4.2 Lance Installation
1. Open the Digilance V by removing the 4 Screws at the Corners of the Front Plate
2. Insert the Lance Pipe and secure with the 2 Bolts
3. Connect the cables from the Housing to the Instrument Assembly according to the diagram
•
Conductive Shielding of the Thermocouple Extension
Cable may not extending into the instrument housing.
•
This could create a short circuit of the electronic
components or measurement circuit.
4.2.1 TC Wire Color Codes
EU
US
T+
T-
T+
T-
Type S/R
Orange
White
Black
Red
Type B
Grey
White
Grey
*Black*
Red
*White*
Type K
Green
White
Yellow
Red
* Color of Contact Block Lead Wires*
Antenna
Connection
Battery Pack
Connection
Thermocouple
Positive (+)
Thermocouple
Negative (-)