2.13 The Trendelenburg or Reverse Trendelenburg Function Does Not Operate
Chapter 2: Troubleshooting Procedures
2
Hill-Rom® Basic Care™ Bed, Hill-Rom® 305 Manual Bed,
Page 2 - 23
Hill-Rom® 405 Electric Bed Service Manual (MAN336 REV 2)
2.13 The Trendelenburg or Reverse Trendelenburg Function Does
Not Operate
When you put the bed into Trendelenburg or Reverse Trendelenburg, the bed
lowers fully but does not go into Trendelenburg or Reverse Trendelenburg.
1. Visually examine the Trendelenburg or Reverse Trendelenburg handles
(A) (see figure 2-1 on page 2-24). The Trendelenburg or Reverse
Trendelenburg handles (A) are connected and not broken.
Yes
No
↓
→
Replace the Trendelenburg box assembly (B) (refer to procedure
4.3 on page 4-7). If this corrects the problem go to “Final
Actions” on page 2-4.
2. Visually examine the Trendelenburg or Reverse Trendelenburg rods (C).
The Trendelenburg or Reverse Trendelenburg rods (C) are not bent.
Yes
No
↓
→
Straighten or replace the Trendelenburg or Reverse
Trendelenburg rods (C) as necessary (refer to procedure 4.4 on
page 4-9). If this corrects the problem go to “Final Actions” on
page 2-4.
3. Visually examine the Trendelenburg or Reverse Trendelenburg
mechanisms (D). The Trendelenburg or Reverse Trendelenburg
mechanisms (D) are connected and not broken.
Yes
No
↓
→
Connect or replace the Trendelenburg or Reverse
Trendelenburg mechanisms (D) as necessary (refer to procedure
4.4 on page 4-9). If this corrects the problem go to “Final
Actions” on page 2-4.
4. Call Hill-Rom Technical Support at 800-445-3720 (USA).