
HBS x-touch 200i
53
EN
Electromagnetic safety
In the development of electrosurgical units, HEBUmedical attaches particular importance to the
stringent electromagnetic emission regulations. Consequently, the solutions we supply have a
correspondingly minimal emission level for the required output. Measurements conform the high level
of electromagnetic safety offered by HEBUmedical electrosurgical units.
Under typical working conditions based on an 8-hour day, a field occurs at a distance of 5 to 15 cm
from the cables. At a distance of 20-40 cm, the value drops well below the prescribed limit.
Electromagnetic fields occur primarily around the cable.
The electrosurgical unit itself does not constitute a significant emitting element.
10 Review of the technical status
Before any procedure, check the cables, accessories (handles, electrodes, forceps) and foot switches
for any signs of visible irregularity. Do not use cables or instruments with broken or defective
insulation. These can result in burns if they come into contact with the skin!
10.1 Mechanical faults
If damage occurs to switches, connecting sockets, houses or the touchscreen, or if the unit is dropped
on the floor, please contact your authorized dealer for advice on further use.
The dealer can arrange for a detailed technical check to be performed.
The manufacturer accepts no liability for calibration work or repairs to the unit carried out by the user.
10.2 Recurring safety inspections
A safety inspection of the electrosurgical system is required once a year. The manufacturer only
permits the use of units for which an up-to-date inspection by suitably trained personnel can be
verified.
All information relating to the inspections (date, inspecting staff member, result etc.) must be recorded
in the unit's log book.
10.3 Servicing and manufacturer address
HEBUmedical GmbH
Badstraße 8
78532 Tuttlingen / Germany
Tel.
+49 7461 94 71 - 0
Fax
+49 7461 94 71 - 22
eMail [email protected]
Web: www.HEBUmedical.de
Содержание HBS x-touch 200i
Страница 6: ...HBS x touch 200i 6 DE Frontansicht des HBS x touch 200i Rückansicht des HBS x touch 200i i ...
Страница 28: ...28 ...
Страница 31: ...HBS x touch 200i 31 EN ...
Страница 34: ...HBS x touch 200i 34 EN Front view of the HBS x touch 200i Rear view of the HBS x touch 200i i ...
Страница 56: ...HBS x touch 200i 56 EN ...
Страница 58: ...HBS x touch 200i 58 FR ...
Страница 84: ...HBS x touch 200 84 FR ...
Страница 86: ...HBS x touch 200i 86 ES ...
Страница 88: ...HBS x touch 200i 88 ES Aplicación universal dado que se puede conectar entre 100 y 260 voltios 50 y 50 Hz ...
Страница 113: ...113 ...
Страница 115: ...HBS x touch 200i 115 IT ...
Страница 141: ...PT HBS x touch 200i 141 ...