bed Service Manual man336 Page 1 9 1 Model Identification Table 1 10 Model Identification Model Number Description P1440 Basic Care Bed electric model P1441 Basic Ca...

Страница 20: ...serious injury or death NOTE Siderails are intended to be a reminder to the patient of the unit s edges not a patient restraining device When appropriate Hill Rom recommends that medical personnel de...

Страница 21: ...vicing by unauthorized personnel could result in personal injury or equipment damage WARNING Obey the product manufacturer s instructions Failure to do so could result in personal injury or equipment...

Страница 22: ...amage WARNING Adhere to appropriate infection control policies and procedures Failure to do so could result in the spread of infection WARNING Failure to install the IV pole correctly could allow it t...

Страница 23: ...ld result in equipment damage CAUTION Do not use harsh cleansers disinfectants such as scouring pads heavy duty grease removers solvents such as toluene xylene or acetone Equipment damage could occur...

Страница 24: ...s footboard or other designated location Failure to do so can result in equipment damage CAUTION Do not lower the bed frame while the trapeze support assembly is attached to the bed Turn off the hilow...

Страница 25: ...Warning and Caution Labels Chapter 1 Introduction Basic Care Bed Service Manual man336 Page 1 15 1 Warning and Caution Labels Figure 1 5 Warning and Caution Labels...

Страница 26: ...Warning and Caution Labels Chapter 1 Introduction Page 1 16 Basic Care Bed Service Manual man336 Figure 1 6 Battery Backup Wiring Diagram...

Страница 27: ...de by answering Yes or No to the statement Your response will lead to another step in the procedure a repair analysis procedure RAP or a component replacement If more than one component is listed repl...

Страница 28: ...correct operator action Yes No Go to Function Checks on page 2 2 4 Instruct the operator to refer to the Basic Care Bed User Manual usr124 To make sure the Basic Care Bed is operating correctly do the...

Страница 29: ...he Hilow Down switch The bed lowers Yes No Go to page 2 18 10 Raise the bed to the high position Pull the Trendelenburg handle and press the Hilow Down switch The bed goes into the Trendelenburg posit...

Страница 30: ...on each caster having that function Try to move the bed out of its position The bed does not move Yes No Replace the casters that are not operating correctly refer to procedure 4 25 on page 4 59 16 I...

Страница 31: ...The power cord is tightly seated in the connector on the control box Yes No Tightly seat the power cord into its connector on the control box If this corrects the problem go to Final Actions on page...

Страница 32: ...or for the head motor into its socket on the control box The head up function does not operate Yes No Go to Final Actions on page 2 4 4 Replace the head motor The head up function does not operate Yes...

Страница 33: ...2 8 Replace the controls on the siderail with the faulty function The head up function operates from this siderail Yes No Replace the siderail with the faulty function If this corrects the problem go...

Страница 34: ...or for the head motor into its socket on the control box The head down function does not operate Yes No Go to Final Actions on page 2 4 4 Replace the head motor The head down function does not operate...

Страница 35: ...Replace the controls having the head down function that does not operate The head down function operates from this siderail Yes No Replace the siderail with the faulty function If this corrects the pr...

Страница 36: ...D for a bed function lockout is on the function is locked out Yes No Turn the head function lockout off If this corrects the problem go to Final Actions on page 2 4 2 Disconnect the bed from its power...

Страница 37: ...ctor for the knee motor into its socket on the control box The knee up function does not operate Yes No Go to Final Actions on page 2 4 4 Replace the knee motor The knee up function does not operate Y...

Страница 38: ...on the siderail having the knee up function that does not operate The knee up function operates from this siderail Yes No Replace the siderail having the knee up function that does not operate If thi...

Страница 39: ...ctor for the knee motor into its socket on the control box The knee down function does not operate Yes No Go to Final Actions on page 2 4 4 Replace the knee motor The knee down function does not opera...

Страница 40: ...n the siderail having the knee down function that does not operate The knee down function operates from this siderail Yes No Replace the siderail having the knee down function that does not operate If...

Страница 41: ...ED for a bed function lockout is on the function is locked out Yes No Turn the knee function lockout off If this corrects the problem go to Final Actions on page 2 4 2 Disconnect the bed from its powe...

