Setup | Single steps for setup
8
Adding a function element
Tap
Measure
in the main menu
Drag an empty field of the Inspector to the left into the
workspace
A dialog box for selecting the function element appears
Tap the desired function element, e.g.
Determining the
circle center
(probing)
Tap
Close
The function element is available in the Inspector
Configuring a function
Drag the function element to the left in the workspace
Use the
ON/OFF
sliding switch in the dialog to activate the
desired functions:
MVO
: The measured value is automatically transferred to
the computer
Select preset
: Once probing has completed, you can use
the acquired coordinates as preset
When using the
Determining the circle center
probing
function, select the plane of the circle via the drop-down
lists;
default: plane of the
X
and
Y
axes
Tap
Close
The selection is saved
The product uses the data format for measured value output that is
selected for manual data transmission in the product settings.
"Selecting a data format", Page 143
8.3.4
Configuring the
MinMax
function
The
MinMax
function helps you to acquire the following measured values:
Minimum
Maximum
Range
You activate and deactivate the function by tapping the
MinMax
function element
in the Inspector. If the function is active, the measured values will be captured
and shown in the position display. You can then transfer the measured values to
a computer via the measured value output. If you activate the
MinMax
function
again, then the existing measured values will be overwritten.
The following steps are necessary in order to configure the
MinMax
function:
Add the
MinMax
function element to the Inspector
Select measured values that are to be captured and displayed when the
MinMax
function is active
132
HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020
Содержание GAGE-CHEK 2000 Series
Страница 1: ...GAGE CHEK 2000 Operating Instructions Evaluation Unit English en 08 2020 ...
Страница 12: ......
Страница 13: ...1 Fundamentals ...
Страница 22: ......
Страница 23: ...2 Safety ...
Страница 28: ......
Страница 29: ...3 Transport and storage ...
Страница 34: ......
Страница 35: ...4 Mounting ...
Страница 41: ...5 Installation ...
Страница 51: ...6 Basic operation ...
Страница 74: ......
Страница 75: ...7 Commissioning ...
Страница 119: ...8 Setup ...
Страница 154: ......
Страница 155: ...9 Measuring ...
Страница 169: ...10 File management ...
Страница 176: ......
Страница 177: ...11 Settings ...
Страница 210: ......
Страница 211: ...12 Servicing and maintenance ...
Страница 220: ......
Страница 221: ...13 What to do if ...
Страница 225: ...14 Removal and disposal ...
Страница 227: ...15 Specifications ...