96
Connection
HBM
QuantumX
6.6.1
MX460 pin assignment
So that insertion or removal of a transducer connection can be unmistakably identified, Pin 4
and Pin 9 in the connector plug must be bridged! If this bridge is missing, no measurement
values will be recorded at the connection!
1
4
5
6
9
11
15
Bridge
Fig.6.9:
Pin arrangement of connection plug, view from the solder side
Pin
Connection
1
TEDS (+)
2
Reference pulse 0
°
(zeroing pulse) (
−
)
3
Reference pulse 0
°
(zeroing pulse) (+)
4
Always connect with Pin 9! (Plug identification)
5
Frequency input f
1
(
−
)
6
TEDS (
−
), signal ground
7
Frequency input f
2
(
−
)
8
Frequency input f
2
(+)
9
Reference voltage V
ref
(2.5 V)
10
Frequency input f
1
(+)
11
Active sensor supply 5 ... 24 V (
−
)
12
Active sensor supply 5 ... 24 V (+)
13
No function
14
f
1
AC+ (for passive inductive transducers)
15
Calibration signal T10F(S) and T40, 5 V/max. 10 mA