82
Connection
HBM
QuantumX
6.1.4
Auto Cal / Auto Adjustment
When the module has been started up, measurement channels with strain gauge full bridges
are cyclically adjusted during run
−
time. This mechanism improves the long
−
term stability
(aging) and, in the event of temperature variation at the place of installation of the measuring
instrument, also the short
−
time stability of a measuring amplifier.
Autocalibration
briefly interrupts measurement and
−
instead of measured values from a
transducer
−
signals from an internal calibration source are transmitted to the A/D converter
(zero and reference signal).
MX840, MX840A and MX440A amplifiers offer an
auto adjustment
function.
In the full/half bridge measurement mode, these amplifiers use a second measuring circuit
that measures in parallel to the input circuit and carries out a calibration cycle at 30 second
intervals. Long
−
term and short
−
time stability are thus guaranteed in this circuit. A patented
method now enables the calibration channel’s accuracy to be transferred to the
measurement channel.
These modules thus attain high stability without the need to cyclically hold the measured
value.
Autocalibration is activated by default. This cyclical autocalibration can also be quickly
deactivated via QuantumX-Assistant and via catmanEASY
®
.
Содержание QuantumX MX1609
Страница 1: ...I3031 8 0 en Operating Manual ...
Страница 2: ...d ...
Страница 31: ...31 Housings HBM QuantumX Fig 4 2 Amplifier MX1609 P in IP65 housing ...
Страница 34: ...34 Housings HBM QuantumX a f 2 5 Fig 4 7 Mounting the CASEPROT ...
Страница 36: ...36 Housings HBM QuantumX Fig 4 10 Attaching the covers ...
Страница 38: ...38 Housings HBM QuantumX Fig 4 13 Close the lever Fig 4 14 Connected housings ...
Страница 99: ...99 Connection HBM QuantumX Connection of the thermocouple plug in the miniature design ...
Страница 144: ......
Страница 180: ...179 Support HBM QuantumX ...