TB20, strain gauge weighing module | Edition 1 | 21.12.2018
40
5.2.7
Linearization through sampling points
In the course of the configuration it can be indicated whether linearization with up to 5 sampling
points should be carried out by the user. The characteristic of the load cell is linearized.
When setting a sampling point, the load cell is loaded with a defined weight. By way of a control
command, the module is informed for which weight and for which sampling point a conversion
factor should be created. The linearization values of the five sampling points are customer parameters.
The sampling point values must be transferred from the user program to the weighing module or be
set with the help of the TB20 ToolBox. The sampling point values are permanently stored in the
weighing module. The sampling point values can be changed by overwriting. Should one or more
sampling points not be used, deactivation must take place via the parameter "Number of sampling
points", see chapter 8.
Linearization through sampling points
Blue line - defined by zero point and max. quiescent point
Green line - linearized with sampling points
Red line - linearized with sampling points and relocated by offset.
The sampling points always refer to the zero point and not to the offset. The sampling point values
must be provided in kg with 2 decimal points.
No linearization takes place if the sampling points are deactivated. The zero point parameter and the
parameter "Characteristic value of the load cell" apply in connection with "Rated load of the load cell."
Linearization can take place with up to five decimal points, whereby the values of the sampling points
must be ascending in a strictly monotonic manner. All sampling points can be created, but not all
sampling points need be used. Usage takes place in an ascending sequence: first sampling point 1,
sampling point 5 last. The amounts of the sampling points must be smaller than the rated load of the
load cell. See graph: Linearization through sampling points.
The number of sampling points to be used is determined through the parameter "Number of sampling
points used."