Electrical connection
10
A4776-1.0
HBM: public
T21WN
5
Electrical connection
5.1
General instructions
We recommend to use shielded, low-capacitance cable from HBM for the elec
trical connection between torque transducer and measuring amplifier.
With cable extensions it is important to ensure that a good connection is pro
vided, with minimum contact resistance and good insulation. All plug connec
tions or cap nuts have to be tightened firmly.
Do not route measurement cables in parallel to power lines and control circuits.
If this is not possible (for example in cable ducts), maintain a minimum dis
tance of 50 cm and protect the cable with a steel tube.
Avoid transformers, motors, contactors, thyristor controllers and similar
sources of stray fields.
5.2
Connector
The transducer is equipped with a permanent housing connector.
It can be connected to the relevant measurement electronics by using the
transducer connection cable (accessories,
for the transducer connection cable can be found in the following table: