97
Device description
Cooling water supply
The connections for the cooling water supply are located on the left-hand side panel of
the device, above the receiving flask(s). The following figure shows the connections for
the external chiller and the condenser of the rotary evaporator:
Connection distribution (from top to bottom):
water cooler ← 2
SUPPLY
connection temperature control fluid
FROM
chiller
water cooler ← 1
RETURN
connection cooling liquid
FROM
condenser
water cooler → 2
RETURN
connection cooling liquid
TO
chiller
water cooler → 1
SUPPLY
connection temperature control fluid
TO
condenser
Keep the
supply
and
return
connections closed if no external
chiller is connected!
WARNING: Risk of explosion
In order to not exceed the maximum permissible pressure in
the condenser (2 bar) during operation, open the
supply
and
return
connections, when an external chiller is connected!