HE 5411 Operating Instructions
#340411 | version 1.7
10
Error Messages
Display
Cause
Troubleshooting
The display shows
alternating with the
currently measured value.
Measurement range overshot
or pressure connections reversed
Check the
measurement range
set
Check the pressure
connections
The displays shows
alternating with the
currently measured value.
Measurement range underrun
or pressure connections reversed
Check the
measurement range
set
Check the pressure
connections
The display shows
alternating with
The calibration data of the sensor
is not valid.
The error cannot be
solved by the
customer.
Please contact
HESCH service
(see
chapter 12
Maintenance and
Service)
The display shows
alternating with
The sensor is missing or faulty
The error cannot be
solved by the
customer.
Please contact
HESCH service
(see
chapter 12
Maintenance and
Service)
The display shows
alternating with the
process value or
displayed value.
The stored parameters are not
valid (e.g. after a firmware
update)
Write parameters with
the software
“EasyTool Controls”
Or set a parameter at
the device , e.g.
‘Zeroing’ or ‘Limit’