en
Rev. 03 / 02.2021
AB3200
54 / 88
Software description
(1) 20 9/;\9 1500 00:00
PROG
T/°C
RCF
RPM
t/min
1 20 9/;\9 1500 2:00
PROG
T/°C
RCF
RPM
t/min
PREC 20 9/;\9 500 01:00
PROG
T/°C
RCF
RPM
t/min
4.
Press the
[START] key to start the run.
🡆
The LED in the
[START] key flashes until the rotor is read in; then
the LED is illuminated. The timer begins at
‘00:00’.
5.
Press the
[STOP/OPEN] key to end the centrifugation run.
🡆
The run-down is performed with the selected run-down parame-
ters
Chapter 7.2.2.2 ‘Braking stage and run-down time’
on page 49. The run-down parameter is displayed, e.g. ‘9’. The
right LED in the
[STOP/OPEN] key lights up. After the rotor comes
to a standstill, the LED in the
[START] key goes out and ‘OPEN /
OEFFNEN’ is displayed. The right LED in the [STOP/OPEN] key
also goes out, the left LED in the
[STOP/OPEN] key starts to flash
and flashes until the lid is opened.
7.2.6
Precooling of the rotor
To quickly precool the unloaded rotor and accessories, a
centrifugation run at a speed of approx. 20 % of the max-
imum rotor speed is recommended. Precooling is done
automatically with the ‘PREC’ (PRECOOLING) program.
Precooling is not possible if program linkages are being
used.
To set the precooling, proceed as follows:
1.
Close the lid.
2.
Press the
[Rotor precooling] key.
🡆
The LED in the
[Rotor precooling] key flashes until the rotor is read
in; then the LED is illuminated.
You can make changes to the ‘PREC’ program in
the same way as all program changes are made.
The ‘PREC’ program is located after program
location 98.