Страница 42: ...r for the hilow motor into its socket on the control box The hilow up function does not operate Yes No Go to Final Actions on page 2 4 4 Replace the hilow motor The hilow up function does not operate...

Страница 43: ...on the siderail having the hilow up function that does not operate The hilow up function operates from this siderail Yes No Replace the siderail having the hilow up function that does not operate If...

Страница 44: ...tor for the hilow motor into its socket on the control box The hilow down function does not operate Yes No Go to Final Actions on page 2 4 4 Replace the hilow motor The hilow down function does not op...

Страница 45: ...n the siderail having the hilow down function that does not operate The hilow down function operates from this siderail Yes No Replace the siderail having the hilow down function that does not operate...

Страница 46: ...LED for a bed function lockout is on the function is locked out Yes No Turn the hilow function lockout off If this corrects the problem go to Final Actions on page 2 4 2 Disconnect the bed from its po...

Страница 47: ...this corrects the problem go to Final Actions on page 2 4 3 Do the procedure on page 2 6 The head section rises when the control for the auto contour up is pressed Yes No Call Hill Rom Technical Supp...

Страница 48: ...this corrects the problem go to Final Actions on page 2 4 3 Do the procedure on page 2 8 The head section lowers when the control for the auto contour down is pressed Yes No Call Hill Rom Technical S...

Страница 49: ...fer to procedure 4 3 on page 4 7 If this corrects the problem go to Final Actions on page 2 4 2 Visually inspect the Trendelenburg or Reverse Trendelenburg rods C The Trendelenburg or Reverse Trendele...

Страница 50: ...rendelenburg or Reverse Trendelenburg Function Does Not Operate Chapter 2 Troubleshooting Procedures Page 2 24 Basic Care Bed Service Manual man336 Figure 2 1 Trendelenburg and Reverse Trendelenburg A...

Страница 51: ...The CPR wire C is held tightly in the bracket D Yes No Tighten the screw E that attaches the CPR wire C under the bracket D If this corrects the problem go to Final Actions on page 2 4 Otherwise go t...

Страница 52: ...2 14 The CPR Function Does Not Operate Chapter 2 Troubleshooting Procedures Page 2 26 Basic Care Bed Service Manual man336 Figure 2 2 CPR Control Assembly...

Страница 53: ...different bed sections Articulating Sleep Deck The articulating sleep deck contains the head seat knee and foot sections These sections except for the seat section articulate in a manner that changes...

Страница 54: ...tions on the control board see figure 3 1 on page 3 2 NOTE The control board is configured for either T1 or T2 transformers Figure 3 1 Transformer Configurations For the correct transformer configurat...

Страница 55: ...g load SWL of 450 lb 204 kg on the bed For the battery backup wiring diagram see figure 1 6 on page 1 16 Controls On the electric Basic Care Bed patients or caregivers can change the angle of the head...

Страница 56: ...Introduction Chapter 3 Theory of Operation Page 3 4 Basic Care Bed Service Manual man336 NOTES...

Страница 57: ...d supplies are necessary Phillips head screwdriver Screwdriver Small screwdriver 13 mm socket 17 mm socket Ratchet 8 mm nut driver deep well 13 mm nut driver 6 mm wrench 8 mm wrench 17 mm wrench 2 5 m...

Страница 58: ...ly Requirements Chapter 4 Removal Replacement and Adjustment Procedures Page 4 2 Basic Care Bed Service Manual man336 Wire cutters Needle nose pliers Rubber mallet Bed supports Straps Bed frame jack C...

Страница 59: ...its power source Failure to do so could result in personal injury or equipment damage 3 Disconnect the bed from its power source 4 Disconnect the cables A from the control box assembly B see figure 4...

Страница 60: ...1 Control Box Assembly Replacement 1 Do the removal procedure in the reverse order CAUTION Connect the cables to the correct socket Failure to do so may cause the equipment to operate incorrectly 2 M...

Страница 61: ...s head screwdriver Wire cutters Removal SHOCK HAZARD Disconnect the bed from its power source Failure to do so could result in personal injury or equipment damage 1 Disconnect the bed from its power s...

