Information for OEM and Setup users | Commissioning
4.5
Configuring a touch probe
You can use a touch probe to probe points. The stylus of the touch probe can be
additionally fitted with a ruby ball tip. In order to use a touch probe, you need to
configure the relevant parameters.
Settings
►
Sensors
►
Touch probe
Parameter
Explanation
Touch probe
Activates or deactivates the connected touch probe for use
Setting range:
ON
or
OFF
Default value:
OFF
Diameter
Diameter of the touch probe
Setting range:
≥
0.0001
Default value:
6.0000
Evaluation of the ready signal
Possibility of setting whether the ready signal of the touch probe
should be evaluated, depending on the touch probe
Setting range:
ON
or
OFF
Default value:
ON
Use the touch-probe triggered output of measured values to
automatically send the measured values to a computer when the stylus
is deflected.
"Configuring the measured value output",
4.6
Configuring the axes
The procedure varies depending on the interface type of the connected encoder:
Encoders with EnDat interface:
The encoder applies the parameters automatically
"Configuring axes for encoders with EnDat interface",
Page 86
Encoders with 1 V
pp
or 11 µA
pp
or TTL interface:
The parameters must be configured manually
For the parameters of HEIDENHAIN encoders that are typically connected to the
product, refer to the overview of typical encoders.
"Overview of typical encoders", Page 83
II
HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 06/2022
81
Содержание GAGE-CHEK 2000
Страница 1: ...GAGE CHEK 2000 Operating Instructions Evaluation Unit English en 06 2022...
Страница 3: ...I General information...
Страница 7: ...1 Fundamentals...
Страница 14: ...2 Safety...
Страница 19: ...3 Basic operation...
Страница 43: ...II Information for OEM and Setup users...
Страница 48: ...1 Transport and storage...
Страница 54: ...2 Mounting...
Страница 60: ...3 Installation...
Страница 72: ...4 Commissioning...
Страница 121: ...5 Setup...
Страница 150: ...6 File management...
Страница 157: ...7 Settings...
Страница 172: ...8 Servicing and maintenance...
Страница 184: ...9 Removal and disposal...
Страница 186: ...10 Specifications...
Страница 194: ......
Страница 195: ...III Information for the Operator user...
Страница 240: ...14 What to do if...