Electrical connection
38
A03452_15_X00_00
HBM: public
T40B
Assignment for plug 3 - Supply voltage and voltage output signal
6
1
5
7
2
4
3
Device plug
Top view
Plug pin
Assignment
Color
code
1
Torque measurement signal
(voltage output ±10 V)
wh
2
Supply voltage 0 V
bk
3
Supply voltage 18 V to 30 V
bu
4
Torque measurement signal (voltage
output ±10 V)
rd
5
Not in use
gy
6
Shunt signal trigger 5 V to 30 V
gn
7
Shunt signal 0 V
gy
Shielding connected to housing
ground
Assignment for plug 4
TMC - only for connection to the TIM 40/TIM-EC Torque Interface Module
within HBM.
6.4
Supply voltage
The transducer must be operated with a separated extra‐low voltage (nominal
(rated) supply voltage 18 to 30 V
DC
). You can supply one or more torque
flanges within a test bench at the same time. Should the device be operated on
a DC voltage network
1)
, additional precautions must be taken to discharge
excess voltages.
The information in this section relates to the self‐contained operation of the
T40B, without HBM system solutions.
The supply voltage is electrically isolated from signal outputs and shunt signal
inputs. Connect a separated extra‐low voltage of 18 V to 30 V to pin 3 (+) and
Distribution system for electrical energy with greater physical expansion (over several test
benches, for example) that may possibly also supply consumers with high nominal (rated) cur
rents.
Содержание T40B
Страница 1: ...Mounting Instructions Montageanleitung English Deutsch T40B ...
Страница 3: ...Mounting Instructions Montageanleitung English Deutsch T40B ...
Страница 48: ...Maintenance 46 A03452_15_X00_00 HBM public T40B 10 Maintenance T40B torque flanges are maintenance free ...
Страница 70: ...Supplementary technical information 68 A03452_15_X00_00 HBM public T40B ...
Страница 71: ...Mounting Instructions Montageanleitung English Deutsch T40B ...
Страница 163: ...Dimensions Abmessungen T40B A03452_15_X00_00 HBM public 93 ...