45
Scout 55
HBM
A0236-5.9 en
MODE UA:
Depending on the analogue signal you select, the following options are
possible:
Display
Meaning
UA OFF
−
0 TO 20mA
Output
"
20 mA
4 TO 20MA
4.. 20 mA
UA OFF
−
10 VOLT
/
−
10 V
NOTE
The current output or voltage output selection is made using jumpers on
the amplifier motherboard. The procedure is described on Page 45.
INPUT SIGN.:
For test purposes, a calibration signal and a zero signal can be displayed
instead of the measurement signal. You can choose the following input
signals:
MEAS.SIGNAL
Measuring mode
CAL SIGNAL *
)
The display corresponds to 50 % of the current full scale value
ZEROSIGNAL *
)
Internal zero point
*)
To display the measurement signal, you must return to measuring mode.
CONTACT 1...6:
Remotes are available on the connector strip for controlling SCOUT 55
functions. The pin assignment or allocation of the remotes is freely
configurable. No function is defined for the remotes at the factory.