Страница 62: ...of the wires ties on the two cables C from the foot end control panel A to the bed frame D 4 Disconnect the two cables C from the control box assembly B 5 Remove the footboard E from the bed frame D 6...

Страница 63: ...rsonal injury or equipment damage 1 Disconnect the bed from its power source 2 Disconnect the battery cable from the control board refer to procedure 4 7 on page 4 16 3 Remove the two retaining rings...

Страница 64: ...pter 4 Removal Replacement and Adjustment Procedures Page 4 8 Basic Care Bed Service Manual man336 Figure 4 3 Trendelenburg Control Box Replacement 1 Do the removal procedure in the reverse order 2 Do...

Страница 65: ...e 3 Put bed supports under the upper bed frame A see figure 4 4 on page 4 10 SHOCK HAZARD Disconnect the bed from its power source Failure to do so could result in personal injury or equipment damage...

Страница 66: ...ndelenburg or Reverse Trendelenburg Assembly Chapter 4 Removal Replacement and Adjustment Procedures Page 4 10 Basic Care Bed Service Manual man336 Figure 4 4 Trendelenburg or Reverse Trendelenburg As...

Страница 67: ...3 Insert the Trendelenburg rod C into the trip lever E 4 Install the retaining ring B to attach the Trendelenburg rod C to the trip lever E 5 Install the retaining ring F to attach the Trendelenburg...

Страница 68: ...equired Screwdriver Rubber mallet Whenever one of the Trendelenburg indicators needs replaced or reset adjust the indicator caps Removal Carefully pry the indicator cap A from the lift arm B see figur...

Страница 69: ...enburg and Reverse Trendelenburg CAUTION Use care when removing indicator caps Failure to do so could damage the indicator cap 4 If the indicator cap A is installed carefully remove it 5 Align the mar...

Страница 70: ...the head section to the highest position 2 Adjust the bed height to have access to the cover A see figure 4 6 on page 4 14 Figure 4 6 Control Box Assembly SHOCK HAZARD Disconnect the bed from its powe...

Страница 71: ...Service Manual man336 Page 4 15 4 5 Disconnect the battery cable D from the control board E 6 Remove the fuse F from the fuse assembly G 7 Remove the fuse H from the control board E Replacement 1 Do...

Страница 72: ...1 Raise the head section to the highest position 2 Adjust the bed height to have access to the cover A see figure 4 7 on page 4 16 Figure 4 7 Control Box Assembly SHOCK HAZARD Disconnect the bed from...

Страница 73: ...E from the control board F 7 Disconnect the battery cable E and the fuse assembly G from the two batteries D CAUTION Dispose of batteries according to your local regulations Failure to dispose of batt...

Страница 74: ...A lower the head section B to the lowest position see figure 4 8 on page 4 19 5 Make a note showing the cable routing from the head motor C to the control box assembly D CAUTION Use caution when cutti...

Страница 75: ...o allow the bed to move through its full range of motion without putting force on any of the cables Failure to do so could result in equipment damage 1 When routing the cables make sure they have enou...

Страница 76: ...power source Failure to do so could result in personal injury or equipment damage 4 Disconnect the bed from its power source 5 Make a note showing the cable routing from the CPR handle assembly A to t...

Страница 77: ...4 9 CPR Cable Chapter 4 Removal Replacement and Adjustment Procedures Basic Care Bed Service Manual man336 Page 4 21 4 Figure 4 9 CPR Cable Assembly...

Страница 78: ...UTION Keep enough slack in the cables to allow the bed to move through its full range of motion without putting force on any of the cables Failure to do so could result in equipment damage 1 When rout...

Страница 79: ...head section A WARNING Attach the foot section to the head section Failure to do so could result in personal injury 5 Use straps to attach the foot section B to the head section A SHOCK HAZARD Disconn...

Страница 80: ...e F 11 Remove the knee motor D Replacement CAUTION Keep enough slack in the cables to allow the bed to move through its full range of motion without putting force on any of the cables Failure to do so...

