RM4220
14
A1091-4.0 en/de
HBM
Null
(
mV
V
)
R
Bridge
1000
4
R
W
R
Bridge
The table shows the adjustable misalignments for R
Bridge
=
350
Ω
and
R
Bridge
=
87
Ω.
ZERO
R
ZERO
Zero for R
Bridge
= 350
W
Zero for R
Bridge
= 87
W
Ω
mV/V
mV/V
1
3.9
⋅
10
6
0.020 (factory setting)
0.005 (factory setting)
2
2.2
⋅
10
6
0.035
0.009
3
750
⋅
10
3
0.110
0.028
4
301
⋅
10
3
0.280
0.070
5
121
⋅
10
3
0.710
0.175
6
49900
1.730
0.430
7
21000
4.000
1.000
8
9090
9.200
2.300
Set the coarse value with the
ZERO
DIP switches, the fine adjustment can be
implemented with the
Zero
potentiometer on the front.
5.4
Set amplification
The amplification is set with
GAIN
Gain
potentiometer
(Front). To achieve the highest possible measured value resolution, you should
use the maximum output voltage of the amplifier (
10V).
Partial
load
Nominal
load
Meas
.
range
in
V
10
V
Sensitivity
in
mV
V
Nom
.
meas
.
range
(
Range
)
in
mV
V
The table shows the measuring ranges that can be adjusted with
GAIN
dependent on the bridge excitation voltage V
E
.
GAIN
V
E
= 5V
V
E
= 10V
mV/V
mV/V
1
11.76 - 25 (factory setting)
5.88 - 12.5 (factory setting)
2
6.06 - 12.9
3.03 - 6.45
3
2.94 - 6.25
1.47 - 3.125
4
1.54 - 3.28
0.77 - 1.64
5
0.84 - 1.74
0.42 - 0.87
6
0.46 - 0.96
0.23 - 0.48
7
0.234 - 0.5
0.117 - 0.25
8
Not used
Not used