On-Board FastRegen Control Module Programming and Operation Instructions
P/N 8040406
3-41
C
HELIX TECHNOLOGY CORPORATION
-
TI
CRYOGENICS
Time Since Last Fast Regeneration
1. Press the SERVICE key on the keypad/display. The serial
number and software version information are displayed.
2. Press the Next key until the Fast RGN Hours are displayed.
3. Press the MONITOR key. The On-Board microprocessor will
monitor system operation.
Password
1. Press the SERVICE key on the keypad/display. The serial
number and software version information are displayed.
2. Press the Next key until the Password prompt is displayed.
3. Press the Enter key. 0 = Password is displayed.
4. Press the Enter key again. The current value is underlined and
ready to accept a new value.
5. Press the numeral keys for the desired password. The range of
values is 0 - 32767.
6. Press the Enter button.
7. Press the MONITOR key. The On-Board microprocessor will
monitor system operation.
NOTE: Refer to Appendix A and contact the local customer support center
if the password is forgotten. A master password can be used to access the
original password.
Lock Mode
NOTE: The Lock Mode parameter cannot be accessed unless a password
has been entered.
1. Press the SERVICE key on the keypad/display. The serial
number and software version information are displayed.
2. Press the Next key until the Password? prompt is displayed.
3. Enter the password and press Enter.
4. Press the Next key until the current status of Lock Mode is dis-
played.
5. Press 1 to enable or press 0 to disable lock mode.
6. Press the MONITOR key. The On-Board microprocessor will
monitor system operation.