Страница 81: ...rol board refer to procedure 4 7 on page 4 16 3 Raise the head section A to its highest position see figure 4 11 on page 4 26 SHOCK HAZARD Disconnect the bed from its power source Failure to do so cou...

Страница 82: ...L and pin M attaching the other half of the hilow motor B to the bed frame E Replacement CAUTION Keep enough slack in the cables to allow the bed to move through its full range of motion without putti...

Страница 83: ...hese sections cannot be raised this procedure can be done from underneath the foot end WARNING Attach the foot section to the head section Failure to do so could result in personal injury 3 Use straps...

Страница 84: ...pter 4 Removal Replacement and Adjustment Procedures Page 4 28 Basic Care Bed Service Manual man336 Figure 4 12 Manual Crank Handle Assembly Replacement 1 Do the removal procedure in the reverse order...

Страница 85: ...the head section Failure to do so could result in personal injury 3 Use straps to attach the foot section to the head section 4 Remove the footboard A from the bed frame B see figure 4 13 on page 4 30...

Страница 86: ...Crank Rod and Drive Screw Assembly Manual Model Only Chapter 4 Removal Replacement and Adjustment Procedures Page 4 30 Basic Care Bed Service Manual man336 Figure 4 13 Head or Knee Crank Rod and Driv...

Страница 87: ...r 4 Removal Replacement and Adjustment Procedures Basic Care Bed Service Manual man336 Page 4 31 4 12 Remove the two bolts L attaching the drive screw assembly H to the bed frame B Replacement 1 Do th...

Страница 88: ...in personal injury or equipment damage 1 Put bed supports under the upper bed frame 2 Raise the knee section and the head section to their highest positions 3 Lift the foot section of the sleep surfac...

Страница 89: ...rew Assembly Including Gas Springs Manual Model Only Chapter 4 Removal Replacement and Adjustment Procedures Basic Care Bed Service Manual man336 Page 4 33 4 Figure 4 14 Hilow Crank Rod and Drive Scre...

Страница 90: ...he bed frame B 12 Remove the two bolts L attaching the drive screw assembly H and the two gas springs M to the bed frame B 13 Remove the two retaining rings N two spacers O and pin P attaching the two...

Страница 91: ...Failure to do so could result in personal injury or equipment damage 3 Disconnect the bed from its power source 4 Disconnect the battery cable from the control board refer to procedure 4 7 on page 4 1...

Страница 92: ...t and Adjustment Procedures Page 4 36 Basic Care Bed Service Manual man336 Figure 4 15 CPR Handle Assembly Replacement Do the removal procedure in the reverse order Adjustment Adjust the tension on th...

Страница 93: ...ction B and the knee section D to their highest positions b Lift the foot section E of the sleep surface and fold it back toward the head section B WARNING Attach the foot section to the head section...

Страница 94: ...4 16 Siderails Chapter 4 Removal Replacement and Adjustment Procedures Page 4 38 Basic Care Bed Service Manual man336 Figure 4 16 Siderails...

Страница 95: ...nd discard the label I on the inside of the siderail control assembly J NOTE The new headrail comes with a new label for the inside of the siderail control assembly b Remove the siderail control assem...

Страница 96: ...d from its power source Failure to do so could result in personal injury or equipment damage 1 Disconnect the bed from its power source 2 Disconnect the battery cable from the control board refer to p...

Страница 97: ...apter 4 Removal Replacement and Adjustment Procedures Basic Care Bed Service Manual man336 Page 4 41 4 Figure 4 17 Slide Bracket Replacement 1 Do the removal procedure in the reverse order 2 Do the Fu...

Страница 98: ...andle Allen wrench Phillips head screwdriver Screwdriver Removal 1 Raise the bed to the highest position 2 Lower the siderail A to the down position see figure 4 18 on page 4 42 Figure 4 18 Siderail R...

Страница 99: ...Failure to do so may result in eye injury 4 Put on eye protection 5 Remove the two screws D attaching the top cane E to the siderail A 6 Remove the two spring assemblies F from the siderail A Replace...

Страница 100: ...iver deep well Removal SHOCK HAZARD Disconnect the bed from its power source Failure to do so could result in personal injury or equipment damage 1 Disconnect the bed from its power source 2 Disconnec...

