
10.4 Current Clamp Recording
57
manually). After an iteration or two, it should be possible to reduce the transient to only a few percent of its
original amplitude. However, if the cell has an unfavorable shape (for example, a long cylindrical cell or one with
long processes), the cell capacitance transient will not be a single exponential, and the cancellation will not be as
complete.
10.3.3
Series Resistance Compensation
Series resistance (Rs) compensation is important when the membrane capacitance is large or when the ionic currents
are large enough to introduce voltage errors. To use Rs-compensation, you first adjust the transient-cancellation
controls (including
C-fast
and
τ
-fast
if necessary) to provide the best cancellation. Then you select
Rs Comp
by
turning up the
%-comp
control to provide the desired degree of compensation.
Note:
The ”R-series” control determines (along with the
%-comp
control) the amount of positive
feedback being applied for compensation. It should be adjusted with some care, since too high a setting
causes overcompensation (the
EPC 10 USB
will think that Rs is larger than it is); this can cause
oscillation and possible damage to the cell under observation.
How you should set the Rs-compensation controls depends on the approximate value of the uncompensated
membrane-charging time constant
τ
u
, which you can calculate as the product of the
C-slow
and
R-series
set-
tings (for example, suppose
C-slow
is 20 pF and
R-series
is 10
M
Ω;
τ
u
is then 20
pF
∗
10
M
Ω = 200
µs
. If
τ
u
is
smaller than about 500
µs
, you should use the 2
µs
setting of the Rs-compensation switch to provide the necessary
rapid compensation; however, the slower settings will provide compensation that is less prone to high-frequency
oscillations from maladjustment of the controls. How much compensation you can apply is also determined by
τ
u
.
If
τ
u
is larger than about 100
µs
, you can use any degree up to the maximum of 90% compensation without serious
overshoot or ringing in the voltage clamp response. For smaller values of
τ
u
, the
%-comp
setting should be kept
below the point where ringing appears in the current trace. As in the case of patch recording, there is rarely need
to use the full bandwidth of the
Filter 1
in whole-cell recording. This is because typical membrane charging time
constants (even after Rs-compensation) are considerably longer than 16
µs
, which is the time constant correspond-
ing to a 10 kHz bandwidth. Thus, the current monitor signal is expected to contain no useful information beyond
this bandwidth. In whole cell recording, the voltage and current monitor signals follow the usual convention, with
outward currents being positive. This is because the pipette has electrical access to the cell interior.
10.4
Current Clamp Recording
To switch from voltage clamp to current clamp recording, just select the
C-Clamp
mode in the amplifier dialog.
This provides a recording of the cell membrane potential, which can be monitored at the
V-mon
output of the
EPC 10 USB
.
When you switch to the current clamp mode, the following things happen inside the
EPC 10 USB
main unit:
I-membrane
is set to a value that will keep the membrane voltage constant, internal stimulation is stopped, the
10 kHz Bessel characteristic for
Filter 1
is selected,
C-Slow Range
is set to ”Off”, and the value of
C-fast
is reduced
by 0.5 pF to avoid oscillations. Rs-compensation will now act as a ”bridge balance”. For the user, these changes
are of little consequence and are mainly designed to make current clamp recording simple and reliable.
Once you have switched to current clamp, you can use
I-membrane
to set a holding current, and you can apply
stimulus pulses via ”External Stim.Input CC”. The scaling of the ”External Stim.Input CC” can be set with the
CC-Gain
control of the
EPC 10 USB
amplifier dialog. With a
CC-Gain
setting of 0.1 pA/mV, the maximum
current for stimulation is
±
1 nA. With a setting of 1 pA/mV, 10 pA/mV or 100 pA/mV the maximum current is
±
10 nA,
±
100 nA or 1
µ
A, respectively. A
CC-Gain
of 0.1 and 1 pA/mV can only be used in the intermediate
Gain
range of the
EPC 10 USB
and a setting of 10 pA/mV is limited to the low
Gain
range.
http://www.heka.com
Содержание EPC 10 USB
Страница 1: ...Hardware Manual Version 2 8 EPC 10 USB Computer controlled Patch Clamp Amplifier...
Страница 6: ......
Страница 10: ...4 Introduction http www heka com...
Страница 16: ...10 Description of the Hardware http www heka com...
Страница 22: ...16 Installation http www heka com...
Страница 32: ...26 Verifying and Testing the EPC 10 USB http www heka com...
Страница 44: ...38 The control software http www heka com...
Страница 48: ...42 Operating Modes http www heka com...
Страница 54: ...48 Compensation Procedures http www heka com...
Страница 58: ...52 Patch Clamp Setup http www heka com...
Страница 64: ...58 Using the Patch Clamp http www heka com...
Страница 74: ...68 Appendix II Probe Adapters http www heka com...
Страница 76: ...70 Appendix III S Probe http www heka com...
Страница 81: ......