background image

POWERING INSTRUMENT

Power ON: Press and hold the middle button for 3 seconds.

RMLD-CS Quick Start Guide

SELF-TEST

Self-Test feature should be performed daily before survey to ensure that the 

instrument is operational. 

Self-Test and Calibration procedure:

1.  Turn on the instrument and allow instrument to successfully boot up.

2.  Press the MENU button and select the SELF-TEST option.

3.  The screen will display and guide the next steps for the operator to  

  follow within 10 seconds (enough time for the steps).

4.  Place the instrument in its designated area in the carrying case, making  

  sure it is all the way in place and flat.

5.  Statuses of test will be present on the display and by audible tones. A  

  continuous one beep every second means the instrument passed the  

 self-test.

  a.  If the instrument fails, two beeps every second will play and operator  

    should repeat self-test. 

  b.  If instrument fails on a third attempt, contact Heath Consultants.

SPOTTER LASER

The green spotter laser in under the control of the operator and is on for one 

minute when the spotter “ON” button has been pressed. 

To use the spotter laser, the user must first ENABLE the laser in the MENU. To 

do this, select MENU, scroll down to SPOTTER CONTROL, then select ENABLE.

www.heathus.com

713.844.1300

Power OFF: Press and hold the middle button for 3 seconds.

The spotter icon will now be visible in the top menu bar. When the spotter laser 

is turned ON, the icon will change from white to green. 

The left menu button will now display the spotter ON 

option. Pressing the left button (on/off) will activate the 

spotter for 1 minute.

DMD MODE

The Digital Methane Detection (DMD) mode greatly enhances the use of the 

RMLD-CS. In most situations, the operator should survey with the DMD mode 

turned on. To turn on the DMD mode, press the MENU button, select AUDIO 

MODE, then select DMD. 