Страница 101: ...iderail control assembly D 10 Remove and discard the two halves of the siderail control assembly D and B Replacement 1 Remove and save the center screw C from the new outer half of the siderail contro...

Страница 102: ...rail Control Assembly Chapter 4 Removal Replacement and Adjustment Procedures Page 4 46 Basic Care Bed Service Manual man336 9 Connect the bed to the correct power source 10 Do the Function Checks on...

Страница 103: ...re 4 7 on page 4 16 4 Make a note showing the cable routing from the siderail A to the control box assembly B see figure 4 20 on page 4 48 5 Remove the two screws C attaching the wire cover D to the s...

Страница 104: ...4 20 Siderail Control Cable Chapter 4 Removal Replacement and Adjustment Procedures Page 4 48 Basic Care Bed Service Manual man336 Figure 4 20 Siderail Control Cable...

Страница 105: ...trol box assembly B Replacement CAUTION Keep enough slack in the cables to allow the bed to move through its full range of motion without putting force on any of the cables Failure to do so could resu...

Страница 106: ...uired Phillips head screwdriver Removal 1 Raise the siderail A to its up and locked position see figure 4 21 on page 4 50 2 Remove the two screws B attaching the top cane C to the siderail A 3 Remove...

Страница 107: ...llips head screwdriver Removal 1 Raise the siderail A to its up and locked position 2 Remove the two screws B attaching the siderail wire cover C to the siderail A 3 Remove and discard the siderail wi...

Страница 108: ...uipment damage 2 Disconnect the bed from its power source Remove the Foot End WARNING Put caster blocks behind the casters at the opposite end of the bed Failure to do so could result in personal inju...

Страница 109: ...he caster the caster will fall off the bed frame 8 If necessary remove the bolt K and nut L attaching the linkage M to the rocker arm J 9 Remove the bolt N and washer O attaching the swivel caster P t...

Страница 110: ...4 Remove the bolt S and washer T attaching the brake caster U to the bed frame D see figure 4 24 on page 4 54 Figure 4 24 Central Brake and Steer Brake Caster Assemblies 5 Remove the bolt V and washe...

Страница 111: ...23 on page 4 53 NOTE When left hand and right hand sides of the bed are identified they are from the patient s point of view CAUTION Make sure that a flat side of the hex rod is facing upward when it...

Страница 112: ...of the hex rod is facing upward when it is installed Failure to do so may result in the central brake and steer not operating correctly 3 Push the hex rod X through the opening in the bed frame D and...

Страница 113: ...bed from its power source 3 Put the central brake and steer pedals A in the neutral position see figure 4 25 on page 4 58 4 Remove the two screws B attaching the two pedals A to the hex rod C 5 Remove...

Страница 114: ...d Steer Pedal Assemblies Replacement CAUTION Make sure that a flat side of the hex rod is facing upward when it is installed Failure to do so may result in the central brake and steer not operating co...

Страница 115: ...the bed from its power source Failure to do so could result in personal injury or equipment damage 2 Disconnect the bed from its power source WARNING Put caster blocks behind the casters at the oppos...

Страница 116: ...er 4 Removal Replacement and Adjustment Procedures Page 4 60 Basic Care Bed Service Manual man336 Figure 4 26 Casters for Individual Brake and Steer Replacement 1 Do the removal procedure in the rever...

Страница 117: ...moval 1 Raise the bed to its highest position 2 Remove the screw A and bushing B attaching the bumper C to the bed frame D see figure 4 27 on page 4 61 Figure 4 27 Bumper 3 Remove the bumper C from th...

Страница 118: ...e sections to their lowest positions 3 If you are replacing a seat panel A or thigh panel B do the following see figure 4 28 on page 4 62 a Remove the four retaining rings C and four pins D attaching...

Страница 119: ...5 If you are replacing a head panel H do the following a Remove the two headrails I from the bed frame E refer to procedure 4 16 on page 4 37 b Remove the CPR handle assembly J from the bed frame E re...

Страница 120: ...4 27 Sleep Surface Panels Chapter 4 Removal Replacement and Adjustment Procedures Page 4 64 Basic Care Bed Service Manual man336 NOTES...

