6.2
Opening the lid
Personnel:
■
Authorized skilled personnel with special instruction
The lid can only be opened if the device is switched on and
the rotor is stopped. If it is not possible to open the lid,
actuate the emergency release
To open the lid, proceed as follows:
1.
Switch the device on.
2.
Press the
[STOP/OPEN] key on the control panel.
🡆
The lid unlocks and the left LED of the
[STOP/OPEN] key goes out.
3.
Open the lid as far as it can go.
6.3
Closing the lid
Personnel:
■
Authorized skilled personnel with special instruction
WARNING
Risk of crushing!
Do not reach with your hands between the lid and the
housing.
To close the lid, proceed as follows:
1.
Close the lid of the device.
2.
Lightly press the lid down.
🡆
The lid is locked. The left LED in the
[STOP/OPEN] key lights up.
6.4
Quick-stop function
If the
[STOP/OPEN] key is pressed twice in a row, the quick-stop function
will be triggered.
With the quick-stop function, the run-down takes place at braking level
;\_
9 (shortest run-down time). Braking level 9 is displayed. If braking level 0
was preselected, the run-down takes place at braking level
;\_
9d. At
braking level 9d, the run-down time is longer than at braking level 9 due to