Do not use an Aspirator unless you have been trained to remove cerumen and have the 
necessary attachments that are available from your equipment distributor.  

 

Set-Up for Aspirator 

 

Carefully remove all packing materials. Confirm that the power is in the 

OFF

 position. 

 

Remove the cover of the collection jar and add diluted germicide into canister to the 300 ML line  
(#5 in drawing) approximately 2 inches from bottom

. To help seal the cover (on) the jar, wet the 

inside rim of the cover with a wet paper towel or tissue.  THEN PRESS DOWN FIRMLY. 
 

Connect the short narrow diameter tube (#1 in drawing) with white filter in the center (#2 in drawing) 
to the stainless steel connector on the top of the Pro-Power.

 

Connect the other end of the tubing (#3 in drawing) to (VACUUM) connection on the cover of the 
collection jar.   
Connect one end of the long opaque tubing to (#4 in drawing) to (PATIENT) connection on the cover 
of the collection jar. 
Connect the other end to the Aspirator suction tubes ordered from your equipment distributor. 
Plug the power cord into a grounded outlet and turn the power to the ON position. 
Adjust the pressure regulator clockwise to increase suction and counter clockwise to decrease 
suction. 

 

Follow safety protocols for the disposal of used germicide. 

 

Set-Up for Cleaning Hearing Aids 

 

Designed for hospital use as an Aspirator: Features include: adjustable suction control, cleaning tools (included 
to help remove wax or debris), suction tips to clear the receiver and microphone ports.  Placing the Hearing Aid 
on the standard open end Black suction tube will complete the cleaning by combining suction and vibration.  

 

 

Remove the 

opaque

 Aspirator tubing and replace it with the 

clear

 tubing with the syringe attachment at the 

end for blunt tip suction needles. 
 
Inspect the receiver tube (earmold tubing), microphone port and surface of the hearing aid. 
 
Remove any loose debris with a brush, cleaning tools and paper towel or tissue prior to using the suction tips.  
 
If the receiver tube is blocked with debris, try to loosen it with a brush and or wax-pick. Be careful not to push 
the debris further into the receiver tube. 
 
If the receiver tube is still clogged attach the canal of the hearing aid to the open suction tube and let it run for 
5 minutes.  Try to loosen the debris again with a brush and wax-pick. 
 
Select a suction tip slightly smaller to the diameter size of the receiver (earmold) or microphone port.  