Страница 121: ...ts List Service Parts Ordering Using the parts lists in this manual identify the necessary part number s Find the product number and serial number on the product identification label A see figure 5 1...

Страница 122: ...ll orders shipped by Hill Rom nominated unless specified Address all inquiries to ATTN TECHNICAL SUPPORT PARTS HILL ROM COMPANY INC 1069 STATE ROUTE 46 E BATESVILLE IN 47006 9167 Address all return go...

Страница 123: ...full selling price only for internal use at Hill Rom Do not confuse this price with your price Do not return any parts without an RMA number When parts products have been requested to be returned Hill...

Страница 124: ...Exchange Policy Chapter 5 Parts List Page 5 4 Basic Care Bed Service Manual man336 NOTES...

Страница 125: ...tion of problem A qualified specialist will provide via telephone United States and Canada or FAX Outside the United States and Canada troubleshooting assistance for facility personnel and provide nec...

Страница 126: ...ox assemblya 120 V 127 V 131523 2 Control box assembly a 220 V 230 V 131524 2 Control box assembly a 240 V 131493 2 Head foot board assembly 131480 2 Head siderail assembly right hand 131482 2 Head si...

Страница 127: ...u Philippines U S Vir gin Islands Lebanon Mexico Saudi Arabia Aruba Antigua Barbuda and Brit ish Virgin Islands 131504 5 Power cordb CG 4 19 Australia and New Zealand 131505 5 Power cord b CG 5 Chile...

Страница 128: ...Base Frame Chapter 5 Parts List Page 5 8 Basic Care Bed Service Manual man336 Base Frame Figure 5 2 Base Frame...

Страница 129: ...ot available 3 Caster hex rod 5 131421 5 Rocker arm assembly Basic Care 6 131467 2 Brake steer pedal pad set 7 131425 1 Brake steer hex rod center 8 131433 1 Brake steer caster central brake and steer...

Страница 130: ...r the Electric Model Basic Care Bed Chapter 5 Parts List Page 5 10 Basic Care Bed Service Manual man336 Upper Frame for the Electric Model Basic Care Bed Figure 5 3 Upper Frame for the Electric Model...

Страница 131: ...95 1 Bolt assembly hilo drive 10 131438 2 Trendelenburg assembly short length as applicable 11 131437 2 Trendelenburg assembly standard length as applicable 12 131526 1 Control panel foot end 13 13152...

Страница 132: ...for the Manual Model Basic Care Bed Chapter 5 Parts List Page 5 12 Basic Care Bed Service Manual man336 Upper Frame for the Manual Model Basic Care Bed Figure 5 4 Upper Frame for the Manual Model Bas...

Страница 133: ...Spring manual crank assembly 3 131438 1 Trendelenburg assembly short length as applicable 4 131437 1 Trendelenburg assembly standard length as applicable 5 131435 4 Trendelenburg indicator 6 131497 3...

Страница 134: ...Siderails Chapter 5 Parts List Page 5 14 Basic Care Bed Service Manual man336 Siderails Figure 5 5 Siderails...

Страница 135: ...traction bar assembly foot left hand 9 131486 1 Siderail assembly foot left hand 10 131485 1 Siderail retraction bar assembly foot right hand 11 131484 1 Foot siderail assembly right hand 12 131477 2...

Страница 136: ...Power Cord Chapter 5 Parts List Page 5 16 Basic Care Bed Service Manual man336 Power Cord Figure 5 6 Power Cord...

Страница 137: ...cord CG 6 Japan 3 131507 Power cord CG 7 Taiwan 4 131509 1 Power cord CG 9 Argentina 5 131510 1 Power cord CG 10 20 India and Pakistan 6 131501 1 Power cord CG 1 15 Bolivia Brazil Indonesia South Kor...

Страница 138: ...p Surface Item Number Part Number Quantity Description 1 131469 1 Head panel electric standard length 2 131470 1 Head panel electric short length 3 131471 1 Head panel manual standard length 4 131472...

