Step 5
Configure setpoint relays for the application
• To configure setpoint relays for the analog version of the module, see
• To configure setpoint relays for the RS–485 version of the module, see
If the module will measure the pressure of a gas other than N
2
or air, you
must
adjust relay setpoints for the process gas. The true pressure of a gas
other than N
2
or air may be substantially different from the pressure that the
output indicates. For example, outputs might indicate a pressure of 10 Torr
for argon, although the true pressure of the argon is 250 Torr. Such a
substantial difference between indicated pressure and true pressure can
cause overpressurization resulting in an explosion.
Step 6
Calibrate the Convectron gauge
Calibration improves the accuracy and repeatability of the Convectron
gauge.
• To calibrate the Convectron gauge for the analog version of the module,
• To calibrate the Convectron gauge for the RS–485 version of the module,
Atmospheric pressure
calibration
An atmospheric calibration is performed on the Convectron gauge, using
N
2
, at the factory before the module is shipped. The factory calibration sets
the atmospheric calibration point to 760 Torr of N
2
.
Because performance varies depending on the process gas, you may wish
to reset the atmospheric calibration point if a gas other than N
2
or air is
being used. Periodic resets of the atmospheric calibration point also
improve the accuracy and repeatability of the Convectron gauge near
atmospheric pressure, even if the process gas is N
2
or air.
Vacuum chamber
pressure calibration
Periodic resets of the vacuum chamber pressure calibration point improve
the accuracy and repeatability of the Convectron gauge.
WARNING
Failure to use accurate pressure conversion data for N
2
or air to other gases can cause an explosion due to
overpressurization.
If the module will measure any gas other than N
2
or air, before
connecting relays to system control devices, adjust setpoints for
the process gas that will be used.
Содержание GRANVILLE-PHILLIPS Convectron ATM
Страница 2: ......
Страница 4: ......
Страница 8: ...iv Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 12: ...4 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 80: ...72 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 90: ...82 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 97: ......