Chapter 4
58
Convectron
®
ATM Vacuum Gauge Module Instruction Manual
4.12
Factory defaults
Table 4-11 lists default settings for the RS–485 version of the module. After
you’ve reconfigured the module, you may restore all parameters to their
default values by cycling power to the module or by sending a factory reset
(FAC) command.
FAC command from host:
#01FAC
↵
Response from module:
*01 PROGM OK
↵
Reset command from host:
#01RST
↵
Response from module:
None
Table 4-11 Factory default settings for RS485 module
Parameter
Default setting
Digital communication
• Parity: 8 data bits, no parity, 1 stop bit
• Baud rate: 19.2 kBaud
• Module address: 01
Relays 1 and 3
• Setpoint: Activates at 1 x 10
3
Torr vacuum chamber pressure
• Activation direction: Relay activates above vacuum chamber pressure setpoint (–)
• State: Deactivated (resets relays 1 and 3 to deactivated state)
• Returning pressure hysteresis: Determined by vacuum chamber pressure
Relays 2 and 4
• Setpoint: Activates at 2.50 x 10
2
Torr differential between atmospheric and vacuum
chamber pressures
• Activation direction: Relay activates above differential pressure setpoint (+)
• State: Deactivated (resets relays 2 and 4 to deactivated state)
• Hysteresis: determined by vacuum chamber pressure
Optional display
• Pressure value: Vacuum chamber pressure
• Pressure unit: Torr
Содержание GRANVILLE-PHILLIPS Convectron ATM
Страница 2: ......
Страница 4: ......
Страница 8: ...iv Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 12: ...4 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 80: ...72 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 90: ...82 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 97: ......