Страница 139: ...h or serious injury personal injury SHOCK HAZARD Disconnect the unit from its power source Failure to do so could result in personal injury or equipment damage SHOCK HAZARD Do not expose the unit to e...

Страница 140: ...urate the spot Disinfecting Dilute disinfectants and germicides as specified on the manufacturer s label Component Handling CAUTION Make sure that your hands are clean and only handle the P C boards b...

Страница 141: ...rd in an antistatic protective bag Lubrication Requirements WARNING Obey the product manufacturer s instructions Failure to do so could result in personal injury or equipment damage CAUTION Do not use...

Страница 142: ...al Procedures Page 6 4 Basic Care Bed Service Manual man336 SA3352 1400 Gear grease use on motor gears only SA0646 1400 Teflon a lubricant dry a Teflon is a registered trademark of E I du Pont de Nemo...

Страница 143: ...only meet JCAHO requirements but will help make sure of a long operative life for the Basic Care Bed PM will minimize downtime due to excessive wear The following PM schedule guides you through a nor...

Страница 144: ...es Examine the caster tires for cuts and wear Replace the caster if nec essary Siderail controls Make sure the switches in the siderails operate correctly Siderail Examine the siderail and make sure i...

Страница 145: ...burg Lockout controls Brake and steer option Individual brake caster Individual brake steer caster Caster tires Siderail controls Siderail Model Number Pivot points Power cord and connector Headboard...

Страница 146: ...s head screwdriver 1 Raise the head section A to its full up position 2 Lower the bed to have access to the cover B on the control box assembly C see figure 6 1 on page 6 8 Figure 6 1 Control Box SHOC...

Страница 147: ...edures Basic Care Bed Service Manual man336 Page 6 9 6 6 Make sure the indicator G turns off 7 Connect the battery cable E to the control board F NOTE The indicator should remain off 8 Install the six...

Страница 148: ...he CPR wire D to the bracket E b Pull a small part of the CPR wire D out of the cable F c While holding the CPR wire D tightly tighten the screw C to attach the wire D to the bracket E NOTE While the...

Страница 149: ...D to the bracket E b Allow a small part of the CPR wire D to go into the cable F c While holding the CPR wire D tightly tighten the screw C to attach the wire D to the bracket E NOTE While the CPR wir...

Страница 150: ...6 2 CPR Cable Adjustment Chapter 6 General Procedures Page 6 12 Basic Care Bed Service Manual man336 NOTES...

Страница 151: ...7 1 7 Chapter 7 Accessories Basic Care Bed Accessories Table 7 1 Accessories List Product Number Description P1445 IV pole P846D48 Trapeze support light neutral P847B Fracture frame adapter set 3 4 pi...

Страница 152: ...two at the head end and two at the foot end Installation WARNING Failure to install the IV pole correctly could allow it to fall resulting in personal injury or equipment damage Put the IV pole A into...

Страница 153: ...445 Chapter 7 Accessories Basic Care Bed Service Manual man336 Page 7 3 7 Adjustment To change the height of the IV pole A pull the knob C out and adjust the upper section D of the IV pole A to the ne...

Страница 154: ...tall the four bolts A and two backing plates B to attach the trapeze support C and on the bed frame D see figure 7 2 on page 7 4 Figure 7 2 Trapeze Support CAUTION Do not lower the bed frame while the...

Страница 155: ...Support P846D48 Chapter 7 Accessories Basic Care Bed Service Manual man336 Page 7 5 7 Removal 1 Loosen and remove the four bolts A and the two backing plates B 2 Remove the trapeze support A from the...

Страница 156: ...ve a 3 4 inside diameter while the P847C sockets have a 1 2 inside diameter Please specify which size is necessary when ordering Installation 1 Put the short adapter tubes A into the IV sockets B at t...

Страница 157: ...B and P847C Chapter 7 Accessories Basic Care Bed Service Manual man336 Page 7 7 7 2 Remove the long adapter tubes C from the IV sockets D at the foot end of the bed by lifting the tubes C upward 3 Do...

Страница 158: ...7 3 Fracture Frame Adapter P847B and P847C Chapter 7 Accessories Page 7 8 Basic Care Bed Service Manual man336 NOTES...

Страница 159: ......