29
WE2108
HBM
A1632-2.0 en
HINWEIS
The major part of the parameters can also be entered through the PC−
interface (RS−232 / RS−485). The CD−ROM available as accessory with
the Order No. 1−WE2108/DOC contains apart from the complete docu-
mentation (Reference Manual) of the unit also the setup program
WE2108 Panel. This PC software serves for measured value display and
configuration of the WE2108 through the serial interface.
5.1.11 Error
displays
The permissible range of the display depends upon the nominal value of the
scale and the set operating mode (scale standard).
The following error appears in the display if the measured value is
•
above the max. display limit:
(Striche oben)
•
below the min. display limit:
(Striche unten)
(Please also refer to Section 5.3.1 ”Troubleshooting”)
Further errors are displayed as two−digit code numbers (e.g. Err44). They
should not occur in normal operation.
For rectification refer to Section 8 ”Troubleshooting”.
5.1.12 Stillstand
The zeroing, taring, summing and printing functions are performed only if a
stable value stands in the display. This is designated as standstill and is dis-
played by the appearance of the unit of measurement. The standstill condition
is that the value changes maximally by a specific variation range per time unit.
In the case of fluctuating (wind) loads or a very high scale resolution, possibly
no standstill is reached. In this case a more strongly damping filter or a lower
resolution must be selected in the parameter setting.
Different options for standstill indication can be selected in the menu
Func
,