Appendix A
76
Convectron
®
ATM Vacuum Gauge Module Instruction Manual
Specifications for analog output version
Analog output
0.375 to 5.659 VDC for 0 to 1000 Torr, non-linear, absolute of N
2
Setpoint relays
Two single-pole, double-throw (SPDT) relays. Use potentiometers and test
points on the module front panel and jumpers inside the module to
configure relays.
Relay 1
• Relay 1 is assigned to the Convectron gauge, which indicates vacuum
chamber pressure
• Relay 1 activates with decreasing and deactivates with increasing
vacuum chamber pressure
• Hysteresis: 30 mV
• Range: 1 x 10
–3
to 1000 Torr
Relay 2
• Relay 2 is assigned to the differential pressure sensor, which indicates
pressure differential between atmospheric and vacuum chamber
pressures
• In default mode, relay 2 activates with increasing and deactivates with
decreasing differential between atmospheric and vacuum chamber
pressures
• Hysteresis: 8 Torr
• Range: –750 to +125 Torr
Analog I/O male connector on module end panel
Analog output
5
Power ground
4
11–26 V power input
3
Setpoint 2 adjust
2
Setpoint 1 adjust
1
15
Relay 1 common
14
Relay 1 normally closed
13
Relay 2 normally closed
12
Relay 2 common
11
Relay 2 normally open
10
Relay 1 normally open
9
No connection
8
No connection
7
Diaphragm zero adjust
6
Signal ground
Содержание GRANVILLE-PHILLIPS Convectron ATM
Страница 2: ......
Страница 4: ......
Страница 8: ...iv Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 12: ...4 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 80: ...72 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 90: ...82 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 97: ......