CHAPTER 3
SENSOR CHARACTERIZATION
3
3
Sensor Characterization
3.1
Spectral Response
Figure 1a shows the typical response between 350 nm and
1100 nm of an US-SQS sensor. The solid line shows the ideal re-
sponse of a photon fluence sensor for photosynthetically active radia-
tion (PAR, defined for the 400-700 nm wavelength range; for defini-
tions see chapter 7). For comparison Figure 1b, shows the absolute
response to energy fluence rate from 350-750 nm.
Fig. 1a:
Absolute spectral sensitivity for Photon Fluence Rate
Fig. 1b:
Absolute spectral sensitivity for Energy Fluence Rate
Absolute Spectral Sensitivity S(
λ
) for Photon Fluence Rate
(current per photon events)
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
350
400
450
500
550
600
650
700
750
800
850
900
950 1000 1050 1100
W avelength (nm)
µ
A
p
e
r mmo
l m
-2
s
-1
SQSA0246
Norm
Absolute Spectral Sensitivity S(
λ
) for Energy Fluence Rate
0.000
0.005
0.010
0.015
0.020
0.025
350 400
450 500 550 600
650 700 750
W avelength (nm)
µ
A
per
W
m
-2
SQSA0246
Norm