15.2 Error messages
In the event of an error during calibration or in weighing
operation, the following messages may be displayed:
Display error mes
sage
Possible causes
Remedy
Calibration Locked
Jumper for calibration lock in
locked position
Move jumper
Error Calibr.
Jumper
Not possible to save be
cause jumper in locked posi
tion
Move jumper, repeat calibra
tion
Not Available
No scale set
Check setting in
ServiceMode
ADC Defect
ADC Error
A/D converter not supplying
data
Replace A/D converter
Short-circuit in load cell cable
Check wiring
Resolution Error
Internal resolution too low,
must be at least 10 times the
set resolution
Set larger interval
Use a load cell with a lower
nominal load
ADC Over
Out Of Range
A/D converter overmodulat
ing because:
Load cell incorrectly con
nected
Check wiring
Load cell defective
Check load cell
Extreme overload on scale
Relieve load on scale
Overload
Scale overloaded
Relieve load on scale
CPU not receiving data from
weighing interface
Check external and internal
wiring
Underload
Gross weight values less
than -20d (below zero)
Load scale
Set parameter
Underload 20d
to N=Off
Содержание WTX120
Страница 1: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Страница 3: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Страница 184: ...Service password 182 A4500 1 1 HBM public WTX120 ...
Страница 185: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Страница 366: ...Service Passwort 182 A4500 1 1 HBM public WTX120 ...
Страница 367: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Страница 549: ...Mot de passe de service WTX120 A4500 1 1 HBM public 183 ...