118
Use keys▲▼to move the indicator „>‟ up and down to the
wanted opt. Press soft key „Enter‟ to confirm your choice. The
instrument will display box of „saved‟ and the setting remains
even after shutting down. Afterwards, the default EDM mode
of the instrument will be that of your choice until next time you
change the setting.
All the other settings under Config menu are similar to
„Mode of EDM‟. Please refer to the explanation above.
Notes: the NEZ or ENZ displaying mode of coordinates
are in relation to the order of the exported coordinate only.
9.6 Calibrate &config
9.6.1 Adjusting index error(I.E)
To adjust index error, the system will first ask you to „Aim
at a target F1‟. Users may press „ENT‟ to confirm after aiming
and the instrument will measure its horizontal angle
automatically. The HA will be displayed for one second and the
system will ask you to „Aim at a target F2‟. Following the same
Calibration
1.Calibrate I.E
2.Calibrate LILT
3.Add const
4. Mul. const
5.Date&Time