Herida Healthcare Argyll II HH/P09 Скачать руководство пользователя страница 1

 

 

1 | 

P a g e  

J A N U A R Y   2 0 2 0

 

 

Argyll II

™ User/Service Manual 

 

 

 

 

 

 

 

 

 

 

 

 

 

Herida II 

 

 
 

 

 
 
 
 

NB: Please ensure this document is fully read & understood  

by the Clinician / Carer / Service Engineer prior to use. 

 
 

User & Full Service Manual 

Pressure Relieving / Alternating Air Mattress & Pump 

 

Pump Model: HH/P09  Mattress Model: HH/T04 

 

Содержание Argyll II HH/P09

Страница 1: ...Service Manual Herida II NB Please ensure this document is fully read understood by the Clinician Carer Service Engineer prior to use User Full Service Manual Pressure Relieving Alternating Air Mattre...

Страница 2: ...2 P a g e J A N U A R Y 2 0 2 0 Argyll II User Service Manual...

Страница 3: ...ON 6 Static Mode Alternating Mode Profiled Bed Clinical Considerations Installation Safety Instructions Pump Operation Mattress Pump Characteristics MAINTENANCE 10 Cleaning Maintenance Repair Storage...

Страница 4: ...g solutions Keep sharp objects away from the mattress Ensure that the system is clean and dry prior to storage Store the pump and mattress in the protective bags supplied We recommend the use of cot s...

Страница 5: ...under patient weight as is often the case Application This product is to be used for those patients who have existing tissue damage or pressure ulcers It can also be used as a high risk preventative...

Страница 6: ...ngly to appropriately accommodate see below Unless the clinical condition dictates it is not advised to adjust the bed profile beyond 45 The Argyll II will continue to operate beyond this but may caus...

Страница 7: ...on the bed before unfolding the mattress laying out flat The air inlet should be located at the feet side of the bed 3 Use the Rapid Inflator if available to inflate the mattress before connecting th...

Страница 8: ...s Pressure Monitor The light is illuminated when pressure is normal extinguished when pressure is abnormal The light will continuously flash during the alternating cycle change over Mute When pushed t...

Страница 9: ...stance to rectify Power Off Alarm The alarm will sound if power is lost to the pump Pump Characteristics Power AC220V AC110V Pressure Range 10 52 mmHg Internal Fuse F0 5A 250V Plug Fuse 3A Frequency 5...

Страница 10: ...da Healthcare Ltd and only original manufacturers parts should be used Failure to follow the instructions could result in a void warranty situation Service Interval Instructions This product must be s...

Страница 11: ...ystem soak test over at least 3 cycles or under weights to replicate patient in situ The tubing and connectors that run down the side of the mattress must be clear of kinks debris or must not be snapp...

Страница 12: ...le on page 9 using a Herida Testing System Calibration training must be conducted applied by a Herida trained technician Other Considerations Original internal fuse is F0 5A 250V When replacing the fu...

Страница 13: ...luding cells tubing and CPR toggle Ensure all cell fasteners are snapped to the mattress base and are not lose or faulty Check that stitching which secures the straps to the base of the mattress is so...

Страница 14: ...14 P a g e J A N U A R Y 2 0 2 0 Argyll II User Service Manual P09 3 Way Pump...

Страница 15: ...against electric shock Class II electrical appliance Double insulated Dispose of electrical equipment correctly WEEE Do not use broken or damaged electrical plugs or cables as this may lead to electri...

Страница 16: ...ctors Repair the air tubes or call for replacement Re attach CPR plug 3 Mattress is too hard or too soft The air pressure is not adjusting to required level Adjust the pump pressure according to the u...

Страница 17: ...Leeds LS27 0LL Freephone 44 800 193 6030 Email sales heridahealthcare co uk Company Number 9918019 VAT Registration Number 233587691 Disclaimer Please be aware that all products should be disposed of...