65 / 88
en
Rev. 03 / 02.2021
AB3200
Software description
-] Settings
PROG
T/°C
RCF
RPM
t/min
Multi programs = on
PROG
T/°C
RCF
RPM
t/min
Store Settings...
PROG
T/°C
RCF
RPM
t/min
-] Settings
PROG
T/°C
RCF
RPM
t/min
PC Address =A
PROG
T/°C
RCF
RPM
t/min
Store Settings...
PROG
T/°C
RCF
RPM
t/min
7.3.5.8
Multi programs
Please observe the notice
Further information on page 55
The program link can be activated or deactivated as follows:
1.
Change to the menu
‘Machine Menu
🠂
2.
Press the
[PROG] key until the ‘Multi programs’ parameter is dis-
played.
3.
Set
‘off’ or ‘on’ using the [adjusting knob].
‘off’
Program link is deactivated.
‘on’
Program link is activated.
4.
Press the
[START] key to save the setting.
5.
[STOP/OPEN] key Press 1x
⇒
Back to the menu
‘Machine Menu’.
6.
[STOP/OPEN] key Press 1x
⇒
Back to the main display
7.3.5.9
PC Address
The ‘PC Address’ is the address of the device (serial inter-
face).
The address is set to ] = 29th address ex works.
To set the PC address, proceed as follows:
1.
Change to the menu
‘Machine Menu
🠂
2.
Press the
[PROG] key until the ‘PC Address’ parameter is displayed.
3.
Use the
[adjusting knob] to set the desired value.
4.
Press the
[START] key to save the setting.
5.
[STOP/OPEN] key Press 1x
⇒
Back to the menu
‘Machine Menu’.
6.
[STOP/OPEN] key Press 1x
⇒
Back to the main display
7.3.5.10
Cool acc time
If necessary, it can be set that cooling should occur with a
time delay after the start of the centrifugation run. The delay
time can be set from 15 to 900 seconds in 1-second incre-
ments. If no time delay is desired, ‘0’ must be set. No delay
time is set ex works.
The delay time can be set as follows: